Description
1. Commercial Introduction
Keeping an automated external defibrillator (AED) operational is a responsibility that cannot tolerate failure: in a cardiac arrest, every minute counts. With the original HeartSine PAD-PAK lithium battery 18 V / 1.5 Ah (Ref. Cahnos 88888948) you will have the correct power replacement for HeartSine Samaritan PAD family defibrillators, ensuring the device remains ready to act throughout its 4-year shelf life.
This battery is the power component of HeartSine’s innovative PAD-PAK cartridge system, which integrates battery and electrodes into a single module with only one expiry date to manage. It is the solution that occupational risk prevention services, sports centres, public administrations, clinics and medical electronics departments need to keep their AED fleet always available, backed by the manufacturer’s original part.
2. Technical explanation of the equipment and the battery
The HeartSine Samaritan PAD defibrillators (models 300/300P, 350P, 360P, 450P and 500P) are public-access AEDs that work with a consumable cartridge called PAD-PAK: an all-in-one module containing the pre-connected defibrillation electrode pair and the system battery. By integrating both consumables into a single piece, maintenance is simplified: there is only one expiry date and a single component to replace after activation or depletion.
The battery we supply (original reference PADPAK, original model) uses lithium-manganese dioxide (LiMnO₂) chemistry, a primary (non-rechargeable) lithium technology with high energy density and very low self-discharge, especially suitable for emergency equipment that remains on standby for long periods. It delivers a nominal voltage of 18 V with a capacity of 1.5 Ah, sufficient for ≈60 shocks at 200 J or ≈6 hours of continuous monitoring. When the remaining charge equals 10 shocks or less, the defibrillator emits an audible warning to plan cartridge replacement.
3. Technical Specifications
| Parameter | Specification |
| Reference Cahnos (SKU) | 88888948 |
| Manufacturer part number | PADPAK (HeartSine PAD-PAK cartridge system) |
| Brand / Manufacturer | HeartSine (HeartSine Technologies Ltd) |
| Product type | Original replacement battery for the PAD-PAK cartridge system |
| Technology | Primary non-rechargeable lithium (lithium-manganese dioxide, LiMnO₂) |
| Nominal voltage | 18 V |
| Capacity | 1.5 Ah (1,500 mAh) |
| Yield | ≈60 shocks at 200 J or ≈6 hours of continuous monitoring |
| Shelf life | 4 years from the date of manufacture |
| Low battery warning | Audible device warning when ≈10 shocks remain |
| Approximate weight | ≈0.31 kg |
| Approximate cartridge dimensions | ≈155 × 115 × 27 mm (according to specialised distribution) |
| Compatibility | HeartSine Samaritan PAD 300/300P, 350P, 360P, 450P and 500P |
| Presentation | 1 unit |
| Transport of dangerous goods | UN3090 (lithium metal batteries) |
| Recycling | Return the depleted unit to a WEEE collection point |
Compatibility by defibrillator model:
| AED model | Compatible PAD-PAK cartridges (18 V / 1.5 Ah battery) |
| Samaritan PAD 300 | PAD-PAK-01 |
| Samaritan PAD 300P | PAD-PAK-01, PAD-PAK-03 |
| Samaritan PAD 350P | PAD-PAK-01, PAD-PAK-03, PAD-PAK-07 |
| Samaritan PAD 360P | PAD-PAK-01, PAD-PAK-03, PAD-PAK-07 |
| Samaritan PAD 450P | PAD-PAK-01, PAD-PAK-07 |
| Samaritan PAD 500P | PAD-PAK-03 |
4. Equivalences and Cross-References
The reference Cahnos 88888948 corresponds to the original battery of the HeartSine PAD-PAK system (original reference PADPAK). The manufacturer does not sell this battery as an independent part: the official replacement route is the complete PAD-PAK cartridge (battery + electrodes), so the manufacturer’s own references listed below are those that include this same 18 V / 1.5 Ah lithium battery. They have been cross-checked against the GUDID/FDA regulatory database and authorised distribution catalogues.
| Manufacturer / Brand | Family / Model | Reference code / SKU | Equivalence level | Technical differences / Remarks |
| HeartSine | Adult PAD-PAK cartridge (battery + electrodes) | PAD-PAK-01 (11516-000003) | Exact reference equivalence (same manufacturer) | Original adult cartridge that integrates this same 18 V / 1.5 Ah lithium battery with a pair of pre-connected defibrillation electrodes; a single expiry date (4 years). Registered in GUDID (05060167120466). Compatible with Samaritan PAD 300, 350P, 360P and 450P. |
| HeartSine | Adult PAD-PAK cartridge (battery + electrodes) | PAD-PAK-03 (EAN 5060167121234) | Exact reference equivalence (same manufacturer) | Commercial variant of the adult cartridge with the same non-rechargeable 18 V / 1.5 Ah lithium battery; ≈60 shocks at 200 J. Documented for Samaritan PAD 300P, 350P, 360P and 500P. |
| HeartSine | PAD-PAK cartridge (battery + electrodes) | PAD-PAK-07 (05060167124488 / M727PADPAK07) | Exact reference equivalence (same manufacturer) | Current version of the adult cartridge registered in GUDID/FDA (HeartSine Technologies Ltd): integrated lithium battery with adult electrodes for Samaritan PAD 350P, 360P and 450P. |
There is no equivalence table validated by the manufacturer that cross-references this battery with third-party manufacturers; consult with the technical department before using in a tender. Generic or compatible lithium batteries for other AEDs are not interchangeable 1:1 with the PAD-PAK: the HeartSine cartridge integrates the electrode connector and the device’s cartridge management, so any substitute must be validated against the official documentation of the defibrillator manufacturer.
5. Selection tips and variants of interest
Confirm the exact model of the device: this battery is the one indicated for HeartSine Samaritan PAD 300/300P, 350P, 360P, 450P and 500P defibrillators. Check the model on the device label and the code of the installed PAD-PAK cartridge before ordering.
Manage a single date: the advantage of the PAD-PAK system is that battery and electrodes share an expiry date (4 years from manufacture). Plan preventive replacement before it expires and after any use of the device in a resuscitation.
Do not mix paediatric cartridges: the paediatric PAD-PAK (Pediatric-Pak) is a different cartridge that adjusts the energy for children aged 1 to 8 years; it is not interchangeable with the adult cartridge in terms of clinical indication.
Check the low battery warning: when the device indicates that approximately 10 shocks or fewer remain, replace the cartridge immediately to avoid compromising AED availability.
Transport and storage: as it is a lithium metal battery (UN3090), transport is carried out as dangerous goods; store between 0 °C and 50 °C, in a dry place away from heat sources.
Environmental management: deliver the depleted unit to a collection point in accordance with WEEE regulations; do not dispose of it in the general waste container.
For institutional orders, tenders or continuous supply contracts, contact our technical-commercial department indicating the defibrillator model and the installed cartridge code; we will confirm compatibility and the formalised quotation.