Description
1. Commercial Introduction
Ensure safe and effective cardiac care for the smallest patients with this pair of standard paediatric defibrillator electrodes with external cable (reference Cahnos F7952PW), a high-reliability consumable compatible with Physio-Control LIFEPAK defibrillators and monitors and with equipment from Bexen Cardio, Cardioline and Mindray. Designed for emergency services, paediatric ICU, consulting rooms, medical transport and cardioprotected areas with a paediatric population, it combines the robustness of the standard series with the operability of the pre-connected external cable (lead-out): the connector is always accessible and the unit ready for immediate connection, a decisive advantage in paediatric cardiorespiratory arrest. A cost-optimised solution, compliant with European medical device standards (class IIb) and latex- and phthalate-free.
2. Technical explanation of the equipment and procedure
The disposable multifunction electrodes are the interface between the defibrillator and the patient: through them the unit analyses the rhythm, delivers therapy and monitors the electrocardiographic response. This paediatric model allows external transthoracic defibrillation, synchronised cardioversion and ECG monitoring (up to 24 h), adapted to paediatric physiology (children under 8 years or weighing less than 25 kg). It is not indicated for transcutaneous cardiac pacing (pacemaker): use it only in the declared functions.
In paediatric patients it is essential to attenuate the delivered energy to avoid myocardial injury; therefore this electrode incorporates a reduced active surface of 40 cm² per electrode and must be used with the unit’s paediatric mode switching or with the reduced energy configuration (50–100 J) indicated in the defibrillator manual. The external cable variant (100 cm) allows the connector to be left accessible and the pad ready, reducing preparation time in emergencies. The low-impedance conductive adhesive gel on 1 mm medical foam with conductive metal foil ensures stable adhesion, uniform current distribution and a reliable signal. They are single-use: they must be replaced after each intervention or upon reaching the expiry date printed on the packaging, being placed according to the anatomical diagrams (sternal and apical position) on the sachet.
3. Technical Specifications
Standard paediatric multifunction electrode from the EuroDefiPads paediatric series, in its 100 cm external cable (lead-out / pre-connected) variant.
| Field | Specification |
|---|---|
| Reference Cahnos (SKU) | F7952PW |
| Product type | Pair of standard paediatric multifunction defibrillation electrodes, with external cable (lead-out) |
| Indicated population | Children under 8 years or weighing less than 25 kg |
| Cable | 100 cm, external (lead-out / pre-connected) |
| Radiotransparency | No (standard version) |
| Functions | Defibrillation, synchronised cardioversion and ECG monitoring (max. 24 h). Not suitable for pacing (pace) |
| Supported shocks | Up to 50 shocks at 100 J (paediatric series) |
| Total area (per electrode) | 75 cm² |
| Active area (per electrode) | 40 cm² |
| Conductive gel | Low-impedance conductive adhesive gel |
| Conductive material | Metal foil |
| Backing material | 1 mm thick medical foam |
| Connection type | Safety shockproof connector / QUIK-COMBO clip type (depending on equipment model) |
| Presentation | 1 pair (2 electrodes); 5 pairs per box (paediatric series) |
| Use | Single-use, disposable; latex-free and phthalate-free |
| Shelf life | Approx. 30 months from date of manufacture (expiry printed on packaging) |
| Conditions of use and storage | Operating temperature from 0 °C to +50 °C; store in a dry place between +5 °C and +35 °C |
| Classification | Medical device class IIb (Directive 93/42/EEC and amendments; EU Regulation 2017/745) |
| Regulatory standard | ANSI/AAMI DF-80; IEC/EN 60601-1, 60601-2-4; ISO 10993-10 |
3.1 Compatibility with defibrillators
| Equipment manufacturer | Compatible models |
|---|---|
| Physio-Control / Medtronic | LIFEPAK 9 (with QUIK-COMBO adapter), LIFEPAK 10, LIFEPAK 12, LIFEPAK 15, LIFEPAK 20 |
| Mindray | BeneHeart D1, BeneHeart D3, BeneHeart D6 |
| Bexen Cardio | REANIBEX-200, REANIBEX-300, REANIBEX-500, REANIBEX-700, REANIBEX-800 |
| Cardioline | ELIFE 700 |
Note: the same product family also documents compatibility with other equipment in the range (e.g. Comen S8) according to the manufacturer’s technical data sheet. Always verify correct insertion of the connector into the equipment socket before clinical use.
4. Equivalences and Cross-References
This electrode (SKU F7952PW) is a consumable compatible with the cited equipment, not an original factory spare part. The correspondence with the equipment manufacturers’ references is as follows; any comparison between brands must be understood as comparable at category level, not 1:1 interchangeable.
| Manufacturer / Brand | Family / Model | Reference code / SKU | Equivalence level | Technical differences / Remarks |
|---|---|---|---|---|
| Stryker / Physio-Control | Radiolucent EDGE System paediatric electrodes with QUIK-COMBO connector (original equipment) | 11996-000090 | Comparable by category (not interchangeable 1:1) | Factory original for LIFEPAK 12, 15 and 20 defibrillators and monitors; QUIK-COMBO connector, paediatric multifunction for defibrillation, pacing and ECG, paediatric use 0–15 kg. It differs from this reference in its radiotransparent (RTS) construction and in that it does allow pacing; this data sheet is the standard non-radiotransparent variant and not suitable for pacing. Code confirmed in the official Stryker store, authorised distributors and GUDID (UDI 00721902195037). Not suitable for equipment of other brands. |
| Mindray | MR61 – Paediatric multifunction electrode (original spare part for BeneHeart series) | 115-040518-00 | Comparable by category (not interchangeable 1:1) | Mindray original for BeneHeart D1, D2, D3, D5, D6 and C series defibrillators; paediatric population under 25 kg; Mindray proprietary connector and algorithm, not compatible with the QUIK-COMBO socket of the LIFEPAK. Code confirmed in the official Mindray pads guide and in manufacturer reference sources. |
For the Bexen Cardio REANIBEX and Cardioline ELIFE 700 equipment, no verifiable original paediatric electrode code has been found in public sources; confirmation with the manufacturer is pending. There is no equivalence table validated by Physio-Control/Stryker, Mindray or Bexen Cardio authorising the replacement of original references 11996-000090 or 115-040518-00 with third-party pads: the connector, analysis algorithm and energy management are proprietary. Any replacement must be validated against the defibrillator’s technical data sheet and IFU. Consult the Cahnos technical department before use in tenders.
5. Selection tips and variants of interest
Confirm the model and connector of the unit: this pair corresponds to Physio-Control LIFEPAK 9 (with QUIK-COMBO adapter), 10, 12, 15 and 20 defibrillators and monitors, and to Mindray BeneHeart, Bexen Cardio REANIBEX and Cardioline ELIFE 700 equipment. Verify the reference and expiry date printed on the packaging before replacement; do not use an expired electrode or reuse a pair after an intervention.
Choose the variant according to your workflow: the external cable (lead-out) of this model allows the unit to be left pre-connected and ready in seconds; if you prefer internal cable, the F7952P reference is available (same standard paediatric series). For an artefact-free imaging requirement (X-ray or CT during care), consider the radiotransparent series F7752P (internal cable) or F7752PW (external cable). For AED equipment with automatic energy reduction, references F7952PL and F7952WPL (200 J → 50 J) are the indicated ones.
Maintenance and stocking: manage stock rotation by expiry (approx. 30 months) and always keep a spare pair next to the unit in centres with a paediatric population. For volume orders or tenders, contact the Cahnos technical department indicating the exact model and serial number of the defibrillator.







Reviews