Description
1. Commercial Introduction
Ensure the operability of your defibrillation point with this pair of preconnected adult/paediatric multifunction electrodes, designed as a compatible replacement for the automatic external defibrillators (AED) Mediana HeartOn AED A15 and Tecno-Gaz Tecno-Heart Plus. Its preconnected configuration with external cable allows the device to remain always ready: without opening the sealed pouch, the connector remains accessible for immediate connection during care for cardiorespiratory arrest. An essential solution for emergency services, companies with AEDs, schools, sports centres, industrial estates and any space with a defibrillator that requires fast, reliable and cost-optimised replacement.
2. Technical explanation of the equipment and procedure
The disposable multifunction electrodes allow, with a single pair of adhesive pads: transthoracic external defibrillation, synchronised cardioversion, transcutaneous cardiac pacing and ECG monitoring. Their low-impedance conductive adhesive gel ensures stable adhesion and a reliable signal, while the 1 mm medical foam support and the internal metal foil guarantee uniform current distribution with minimal risk of burns.
As it is an adult/paediatric model, the same pair of electrodes serves patients over 10 kg: both the Mediana HeartOn AED A15 and the Tecno-Gaz Tecno-Heart Plus have energy switching via a physical ‘adult/paediatric’ switch, which automatically reduces the shock in paediatric mode. The anti-shock safety connector is inserted into the device socket following the anatomical diagrams on the packaging. It is a single-use product: it must be replaced after each use or upon reaching the expiry date printed on the packaging.
3. Technical Specifications
| Field | Specification |
| Product type | Disposable adult/paediatric multifunction electrode, preconnected (lead-out) |
| Compatibility | Mediana HeartOn AED A15 and Tecno-Gaz Tecno-Heart Plus |
| Adult mode | Application for adult patients with optimised energy |
| Paediatric mode | Reduced energy setting for paediatric patients |
| Population | Adults and paediatric > 10 kg (> 12 months) |
| Functions | Defibrillation, synchronized cardioversion, transcutaneous pacing and ECG monitoring (max. 24 h) |
| Backing material | Medical foam (FOAM) 1 mm thick |
| Conductive material | Metal foil |
| Conductive gel | Low-impedance conductive adhesive gel |
| Adhesive ring width | 1 cm |
| Total area (per electrode) | 148 cm² |
| Active area (per electrode) | 95 cm² |
| Supported shocks | 50 shocks at maximum power of 360 J |
| Cable length | 180 cm, connector accessible from outside the pouch (preconnected) |
| Connection type | Anti-shock safety connector |
| Presentation | 1 pair in heat-sealed polyethylene and aluminium pouch (10 pairs per box) |
| Classification | Medical device class IIb (Directive 93/42/EEC) |
| Use | Single-use, disposable, non-sterile |
| Exempt materials | Latex-free and phthalate-free |
| Shelf life | Approx. 30 months from the date of manufacture; expiry date indicated on the packaging |
| Operation | 0 °C to +50 °C |
| Storage | Dry place, between +5 °C and +35 °C |
4. Equivalences and Cross-References
This product is a compatible consumable for Mediana and Tecno-Gaz defibrillators, not an original spare part from these manufacturers: the connector and monitoring system are proprietary to each device, so replacing the original pad with a compatible accessory must always be validated against the technical data sheet and IFU of the defibrillator. The original reference numbers of the device manufacturers that this electrode functionally covers are:
| Manufacturer / Brand | Family / Model | Reference code / SKU | Equivalence level | Technical differences / Remarks |
| Mediana | Disposable adult/paediatric electrode for HeartOn AED A15 (original equipment) | A0649-0 (current ref.) / A0401-0 (previous ref.) | Comparable by category (not interchangeable 1:1) | Factory original: connector, gel and labelling specific to Mediana; same dual adult/paediatric use concept with switching via the AED switch (reference numbers confirmed by authorised distributors of the manufacturer) |
| Tecno-Gaz | Disposable adult/paediatric electrode for Tecno-Heart Plus (original equipment) | 1Z12A0003 | Comparable by category (not interchangeable 1:1) | Factory original (official Tecno-Gaz catalogue); Tecno-Gaz S.p.A. expressly advises against the use of third-party electrodes in Tecno-Heart Plus and Tecno-Heart Smart (official note 0Z12H0052, 29/09/2023) |
There is no equivalence table validated by Mediana or Tecno-Gaz that authorises replacing reference numbers A0649-0/A0401-0 or 1Z12A0003 with third-party consumables: any comparison should be understood as comparable at category level, not 1:1 interchangeable, and must be confirmed with the technical department before use in tenders. In particular, Tecno-Gaz S.p.A. has published an official communication (0Z12H0052, 29/09/2023) in which it warns that the use of unauthorised pads may cause incorrect AED operation and erroneous measurements. The correspondence of this SKU (F7986UW) with the official Mediana and Tecno-Gaz constructions is not published by these device manufacturers.
5. Selection tips and variants of interest
If your AED is a Mediana HeartOn AED A15 or a Tecno-Gaz Tecno-Heart Plus and serves the general population (adults and children), this pair of preconnected adult/paediatric electrodes is the most versatile option: a single pair of pads covers both populations and paediatric energy is selected with the device switch, without the need for an additional set. Always consult the IFU of your defibrillator and, if you have any doubt about compatibility, contact the technical department of Cahnos.
For AED fleets or high-volume centres, manage stock rotation by expiry (approx. 30 months) and always replace the electrode after use or expiry. As variants of interest in the catalogue, if you have Nihon Kohden equipment from the Cardiolife family, consider the pair compatible with AED 2100, 2150K, 2151K and 2152K (SKU F7948UW) or the preconnected lead-out variant for the same Nihon Kohden platform (SKU F7989UW). Consult the technical department of Cahnos to confirm exact compatibility with your defibrillator model before placing large-volume orders or tenders.






Reviews
There are no reviews yet.