Description
1. Reliable power for your Biocare electrocardiograph
The battery for the Biocare ECG-9803 MFR electrocardiograph (ref. Cahnos 88889365) is the power replacement indicated for the portable Biocare ECG-9803 and ECG-9803G electrocardiographs, replacing the original reference HYLB-114A. Designed for cardiology departments, clinics, emergency and primary care, it allows complete cardiac examinations to be performed without depending on a mains power point, keeping the equipment operational during transport and in clinical situations where continuity of recording is critical. It is a rechargeable lithium-ion cell, ready to install and compatible with the device’s original charger.
2. How this battery works
The Biocare ECG-9803 electrocardiograph is a portable electrocardiographic recording device that integrates signal acquisition, display and internal storage. This battery of 14.8 V and 1.35 Ah (19.98 Wh) powers the acquisition module, the display and the recording electronics during away-from-mains working sessions. Its lithium-ion chemistry provides a stable discharge curve: the voltage remains practically constant until the end, which prevents abrupt cut-offs mid-examination and protects the integrity of stored traces. The positive-locking connector fits directly into the equipment compartment, without the need for adapters or wiring modifications, and the internal management system supervises the charge and discharge of the cell, adding an extra layer of safety to the whole.
3. Technical Specifications
| Application | Electrocardiographs Biocare ECG-9803 and ECG-9803G |
| Original reference | HYLB-114A |
| Battery type | Lithium-ion (Li-ion), rechargeable |
| Nominal voltage | 14.8 V |
| Capacity | 1.35 Ah (1350 mAh) |
| Energy | 19.98 Wh |
| Weight | Approximately 0.22 kg |
| Dimensions | 55 x 65 x 28 mm |
| Device manufacturer | Biocare |
| Presentation | Unit |
Note: the voltage, capacity and energy indicated correspond to the declared specification for reference HYLB-114A. Always check the compartment and connector of your unit before replacement.
4. Equivalences and Cross-References
There is no equivalence table validated by the manufacturer Biocare that authorises third-party batteries for the ECG-9803. The references listed below come from replacement battery manufacturers that declare they replace reference HYLB-114A; they are classified as comparable by category (not interchangeable 1:1), so before replacing, each one must be verified individually: voltage (14.8 V), capacity, chemistry (Li-ion), connector and compartment dimensions.
| Manufacturer / Brand | Family / Model | Code / Reference | Equivalence level | Technical differences / Remarks |
| Cameron Sino | Medical Series (Biocare) | CS-ECG803MD | Comparable by category (not interchangeable 1:1) | Li-ion 14.8 V / 1350 mAh (19.98 Wh), declared dimensions 44.40 x 35.70 x 35.00 mm, ≈150-173 g, blue, EAN 4894128124894. Declared as a replacement for HYLB-114A for ECG-9803 and ECG-9803G in multiple specialised distribution datasheets; dimensions different from those in this datasheet: check connector and compartment before replacement. |
| Kingsener | Medical series (Biocare) | HYLB-114A (compatible ECG-9803 / ECG-9803G) | Comparable by category (not interchangeable 1:1) | Li-ion 14.8 V / 1350 mAh (19.98 Wh), 100% new condition with a declared 1-year warranty. Replaces HYLB-114A for the ECG-9803 and ECG-9803G models; confirm pack dimensions and compatibility with the device’s original charger. |
Reference HYLB-1080 appears associated in distribution with the same pack for the ECG-9803 / ECG-9803G platform (same 14.8 V and 1350 mAh cell); it has not been verified against an official Biocare table, so it is considered pending confirmation with the manufacturer before use.
There is no equivalence table validated by the manufacturer that authorises third-party batteries; consult the technical department before use in tenders. Before any replacement with other brands’ spare parts, voltage, capacity, chemistry, connector and dimensions must be checked against the specific unit.
5. Selection tips and variants of interest
Identify the exact model: this battery corresponds to the Biocare ECG-9803 and ECG-9803G electrocardiographs. Also check the label of the installed battery: the original reference is HYLB-114A. Do not swap it for batteries from other models of the same brand without technical validation: each device mounts a cell with specific voltage, capacity and dimensions.
Battery care: keep the cell charged if the device is to remain idle for long periods and replace the battery when the autonomy is noticeably reduced or the device indicates low battery. Use only the electrocardiograph’s original charger to respect the Li-ion charging parameters and avoid habitual deep discharges.
Variants of the same Biocare family at Cahnos: battery for Biocare ECG-3010 MFR HYLB-947 (ref. 88889367, 14.8 V / 1.6 Ah), battery for Biocare ECG-1215 MFR HYLB-952 (ref. 88889366, 14.8 V / 2.6 Ah) and battery for Biocare ECG-1200 MFR HYLB-293 (ref. 88881242, 14.8 V / 2.6 Ah). Do not swap these references between models: each electrocardiograph accepts a specific voltage, capacity and geometry.
For institutional orders, tenders or continuous supply contracts, contact our technical-commercial department to confirm compatibility, lead times and a formalised quotation.