Description
1. Introduction
The Vasofix Safety PUR safety catheter from B. Braun in gauge 14G (2.2 × 50 mm) is the reference solution for obtaining large-bore peripheral venous access in critical situations: emergency, resuscitation, surgery, ICU and blood donation. Its fully automatic passive protection system drastically reduces the risk of accidental needlesticks for healthcare staff, in line with current regulations on handling sharps, without compromising the speed or precision of cannulation. The 14G gauge (orange colour) is the highest-flow option in the range, designed for rapid transfusion of blood and blood products, high-viscosity fluids or high-flow intravenous therapies.
2. Technical explanation of the procedure
In peripheral intravenous cannulation, the catheter is introduced mounted on a steel needle-cannula that guides advancement through the skin and the venous wall. The flashback chamber technology allows immediate confirmation of correct positioning through rapid visualisation of blood return in the catheter and needle. The universal back cut bevel offers a wide choice of insertion angles with minimal puncture trauma. Once the vein is cannulated, the needle is withdrawn: the passive safety shield deploys fully automatically covering the tip, eliminating re-exposure and the risk of accidental inoculation. The radiopaque polyurethane (PUR) catheter remains in the vein, flexible and kink-resistant, and its injection port with valve allows intermittent administration of medication without puncturing the line, with closing cap included. The set is compatible with standard Luer ports and is free from latex, PVC and DEHP.
3. Technical Specifications
| Parameter | Specification |
| Manufacturer reference | 4269225S-01 (VASOFIX SFTY PUR 14G, 2 IN, 2.2X50MM-EU) |
| Family / Brand | Vasofix Safety · B. Braun |
| Gauge | 14G |
| Size | 2.2 × 50 mm (2.2 mm outer diameter; 50 mm / 2 inches length) |
| Flow rate | Approx. 343 ml/min |
| Connection | Luer-Lock / female connector |
| Material | Radiopaque polyurethane (PUR) |
| Injection port | Yes, with injection valve and closing cap |
| Safety system | Fully automatic passive protection against needlesticks |
| Identification colour | Orange (14G gauge code, according to ISO 10555-5) |
| Sterility / use | Sterile · Single-use |
| Without latex, PVC or DEHP | Yes |
| Device class | Medical device class IIa (EU) |
| Presentation | Box of 50 units |
Complete Vasofix Safety PUR range available at Cahnos:
| Reference | Gauge | Size |
| 4269225S-01 | 14G | 2.2 × 50 mm (product analysed) |
| 4269179S-01 | 16G | 1.7 × 50 mm |
| 4269330S-01 | 18G | 1.3 × 33 mm |
| 4269136S-01 | 18G | 1.3 × 45 mm |
| 4269110S-01 | 20G | 1.1 × 33 mm |
| 4269217S-01 | 20G | 1.1 × 25 mm |
| 4269098S-01 | 22G | 0.9 × 25 mm |
| 4269071S-01 | 24G | 0.7 × 19 mm |
4. Equivalences and Cross-References
Reference 4269225S-01 corresponds to the official manufacturer nomenclature B. Braun for the Vasofix Safety PUR safety catheter in 14G × 2″ (2.2 × 50 mm, orange colour), confirmed in the manufacturer’s official catalogue, so the code is considered valid. For functional comparison purposes (not 1:1 interchangeable), the closest alternative from another brand by gauge and application is BD:
| Manufacturer / Brand | Family / Model | Code / SKU | Equivalence Level | Technical Differences |
| BD (Becton Dickinson) | Insyte Autoguard (Vialon biomaterial) | 381467 | Comparable by category (not interchangeable 1:1) | Same 14G gauge (orange) and same outer diameter 2.1-2.2 mm, but length 1.75″ (≈45 mm) versus 50 mm for the Vasofix Safety; Vialon biomaterial versus radiopaque PUR; button-activated safety versus automatic passive protection; flow rate ≈330 ml/min. |
On the market there are also equivalent safety lines at category level from manufacturers such as BD (Venflon Pro Safety), ICU Medical (ProtectIV range), Vygon (Leaderflex) or Nipro (SafeTouch); their suitability must be confirmed against each manufacturer’s IFU before any substitution. There is no manufacturer-validated equivalence table that allows this catheter to be substituted 1:1 by another brand; consult the technical department before use in tenders.
5. Selection tips and variants of interest
The 14G gauge (orange colour) is the highest-flow in the Vasofix Safety family: it is indicated for adult patients with good venous access who require rapid transfusion of blood and blood products, high-viscosity fluids or high-volume therapies in critical settings (emergency, operating theatre, ICU, blood donation). For smaller-gauge lines or maintenance therapies, consider stepping down within the same family: 16G × 50 mm (4269179S-01), 18G × 33/45 mm (4269330S-01 / 4269136S-01), 20G × 33/25 mm (4269110S-01 / 4269217S-01), 22G × 25 mm (4269098S-01) or 24G × 19 mm (4269071S-01). To maintain line patency between administrations, it is recommended to complete the set with the obturator stylet of the same gauge: for the 14G × 50 mm, the Vasofix stylet Ref. 4219201 (box of 50 units). Consult your sales representative to configure the combined stock of gauges according to the patient profile of your department.