Description
1. Commercial Introduction
The COVID-19 IgM/IgG Antibody Rapid Test is a rapid serology diagnostic solution designed for the qualitative and differential detection of IgM and IgG antibodies against SARS-CoV-2 in a single sample of whole blood (capillary or venous), serum or plasma. In a single dual-band cassette and in just 10–15 minutes, the healthcare professional obtains a reliable visual result, without the need for laboratory equipment. It is the ideal tool for clinics, hospitals, primary care consultations, occupational health units, emergency departments, care homes and serological screening campaigns that need to assess the immune response to the virus, facilitating clinical decision-making at the point of care.
2. Technical explanation of the procedure
The test is a lateral flow chromatographic immunoassay with colloidal gold. Anti-human IgM antibodies (IgM test line), anti-human IgG antibodies (IgG test line) and a control (C line) are immobilised on the nitrocellulose strip. The conjugate pad contains recombinant SARS-CoV-2 antigens conjugated with colloidal gold. When the sample is added followed by the assay buffer, any IgM and/or IgG antibodies present bind to the conjugates and the complex migrates by capillary action; if the complex is trapped at the corresponding line, a burgundy-coloured band forms, confirming a reactive result. The control line must always appear, validating the correct functioning of the device. It is a single-use test, exclusively for professional in vitro diagnostics.
Summary procedure:
- Massage the fingertip, disinfect with the alcohol swab and perform the puncture with the included lancet.
- Clean the first drop of blood with dry cotton and aspirate a small amount with the micropipette.
- Deposit the sample in the cassette well and add two drops of the reagent (buffer).
- Leave the plate horizontal for 10 minutes and proceed to read.
Interpretation of results:
- Negative: only the control line (C) appears.
- IgM positive: appearance of the C band and the IgM band (recent or active infection).
- IgG positive: appearance of the C band and the IgG band (previous exposure or convalescent phase).
- IgM and IgG positive: both test bands appear together with the control.
- Invalid: absence of the control line; repeat with a new cassette.
3. Technical Specifications
| Parameter | Specification |
|---|---|
| Test type | Lateral flow immunochromatographic assay (colloidal gold method) |
| Analytes detected | Anti-SARS-CoV-2 IgM and IgG antibodies (qualitative and differential detection) |
| Sample type | Whole blood (capillary or venous), serum or plasma |
| Format | Double band cassette (IgM, IgG and control C lines) |
| Time to result | 10–15 minutes (visual reading) |
| Internal control | Control line (C) on each device |
| Storage | 2–30 °C; do not freeze; keep the cassette in the sealed pouch until use |
| Use | In vitro diagnostic for professional use |
| Presentation | Box of 20 units (minimum order: full box) |
Box contents (20 tests):
| Material | Quantity |
|---|---|
| Sealed pouches (test cassette, dropper and buffer) | 20 |
| Lancets (for fingerstick whole blood) | 20 |
| Alcohol swabs for disinfection | 20 |
| Package insert / instructions for use | 1 |
Clinical performance:
| Parameter | Value |
|---|---|
| Sensitivity (according to technical data sheet) | 98.6 % |
| Specificity (according to technical data sheet) | 99.1 % |
| IgM performance vs. EIA (IFU) | Relative sensitivity 95.7 % · Relative specificity 97.3 % · Overall agreement 96.8 % |
| IgG performance vs. EIA (IFU) | Relative sensitivity 91.8 % · Relative specificity 96.4 % · Overall agreement 95.0 % |
Note: positive or equivocal results must be confirmed with alternative assay methods (RT-PCR, ELISA or quantitative serology) and clinical findings, as indicated in the manufacturer’s package insert.
4. Equivalences and Cross-References
This antibody test falls within the category of rapid serology tests (IgM/IgG) on whole blood, serum or plasma by lateral flow immunochromatography. The references from other manufacturers listed below are functional category comparisons verified against catalogues, IFUs and regulatory databases of the respective manufacturers; they are not interchangeable 1:1 and must be validated property by property (format, analytes, read time, presentation and performance) before any substitution.
| Manufacturer / Brand | Family / Model | Code / SKU | Equivalence Level | Technical Differences / Remarks |
|---|---|---|---|---|
| SD Biosensor | STANDARD Q COVID-19 IgM/IgG Duo | Q-NCOV-01D (presentation 25 tests, cat. RK054-25) | Comparable by category (not interchangeable 1:1) | Same analyte (anti-SARS-CoV-2 IgM/IgG) and same lateral flow technique on blood/serum/plasma (sample of 10 µL, 10–15 min, stability at 2–30 °C); presentation of 25 tests per kit. Confirm cassette format and volume per pack. |
| Healgen Scientific | COVID-19 IgG/IgM Rapid Test Cassette (Whole Blood/Serum/Plasma) | GCCOV-402A | Comparable by category (not interchangeable 1:1) | Immunochromatography cassette for qualitative detection of IgM/IgG in venous whole blood, serum or plasma (FDA EUA, professional use); usual presentation of 25 tests per box. Read at the time indicated in its IFU. |
| CTK Biotech | OnSite COVID-19 IgG/IgM Rapid Test | R0180C | Comparable by category (not interchangeable 1:1) | Single-use lateral flow immunoassay for detection and differentiation of anti-SARS-CoV-2 IgG/IgM in serum, plasma or whole blood (EDTA, heparin or citrate); readout in 15 minutes, sensitivity 97.1% according to published data. Verify the presentation reference (number of tests per kit) in the manufacturer’s catalogue. |
There is no equivalence table validated by the manufacturer for this reference; consult the technical department before using in a tender. The equivalences indicated are functional category comparisons, not interchangeable 1:1: before substituting, verify in each technical data sheet the device format, the analytes detected (IgM/IgG), the authorised sample type, the read time, the scope of use (professional) and the performance data against the reference method (EIA or PCR).
This antibody test is a serology test and should not be confused with the antigen tests in the catalogue (which detect active infection in a nasal sample); both approaches are complementary depending on the phase of infection.
5. Selection tips and variants of interest
- Phase of infection: IgM antibodies appear in the early phase of the immune response and IgG in the convalescent phase or prior exposure. This test is suitable for assessing serological response; for diagnosis of acute infection in the first few days, combine with PCR or with antigen tests from the catalogue according to protocol.
- Sample type: accepts whole blood (finger prick), serum or plasma; lancets and alcohol pads are included for capillary collection.
- Care volume: the presentation of 20 units per box optimises the cost per test in centres with medium demand; consult our sales team for special conditions for volume orders.
- Serology panel: to complete the rapid serology portfolio on blood samples from a single supplier, the catalogue includes antibody/antigen tests for hepatitis (IHC-402, IHBSG-402, IHE-302), HIV (IHI-402), syphilis (ISY-N402), chikungunya (ICHI-402), dengue (IDEC-425) and tetanus (ITE-402), with the same cassette technology and visual readout.
- Storage: keep the kit between 2 and 30 °C, protected from moisture and direct sunlight, and observe the expiry date printed on the sealed pouch.
- Confirmation: reactive or equivocal results must be confirmed by reference laboratory methods before making clinical decisions.