Description
1. Introduction
Ensure safe and effective cardiac care for the smallest patients with this pair of standard paediatric defibrillator electrodes with external cable (reference Cahnos F7961PW), a replacement consumable compatible with Nihon Kohden defibrillators from the Cardiolife TEC-5500, TEC-5600 and TEC-7600 series. It is the ideal solution for emergency departments, paediatric intensive care units (PICU), ambulances and any clinical setting that cares for children: a single pair of disposable pads keeps the equipment ready to act in the event of a cardiorespiratory arrest. Available with immediate dispatch from Cahnos and specialised technical support for compatibility verification by model and connector.
2. Technical Explanation
Disposable multifunction electrodes are the interface between the defibrillator and the patient: through them the device analyses the rhythm, delivers the transthoracic shock and monitors the response. This paediatric model allows external transthoracic defibrillation, synchronised cardioversion, non-invasive transcutaneous cardiac pacing and ECG monitoring (up to 24 h), adapted to paediatric physiology. The configuration with external cable (pre-connected electrode) keeps the connector accessible and allows the device to be ready for immediate connection, a decisive advantage in paediatric resuscitation.
In patients under 8 years of age or weighing less than 25 kg, it is essential to attenuate the delivered energy to avoid myocardial injury; therefore, this electrode incorporates a reduced active surface of 40 cm² per electrode and must be used with the device’s paediatric mode switching or with the reduced energy configuration indicated in the defibrillator manual. The low-impedance conductive adhesive gel ensures stable adhesion and a reliable signal, while the metal foil and 1 mm medical foam (FOAM) guarantee a uniform distribution of current. They are single-use: do not reuse under any circumstances and replace after each intervention or upon reaching the expiry date printed on the packaging.
3. Technical specifications and compatibility
3.1 Technical specifications
| Parameter | Specification |
| Patient type | Paediatric (<8 years / <25 kg) |
| Cable configuration | External cable (pre-connected), 100 cm |
| Total surface per electrode | 75 cm² |
| Active surface per electrode | 40 cm² |
| Adhesive ring width | 1 cm |
| Supported shocks | Up to 50 shocks at 100 J |
| Conductive gel | Low-impedance conductive adhesive gel |
| Conductive material | Metal foil |
| Backing material | Medical foam (FOAM) 1 mm thick |
| Connection type | Safety anti-shock connector |
| ECG monitoring | Up to 24 hours |
| Regulatory classification | Medical device class IIb (Directive 93/42/EEC; EU Regulation 2017/745) |
| Reference standards | ANSI/AAMI DF-80; IEC/EN 60601-1, IEC/EN 60601-2-4; ISO 10993 |
| Primary packaging | 1 pair in heat-sealed polyethylene and aluminium bag |
| Sales format | 5 pairs per box (paediatric series) |
| Shelf life | 30 months from the date of production (see expiry date on the packaging) |
| Storage | Dry place, between +5 °C and +35 °C |
| Composition | Latex-free and phthalate-free · Single-use |
3.2 Compatibility with defibrillators
| Equipment brand | Compatible models | Cable / required accessories |
| Nihon Kohden | Cardiolife TEC-5500 Series, TEC-5600 Series, TEC-7600 Series | Cable JC-755V (TEC-5500, TEC-7600) · Cable K342A / JC-855V (TEC-5600) |
Note: always verify the fit of the connector and the cable of your device before use; the TEC-5600 series use the K342A / JC-855V cable, while TEC-5500 and TEC-7600 use JC-755V.
4. Equivalences and Cross-References
The reference F7961PW is a consumable compatible with the aforementioned Nihon Kohden Cardiolife TEC defibrillators, not an original spare part certified by the device manufacturer: the connector and energy management are proprietary to Nihon Kohden. After verifying official manufacturer catalogues and regulatory databases, the following cross-references with codes confirmed in public sources have been identified. In no case is there an official equivalence table validated by the manufacturer for this reference.
| Manufacturer / Brand | Family / Model | Reference code / SKU | Equivalence level | Technical differences / Remarks |
| Nihon Kohden | Disposable adult/child/infant pad (original equipment) | P-740K | Comparable by category (not interchangeable 1:1) | Factory original with 1.5 m cable; reference confirmed in the manufacturer’s official IFU and authorised distributors. Covers AED-2100K/2150K/2151K/2152K, AED-3100 and TEC-5500/5600/7600/7700/8300 series. Connector, gel and labelling proprietary to Nihon Kohden; verify paediatric size and connector fit before replacement. |
| Nissha Medical Technologies (HeartSync) | Defibrillation pads compatible with Nihon Kohden TEC-5500, TEC-5600, TEC-7600, TEC-7700 and TEC-8300 | T100AC-NIHONKOHDEN (ref. 6600265H) | Comparable by category (not interchangeable 1:1) | Code confirmed in the manufacturer’s official catalogue: connector comparable to the original Nihon Kohden and cable included in the packaging. The catalogued version is adult (total surface 371 cm²); verify paediatric size, connector and cable (JC-755V / K342A-JC-855V depending on series) before replacement. |
In public sources, no verifiable original paediatric electrode from Nihon Kohden specific to the TEC-5500/5600/7600 series with JC-755V or K342A cable has been found; the manufacturer publishes electrodes with paediatric capability aimed at its AED and TEC-8300 families (P-711, P-713, P-740K). There is no equivalence table validated by the manufacturer that authorises the replacement of original electrodes with third-party pads: the connector and energy management of each defibrillator are proprietary. Any replacement must be validated against the technical data sheet and instructions for use of the defibrillator; consult the technical department of Cahnos before use in tenders.
5. Selection tips and variants of interest
- Select according to the patient: this paediatric reference is indicated for children under 8 years of age or weighing less than 25 kg. For patients over 25 kg, use the adult version of the same Nihon Kohden connector family (F7961W, external cable JC-765V).
- Verify the device cable Nihon Kohden: the TEC-5600 series models use the K342A / JC-855V cable; the TEC-5500 and TEC-7600 series use JC-755V. Always check the connector fit before clinical use.
- Variants of the same Cahnos family: paediatric version with internal cable for Nihon Kohden (F7961P, for TEC-2200, TEC-2300 and TEC-5500 with JC-765V), standard multi-brand version with external cable (F7955PW) or with internal cable (F7955P) which also covers these TEC series, and radiolucent version with internal cable (F7755P) if X-rays or CT scans are anticipated during care.
- Paediatric mode: ensure the device switches to reduced paediatric energy (up to 100 J) according to the defibrillator manual; on manually adjustable models always select it before the shock.
- Storage: keep in a dry place between +5 °C and +35 °C and observe the expiry date stated on the packaging (30 months from production); the conductive gel may dry out and reduce its conductivity after the expiry date.
- Single-use: do not reuse; one pair per patient and clinical event. It is recommended to maintain minimum stock per unit and rotate stock (FIFO).
Immediate availability from Cahnos with technical support for compatibility verification by model and connector before placing volume orders or submitting tenders.






Reviews