Description
1. Commercial Introduction
Keep your defibrillator or defibrillator monitor from the Schiller family always ready to treat even the smallest patients. This standard paediatric electrode pair with external cable (reference Cahnos F7967PW) is the replacement consumable indicated for paediatric defibrillation with the FRED EasyPort, FRED PA-1, Argus Pro LifeCare, Argus Pro LifeCare 2 and Cardiovit AT-101 Ea equipment. It is sold as a pair (2 units) in an individual heat-sealed package and is ready to be integrated into cardiopulmonary resuscitation protocols in emergency services, paediatric intensive care units (PICU), ambulances, health centres, schools and heart-safe spaces equipped with Schiller.
2. Technical explanation of the equipment and procedure
Multifunction defibrillation electrodes are the interface element between the defibrillator and the patient: through them the device analyses the cardiac rhythm, delivers the transthoracic shock and, in many models, allows emergency ECG monitoring. In paediatric patients (under 8 years of age or weighing less than 25 kg) it is essential to attenuate the delivered energy to avoid myocardial injury; therefore this electrode features a reduced active surface of 40 cm² per pad and must be used with the device’s paediatric mode switch or with the reduced energy setting indicated in the defibrillator manual.
The recommended placement is anterolateral (anteroapical position), so that both pads do not come into contact with each other on the child’s chest. The low-impedance conductive adhesive gel ensures stable adhesion and a reliable signal, while the metal foil and 1 mm thick medical foam guarantee a uniform distribution of current. The external cable configuration leaves the connector accessible and allows the device to be kept ready for immediate connection, a decisive advantage in paediatric resuscitation. They are single-use and disposable: do not reuse under any circumstances and replace after each intervention or upon reaching the expiry date printed on the packaging.
3. Technical Specifications
| Parameter | Specification |
|---|---|
| Reference Cahnos (SKU) | F7967PW |
| Compatible equipment brand | Schiller – FRED EasyPort, FRED PA-1, Argus Pro LifeCare, Argus Pro LifeCare 2, Cardiovit AT-101 Ea |
| Product type | Standard paediatric electrode pair with external cable, multifunction |
| Target population | Paediatric (< 8 years / < 25 kg) |
| Total surface (per electrode) | 75 cm² |
| Active surface (per electrode) | 40 cm² |
| Adhesive ring width | 1 cm |
| Supported shocks | Up to 50 shocks (reduced paediatric energy 50–100 J; verify according to the device indications) |
| Backing material | Medical foam (FOAM) 1 mm thick |
| Conductive material | Metal foil |
| Conductive gel | Low-impedance conductive adhesive gel |
| Connection type | Safety anti-shock connector / clip, with external cable |
| Maximum ECG monitoring time | Up to 24 h |
| Operating temperature | Between 0 °C and +50 °C |
| Use | Single-use, disposable, non-sterile |
| Latex / phthalates | Latex-free and phthalate-free |
| Regulatory classification | Medical device class IIb (Directive 93/42/EEC) |
| Reference standards | ANSI/AAMI DF-80; IEC/EN 60601-1, IEC/EN 60601-2-4; ISO 10993-10 |
| Primary packaging | 1 pair in heat-sealed polyethylene and aluminium bag, with laser pre-cut and opening notch |
| Presentation | Price per pair of electrodes (2 units); 5 pairs per box in the paediatric series |
| Shelf life | Approx. 30 months from the date of production; expiry indicated on the packaging |
| Storage | Dry place, between +5 °C and +35 °C |
Note: the active surface (40 cm²), total area (75 cm²), shocks (50–100 J) and adhesive ring (1 cm) values correspond to the standard paediatric electrode series with external cable in this product family; confirm them with the technical data sheet of the supplied batch and the device manual before use in tendering.
4. Equivalences and Cross-References
This electrode (SKU F7967PW) belongs to the family of standard paediatric consumables with external cable for defibrillation. The original replacement from the equipment manufacturer, Schiller, is identified by code 2.155067 (paediatric electrodes for FRED easy G2, FRED PA-1, FRED easyport/easyport plus, as well as Defigard Touch 7, Physiogard Touch 7 and Defigard HD-7), consistently verified at authorised distributors of original Schiller spare parts. Equivalences with other brands are comparable only at category level: the connector, analysis algorithm and energy management are proprietary to each device, so they are not interchangeable 1:1.
| Manufacturer / Brand | Family / Model | Reference code / SKU | Equivalence level | Technical differences / Remarks |
|---|---|---|---|---|
| Schiller (original equipment replacement) | Paediatric electrodes FRED PA-1 / FRED EasyPort (child pads) | 2.155067 | Comparable by category (not interchangeable 1:1) | Original factory replacement for the same equipment family (FRED easy G2, FRED PA-1, FRED easyport/easyport plus, Argus Pro LifeCare, Defigard Touch 7, Physiogard Touch 7, Defigard HD-7 and Cardiovit AT-101 easy); pre-connected, with integrated cable instead of plug-in external cable. Code confirmed at authorised distributors of original Schiller spare parts; there is no manufacturer validation authorising the replacement of the original with the compatible Cahnos replacement: the substitution must be validated against the equipment’s technical data sheet and IFU. |
| ZOLL Medical | Pedi·padz® II (original paediatric electrodes) | 8900-0810 / 8900-0810-01 | Comparable by category (not interchangeable 1:1) | Factory original for AED Plus and AED Pro (pre-connected, without external cable); conductive hydrogel and paediatric levels of 50/70/85 J. Code confirmed on the manufacturer’s website and GUDID (UDI 00847946001438). Not suitable for Schiller equipment due to proprietary connector. |
Within the Cahnos catalogue there are other references in the same family (standard paediatric with external cable) intended for equipment from other manufacturers, so they are not interchangeable with each other as the connector and algorithm differ: F7950PW (Philips), F7961PW (Nihon Kohden), F7964PW (Weinmann MEDUCORE), F7969PW (Mediana HeartOn), F7984PW (Instramed), F7986PW (Mediana D500) and F7987PW (ZOLL). The adult version of this same Schiller equipment is the F7967W, and the paediatric variant with internal cable for the same family of equipment is the F7967P.
There is no equivalence table validated by the manufacturer; consult the technical department before use in tenders.
5. Selection tips and variants of interest
- Confirm the equipment model: verify that your defibrillator is a Schiller FRED EasyPort, FRED PA-1, Argus Pro LifeCare, Argus Pro LifeCare 2 or Cardiovit AT-101 Ea and that the external cable connector fits correctly into the equipment socket before clinical use.
- Do not confuse the paediatric and adult versions: this pair is for patients under 8 years of age or under 25 kg; for adult patients in the same Schiller family use reference F7967W (external cable) or F7967 (internal cable).
- Paediatric mode: ensure the equipment switches to paediatric mode or energy (reduced energy setting 50–100 J) according to the defibrillator manual; on manually adjustable models, always select it before the shock.
- Storage and expiry: always check the date printed on the packaging and store in a dry place between +5 °C and +35 °C; expired electrodes or those with damaged packaging must not be used (the gel may lose conductivity).
- Plan your replacement: as this is a single-use consumable, maintain a minimum stock per defibrillation unit and rotate stock according to the expiry date. For volume orders or tenders, contact the Cahnos technical department stating the exact model and serial number of your equipment.