Description
1. Commercial Introduction
Keep your Physio-Control defibrillator or monitor always operational with this pair of standard adult defibrillator electrodes with internal cable (reference Cahnos F7971), a highly reliable replacement consumable compatible with LIFEPAK family equipment. Designed for emergency services, ICU, ambulances, operating theatres, hospitalisation and any cardio-protected space, it simplifies consumable management with an internal cable configuration that keeps the connector protected in the packaging until the moment of use: simply open the sachet, place the pads following the anatomical diagrams and connect. Available with immediate dispatch from Cahnos and specialised technical support for compatibility verification by model and connector before placing the order.
2. Technical explanation of the equipment and procedure
Multifunction defibrillation electrodes are the interface between the defibrillator and the patient: through them the device analyses the cardiac rhythm, delivers the transthoracic shock, enables synchronized cardioversion, non-invasive transcutaneous pacing (external pacemaker) and emergency ECG monitoring.
This adult model (patients over 8 years old or weighing more than 25 kg) is constructed with low-impedance conductive adhesive gel, which ensures homogeneous contact with the skin and optimal energy transmission; the internal metal foil distributes the current uniformly and the 1 mm medical foam backing adapts to the contours of the chest, minimising the risk of burns. The configuration with integrated internal cable (leads-in) and Snap-type connector provides a clean, stable and direct connection to the device socket, without loose cables during resuscitation. The recommended placement is anterolateral (sternum-apex), according to the anatomical diagrams on the packaging. They are single-use: they must be replaced after each intervention or upon reaching the expiry date printed on the packaging; do not reuse under any circumstances.
3. Technical Specifications
Main features of the standard adult electrode series with internal cable to which this reference belongs:
| Parameter | Specification |
|---|---|
| Reference Cahnos (SKU) | F7971 |
| Product type | Pair of standard multifunction adult defibrillator electrodes with internal cable (leads-in) |
| Patient type | Adult (older than 8 years / more than 25 kg) |
| Connection | Internal cable with Snap connector (Physio-Control / QUIK-COMBO connection platform) |
| Functions | Defibrillation, synchronized cardioversion, non-invasive transcutaneous pacing and ECG monitoring |
| Total surface (per electrode) | 148 cm² (standard adult series) |
| Active surface (per electrode) | 95 cm² (standard adult series) |
| Adhesive ring width | 1 cm |
| Supported shocks | From 50 to 360 J (up to 50 shocks) |
| ECG monitoring | Max. 24 h |
| Conductive gel | Low-impedance conductive adhesive gel |
| Conductive material | Metal foil |
| Backing material | Medical foam (FOAM) 1 mm thick |
| Use | Single-use, disposable, non-sterile; latex-free and phthalate-free |
| Presentation | 1 pair in heat-sealed polyethylene and aluminium bag; 10 pairs per box (adult series) |
| Shelf life | Approx. 30 months from the date of manufacture (expiry date indicated on the packaging) |
| Storage | Dry place, between +5 °C and +35 °C; operating from 0 °C to +50 °C |
| Regulatory classification | Class IIb medical device (Directive 93/42/EEC; EU Regulation 2017/745) |
| Reference standards | ANSI/AAMI DF-80; IEC/EN 60601-1, IEC/EN 60601-2-4; ISO 10993-10 |
Note: the active surface values (95 cm²), total area (148 cm²), adhesive ring (1 cm) and shocks (50 to 360 J) correspond to the standard adult electrode series of this product family; confirm with the technical data sheet of the supplied batch and the device manual before use in tendering.
3.1 Compatibility with defibrillators
| Equipment brand | Compatible models |
|---|---|
| Physio-Control | LIFEPAK 9 (with QUIK-COMBO adapter), LIFEPAK 10, LIFEPAK 12, LIFEPAK 15, LIFEPAK 20, LIFEPAK 20E, LIFEPAK 500 AED and LIFEPAK 1000 AED |
Note: compatibility refers to the family’s connector platform. Always verify the correct fit of the Snap connector and the cable of your device before clinical use.
4. Equivalences and Cross-References
Reference F7971 is a consumable compatible with the cited Physio-Control defibrillators, not an original spare part certified by the device manufacturer: the connector, analysis algorithm and energy management are proprietary. After verifying official manufacturer catalogues, authorised technical distribution and regulatory databases (GUDID/FDA), the following cross-references with confirmed codes in public sources are identified. Any comparison between brands should be understood as comparable at category level, not interchangeable 1:1.
| Manufacturer / Brand | Family / Model | Reference code / SKU | Equivalence level | Technical differences / Remarks |
|---|---|---|---|---|
| Stryker / Physio-Control | EDGE System Electrodes with QUIK-COMBO Connector (original for the device, internal cable / not pre-connected) | 11996-000091 | Comparable by category (not interchangeable 1:1) | Factory original for LIFEPAK 12, 15 and 20; cable inside the packaging (not pre-connected), same internal cable configuration and same clinical function (defibrillation, pacing, cardioversion and ECG). Code verified in GUDID/FDA (UDI 00883873784945). Verify connector, gel and labelling before replacement. |
| Stryker / Physio-Control | EDGE System Electrodes with QUIK-COMBO Connector & REDI-PAK Preconnect System (original equipment) | 11996-000017 | Comparable by category (not interchangeable 1:1) | Factory-original pre-connected for LIFEPAK 12, 15, 20, 20E, 500 AED and 1000 AED; 42″ cable (≈107 cm) with REDI-PAK pre-connection system. Code verified in GUDID/FDA (UDI 00883873784969) and in the official Stryker store. Same connector platform; pre-connected configuration different from this reference. |
| Leonhard Lang | Skintact® Easibeat – adult multifunction electrode, Physio-Control connection platform | DF20N | Comparable by category (not interchangeable 1:1) | Multifunction (defibrillation, pacing, cardioversion and monitoring) for LIFEPAK 9, 10, 12, 15, 20, 500 and 1000; 1 pair per pouch and 10 pouches per box. Code confirmed in authorised technical distribution catalogue (CNA Medical) and in the official documentation of Leonhard Lang. Connector, gel and materials are proprietary; verify fit before replacement. |
| Bexen Cardio | Original adult electrodes for Reanibex 200 / 300 / 500 / 700 / 800 (not pre-connected) | TKL0311A | Comparable by category (not interchangeable 1:1) | Factory-original replacement for the REANIBEX series, same connection platform type QUIK-COMBO; non-pre-connected presentation (internal cable). Code confirmed in authorised distribution. Verify connector fit before replacement. |
| Mindray | Adult Multifunction Electrode Pads (Basic) for BeneHeart / MR60 series | 115-040517-00 | Comparable by category (not interchangeable 1:1) | Original Mindray for BeneHeart D1, D3, D6 and MR60 series defibrillators; adult >25 kg, up to 50 shocks at 360 J. Code confirmed in the official Mindray electrode guide and in authorised distribution. Proprietary Mindray connector and algorithm; verify compatibility with the equipment connector. |
There is no equivalence table validated by the manufacturer that authorises the replacement of the original electrodes of these devices with third-party pads: the connector, the analysis algorithm and the energy management are proprietary to each defibrillator. Any replacement must be validated against the technical data sheet and IFU of the device; consult with the technical department of Cahnos before use in tenders.
5. Selection tips and variants of interest
Confirm the model and connector of your device: this pair corresponds to Physio-Control LIFEPAK 9 (with QUIK-COMBO adapter), 10, 12, 15, 20, 20E, 500 AED and 1000 AED defibrillators and monitors. Always verify the Snap connector fit and the expiry date printed on the packaging before replacement; do not use an expired electrode or reuse a pair after an intervention.
Choose the variant according to your workflow: the internal cable of this model provides a clean and robust connection during resuscitation; if you prefer to leave the equipment pre-connected and ready in seconds, you have the external cable version of the same platform (F7952W). For image-guided procedures (X-ray, fluoroscopy or CT scan), consider the radiolucent versions F7752 (internal cable) or F7752W (external cable). For paediatric patients (under 8 years or 25 kg), use the paediatric references F7952P and F7952PW (or F7752P / F7752PW in radiolucent version).
Storage and expiry: store in a dry place between +5 °C and +35 °C and respect the expiry date on the packaging; the conductive gel may dry out and reduce its conductivity after the expiry date. As it is a single-use consumable, maintain a minimum stock per defibrillation unit and rotate stock (FIFO).
For volume orders or tenders, contact the technical-commercial department of Cahnos indicating the exact model and serial number of your equipment; we will confirm compatibility before placing the order.