Description
1. Commercial Introduction
Keeping the infusion running is a matter of clinical safety, and the internal battery is the component that guarantees treatment continuity when the pump is not connected to mains power. With this rechargeable battery of 9.6 V and 1.8 Ah for Terumo 171/172 (TE-171/TE-172) infusion pumps, your service will have a reliable, proven-quality replacement ready to install, suitable both for hospital medical electronics fleets and for preventive maintenance programmes for pump fleets. It is the natural replacement for the original factory reference 8N-1200SCK from Terumo. Ref. Cahnos 88888916.
2. Technical Explanation
The Terumo TE-171 (single-channel) and TE-172 (dual-channel) infusion pumps, commercially known as Terumo 171/172, administer fluids, drugs and nutrition at a controlled flow rate over prolonged periods. To guarantee therapy continuity when the device is not connected to the mains —patient transfers, movement within the ward or power cuts—, the pump incorporates an internal rechargeable battery that recharges automatically while it remains plugged into the mains.
This battery is a rechargeable pack of nickel-cadmium (Ni-Cd) chemistry made up of 1.2 V cells in series that deliver a nominal voltage of 9.6 V, with a nominal capacity of 1.8 Ah (1,800 mAh). It is the natural replacement for the original factory reference 8N-1200SCK from Terumo. When the autonomy notably decreases or the pump gives a premature low battery warning, it is time to replace the accumulator. Installation is direct into the device’s battery compartment, without the need for tools or adaptations.
3. Technical Specifications
| Characteristic | Specification |
| Reference Cahnos (SKU) | 88888916 |
| Original equipment reference (OEM) | 8N-1200SCK (Terumo); the previous datasheet transcribed 8N-1200SCX, not verifiable in public sources |
| Compatibility | Terumo TE-171 and TE-172 infusion pumps (Terumo 171/172) |
| Accumulator type | Ni-Cd (nickel-cadmium), rechargeable |
| Nominal voltage | 9.6 V (pack of 8 cells of 1.2 V in series) |
| Nominal capacity | 1.8 Ah (1,800 mAh) |
| Format / approximate dimensions | 89.5 x 45.0 x 44.8 mm (documented format for packs 8N-1200SCK; verify device compartment) |
| Recharge | Automatic, integrated into the pump when connected to mains power |
| Installation | Direct replacement in the equipment compartment, without tools |
| Intended use | Hospital infusion and perfusion: ICU, anaesthesia, oncology, emergency and hospitalisation |
| Presentation | 1 unit |
4. Equivalences and Cross-References
The official factory reference of the original accumulator is 8N-1200SCK, a code consistently documented in industry catalogues and specialist distribution for Terumo TE-171 and TE-172 pumps. It has been detected that the previous datasheet transcribed the code as «8N-1200SCX», which is not found in public sources: before ordering or replacing, confirm the code printed on the label of the pack installed in your pump.
| Manufacturer / Brand | Family / Model | Reference code / SKU | Equivalence level | Technical differences / Remarks |
| Terumo (device manufacturer) | Original battery TE-171 / TE-172 | 8N-1200SCK | Exact reference equivalence (same manufacturer) | Documented OEM reference of the original accumulator; same voltage (9.6 V), same pump family and same pack format. |
| Cameron Sino | Medical battery for Terumo TE-171 / TE-172 | CS-TER171MD / MED-CS-TER171MD | Comparable by category (not interchangeable 1:1) | Ni-MH 9.6 V / 3,000 mAh (28.8 Wh), 89.52 x 45.04 x 44.82 mm; replaces 8N-1200SCK. Different chemistry (Ni-MH vs Ni-Cd) and capacity (3,000 vs 1,800 mAh); verify compartment and connector before replacing. Verified in manufacturer catalogues and authorised distribution. |
There is no equivalence table validated by the device manufacturer (Terumo) that authorises third-party batteries for reference 8N-1200SCK; the references indicated have been cross-checked against battery manufacturer datasheets and authorised distribution catalogues, but the replacement must be validated field by field (voltage, chemistry, capacity, dimensions and connector) before being used in tenders. Additionally, generic Ni-MH packs of 9.6 V / 3,000 mAh advertised as a replacement for the 8N-1200SCK (e.g. code SS-002615 in authorised distribution) circulate in specialist distribution, but are not listed as equivalents because they have not been validated against an official datasheet from the manufacturer that issues them; confirm with the technical department before use. Consult with the technical department before using this reference in public tenders.
5. Selection tips and variants of interest
- Confirm the exact model of your pump: this battery is the one indicated for Terumo TE-171 and TE-172 (171/172). Do not use it in other equipment of the Terumo family with different voltage or connector, such as the Terumo 112 pump (7.2 V / 1.9 Ah, Ref. 88881003), even if the shape of the compartment may appear similar.
- Verify the label of the installed pack: the original reference is 8N-1200SCK. The code «8N-1200SCX» is not verified in public sources, so it must be confirmed before placing the order.
- Ni-Cd chemistry maintenance: to avoid the memory effect, it is recommended to periodically perform full discharge and recharge cycles; replace the battery when the autonomy is noticeably reduced or the device warns of low battery within a few hours of charging.
- Always order by reference: indicate code 88888916 or the original equipment reference (8N-1200SCK) to avoid confusion with other batteries in the Terumo family.
- Mixed fleets and institutional orders: in our catalogue of batteries for infusion and perfusion pumps we have spare parts for the Terumo, B. Braun, Fresenius, Graseby, Carefusion, Foures and other families; for tenders or continuous supply contracts, contact our technical-commercial department to confirm compatibility and a formal quotation.