Description
1. Commercial Introduction
The disposable SpO₂ sensor (ref. 33755) is the practical and safe solution for continuous monitoring of oxygen saturation in two very common patient profiles: the neonate under 3 kg (placement on the foot) and the adult over 40 kg (placement on the finger). Compatible with monitors and pulse oximeters with Nellcor® connection (except Nellcor Oximax™ technology), this single-use consumable is supplied in sterile packaging with a 70 cm cable, so it is available immediately, hygienically and without any possibility of cross-contamination between patients. An ideal choice for maternity, neonatal units, emergency departments, hospital wards, outpatient clinics and diagnostic centres.
2. Technical Explanation
Pulse oximetry non-invasively measures blood oxygen saturation (SpO₂) and pulse rate by means of the difference in absorption of red light (650–670 nm) and infrared light (880–910 nm) through the vascular bed. This disposable optical sensor incorporates both emitters with powers of 6.65 mW and 6.75 mW respectively, and its dual design allows the application area to be adapted according to the patient’s weight: on the neonate’s foot (40 kg). Being single-use and supplied sterile, traceability and maximum hygiene are guaranteed in each measurement. It is not compatible with Nellcor Oximax™ technology.
3. Technical Specifications
Technical data verified according to the manufacturer’s catalogue:
| Parameter | Specification |
|---|---|
| SpO₂ measurement range | 70 % ~ 100 % |
| SpO₂ accuracy | 70 % ~ 100 % ±2 %; below 70 %: not specified |
| Pulse measurement range | 30 ~ 250 bpm |
| Pulse accuracy | ±2 bpm or 2% (whichever is greater) |
| Optical sensor – red light | Wavelength 650~670 nm, 6.65 mW |
| Optical sensor – infrared light | Wavelength 880~910 nm, 6.75 mW |
| Target population | Neonate <3 kg (foot) and adult >40 kg (finger) |
| Compatibility | Nellcor® (not compatible with Nellcor Oximax™) |
| Presentation | Individual sterile packaging with 70 cm cable |
Product data
| Manufacturer | GIMA |
|---|---|
| Manufacturer reference | 33755 |
| SKU Cahnos | GM33755 |
| Sales unit | Box of 10 units (box of 10 pcs) |
| Device type | Medical device (Medical device) |
| Device class | II B |
| CND | C900301 |
| EAN | 6945040104346 |
Available references
| Reference | SKU Cahnos | Description | Sales unit |
|---|---|---|---|
| 33755 | GM33755 | Disposable SpO₂ sensor for neonates (<3 kg, foot) and adults (>40 kg, finger) compatible with Nellcor® | Box of 10 units |
4. Equivalences and Cross-References
The manufacturer’s original reference is 33755 (.), which corresponds to SKU GM33755 on Cahnos. No validated 1:1 equivalence table of this sensor with codes from other manufacturers is available from reliable sources, so no equivalent third-party references are given. The compatibility verified by the manufacturer is limited to equipment with Nellcor® connection (excluding Nellcor Oximax™ technology); before replacing any pulse oximetry consumable, always confirm the connector and measurement system of your monitor with our technical-commercial department.
5. Selection tips and variants of interest
Check the connection system of your pulse oximeter or monitor: this sensor is compatible with conventional Nellcor® technology and not with Nellcor Oximax™. Select the application area according to the patient’s weight: neonate’s foot for those under 3 kg and adult’s finger for those over 40 kg. As it is a single-use product supplied in sterile packaging, use a new unit per patient and check the integrity of the packaging before opening it. Its presentation in a box of 10 units is the most efficient option for services with high monitoring turnover. For volume orders, hospital supplies or technical advice, contact the Cahnos team. For large orders you can request a quote.