Description
1. Introduction
The ARTro radiofrequency generator from REGER Medizintechnik GmbH is the high-end energy unit of the REGER arthroscopic plasma system. Developed specifically for arthroscopy, it incorporates the latest technology and software from the German manufacturer to ensure precise and safe electrosurgical treatment of intra-articular soft tissue. This product includes the ARTro unit (ref. 897-10) together with its foot pedal (ref. 897-7), forming the complete configuration for working with the handles, BiVac bipolar cables and electrodes of the system. A reliable, high-performance solution for hospitals, clinics and arthroscopic surgery centres.
2. Technical explanation of the equipment
The ARTro generator concentrates radiofrequency energy at the tip of the active electrode to vaporise, ablate and coagulate soft tissue within the joint space in arthroscopic procedures of the knee, shoulder, elbow and ankle. The unit works with an arthroscopic plasma system and allows precise control of the energy applied to the tissue, with a controlled thermal effect on adjacent structures. The foot pedal (ref. 897-7) allows hands-free activation of the generator during the procedure, leaving both of the surgeon’s hands free to handle the arthroscopic instruments.
3. Technical specifications and references
ARTro system specifications
| Characteristic | Value |
|---|---|
| Brand | REGER |
| Manufacturer | REGER Medizintechnik GmbH |
| Set type | High-end high-frequency (HF) / radiofrequency (RF) generator |
| Indication | Arthroscopy (developed specifically for this specialty) |
| Working frequency | 333 kHz |
| Operating modes | Bipolar and monopolar |
| Actuation | Foot pedal (footswitch) and hand switch |
| Technology and software | Latest technology and software from the manufacturer |
| Supply | ARTro unit (897-10) and foot pedal (897-7) |
Version data with technical detail (variant 897-10V)
| Characteristic | Value |
|---|---|
| Voltage | 110V |
| Output power | Bipolar cut: 250 W; bipolar coagulation: 200 W; monopolar cut: 200 W; monopolar coagulation: 180 W |
| Rated input power | 1350 VA |
| Frequency | 50/60 Hz |
| Equipment protection | 2 x T 8A/250V |
| Protection against excessive heating factor | Automatic current control so that the heating factor does not exceed 30 A²s |
| NEM system (neutral electrode monitoring) | Neutral electrode monitoring |
| Protection / safety / risk class | I / CF / IIb |
| IP code | IP 2X according to EN 60529 |
| Low/high frequency leakage currents | IEC 60601-2-2 / IEC 60601-1-1 |
| Power selection | Levels 1-9 |
| Scope of delivery | Mains cable and user manual |
Available references
| REGER reference | Type | Compatible with |
|---|---|---|
| 897-7 | Foot switch | ARTro |
| 897-10 | ARTro unit |
4. Equivalences and Cross-References
This set is identified in the REGER Medizintechnik catalogue by the references RGR-ARTRO-GEN (SKU Cahnos), RGR-897-10 / 897-10 (ARTro unit) and RGR-897-7 / 897-7 (foot pedal). Reference 897-10V corresponds to the version of the ARTro radiofrequency generator with a published detailed technical data sheet (supplier path: LYNX-4K/HD Tower > RF-Plasma System > RF-Generator). The ecosystem is completed by the BiVac bipolar handles and cables (series 92888/92889), the single-use BiVac vaporisation electrodes (series 890/891), the BiCart bipolar electrodes and the monopolar arthroscopy electrodes from series 800/802/805. No direct 1:1 equivalence tables with generators from other brands have been identified in reliable sources; to validate the replacement or integration of platforms, please consult our technical department.
5. Selection tips and variants of interest
This ARTro generator with foot pedal is the ideal option if you have or plan to install a REGER arthroscopic plasma system (BiVac/ARTro) and want the complete configuration of energy unit and activation. The foot pedal (ref. 897-7) is the recommended complement for hands-free activation during the procedure. Complete your configuration with the BiVac bipolar handles and cables (references 92889-6, 92889-14, 92888-6 and 92888-16), the single-use vaporisation electrodes and the BiCart bipolar electrodes of the system. Always check the supply voltage of your installation before commissioning. Product without published price: request your personalised quote from the technical-commercial team at Cahnos.







Reviews