Description
Battery for Welch Allyn CP100 and CP200 electrocardiograph
Keep your 12-lead Welch Allyn CP100 and CP200 electrocardiographs always operational with this sealed lead-acid rechargeable battery, 6 V and 4 Ah. This is a quality replacement that substitutes the original battery, reference 5200-84, designed to provide stable and long-lasting power to the equipment in demanding clinical environments: practices, hospitals, health centres and home care.
With this battery, your electrocardiograph can perform multiple ECG recordings without interruption, even in emergency situations or when no mains power is available. It is the reliable power solution your technical service needs to extend the equipment’s service life and ensure continuity of care.
Technical explanation: why this battery is critical
The Welch Allyn CP100 and CP200 electrocardiographs are portable devices designed for recording resting 12-lead ECGs. When disconnected from the mains, all the electronics —screen, keyboard and thermal printer— are powered exclusively by the internal battery. A degraded battery causes voltage drops during printing, unexpected shutdowns or the extremely low battery message, which compromises the operability of the service.
According to the manufacturer’s original service manual (reference 704340 Rev D), the device uses a 6 V rechargeable lead-gel battery that allows printing up to 100 continuous ECGs per charge, with a recharge time of approximately 12 hours. The manual also establishes the device’s charging parameters: fast charge voltage between 6.8 and 7.3 V and maintenance voltage between 6.3 and 6.8 V. After replacing the battery, the device asks for date and time to be entered, and it is recommended to reset the battery life counter from the service menu (battery life test) for proper control of the replacement’s condition.
Replacement is carried out through the device’s lower compartment: remove the cover (secured with two screws), disconnect the two connectors respecting polarity and fit the new battery. A simple process that requires no special tools or advanced electronics knowledge.
Technical specifications
| Battery type | Rechargeable sealed lead (SLA/VRLA) |
| Nominal voltage | 6 V |
| Capacity | 4 Ah (24 Wh) |
| Original manufacturer reference | 5200-84 (alternative equipment code: 401370, battery assembly) |
| Compatible equipment | Welch Allyn CP100 and CP200 (12-lead electrocardiographs) |
| Battery life | Up to 100 continuous ECGs per charge (according to manufacturer’s specification) |
| Recharge time | Approximately 12 hours |
| Equipment charging parameters | Fast charge 6.8–7.3 V; maintenance charge 6.3–6.8 V |
| Installation | Lower compartment, two connectors, cover with screws; no advanced knowledge required |
| Yield | Constant and stable power during ECG recording and printing |
| Professional use | Hospitals, clinics, consulting rooms, primary care and medical transport |
| Recycling | Do not dispose of as municipal waste; recycle in accordance with RAEE/WEEE regulations |
Note: the manufacturer’s service manual specifies the lead-gel battery as 6 V with 5 Ah nominal; versions of 4 to 4.5 Ah are commonly distributed on the market for this device. This datasheet corresponds to the marketed version of 6 V and 4 Ah, fully compatible with the charging electronics of the CP100 and CP200.
Equivalences and cross-references
The Cahnos reference 88888388 corresponds to the original Welch Allyn battery 5200-84 (lead, 6 V) used in the CP100 and CP200 electrocardiographs. In the manufacturer’s official service manual (ref. 704340 Rev D), the field-replaceable parts list identifies this battery as ‘battery assembly’ with code 401370. Specialised biomedical battery distributors also document for this device the alternative codes 100660 (SLA version, 6 V and 6 Ah) and 105204 (later lithium iron phosphate version, 6.4 V and 6 Ah).
| Manufacturer / Brand | Family / Model | Reference code / SKU | Equivalence level | Technical differences / Remarks |
|---|---|---|---|---|
| Welch Allyn | CP100/CP200 battery assembly (service manual 704340 Rev D) | 401370 | Exact reference equivalence | Official manufacturer part for the same equipment; same function and same assembly as the 5200-84 battery. |
| Welch Allyn | CP200 battery (SLA 6 V variant) | 100660 | Comparable by category (not interchangeable 1:1) | Documented by specialised biomedical battery distributors as an OEM replacement for CP200; declared capacity of 6 Ah compared to the 4 Ah of this datasheet; verify fit and charging electronics before replacing. |
| Welch Allyn | CP200 battery (LiFePO4 variant) | 105204 | Comparable by category (not interchangeable 1:1) | Later lithium iron phosphate variant (6.4 V / 6 Ah): different chemistry and voltage from the 6 V sealed lead-acid; requires technical verification before any replacement. |
There is no equivalence table validated by the manufacturer; consult the technical department before use in tenders.
Selection tips and variants of interest
- Check the device reference: this battery is intended for the Welch Allyn CP100 and CP200 electrocardiographs. Before purchasing, confirm the device label.
- Verify the chemistry and voltage: the device is designed for 6 V sealed lead-acid; do not install batteries of other chemistry or voltage without prior technical validation.
- After replacement: the device will ask for date and time; reset the battery life counter from the service menu for correct tracking.
- Market documentation: various specialised distributors also associate reference 5200-84 with Welch Allyn monitors from the Lifesign family (4200, 5000, 5200, 52000 and Quick Sign); this compatibility comes from commercial documentation, not from official manufacturer specification.
- Recycling: lead batteries must be recycled at authorised points in accordance with current regulations.
- Other catalogue variants: if you manage other Welch Allyn equipment or electrocardiographs from other brands, see our ECG and monitoring batteries with references 88888655, 88888799, 88888849, 88888181, 88888889 and 88888384.
Are you organising a maintenance order or a tender? Contact our technical department to confirm the exact compatibility of your equipment fleet and to obtain special volume conditions.