Description
1. Introduction
The Set Contiplex® C from B. Braun is the continuous regional anaesthesia solution that simplifies perineural catheter placement: a catheter-over-needle system that combines the simplicity of a single puncture with the analgesic power of a continuous infusion catheter. Designed for anaesthesiology departments, pain units and enhanced recovery after surgery (ERAS) protocols, it allows continuous blockade of plexuses and peripheral nerves with less trauma, less complexity and greater patient comfort, since the puncture cannula is of fine gauge (25 G).
2. Technical explanation of the equipment and procedure
The set is based on a differentiating design: the Contiplex® C puncture needle (25 G) is mounted inside the Contiplex® C catheter (19 G) and both components advance together during puncture. Once the target structure is located using neurostimulation and/or ultrasound imaging (dual guidance), the needle is removed and the echogenic catheter remains in final position in a single step, without the need to thread or advance it secondarily. This reduces the number of manoeuvres, tissue trauma and the risk of catheter displacement.
The echogenic catheter offers excellent visibility under ultrasound control, facilitating verification of local anaesthetic spread around the nerve. The system is completed with the C-grip (control handle during puncture), the Perifix® catheter connector, the Perifix® 0.2-micron filter, the Perifix® PinPad (adhesive fixation of the filter) and the infusion extension line. It also includes an adhesive label identifying the catheter location.
Included components
- Contiplex® C puncture needle (25 G).
- Contiplex® C echogenic catheter (19 G).
- Adhesive label identifying the catheter location.
- C-grip.
- Perifix® catheter connector.
- Perifix® 0.2-micron filter.
- Perifix® PinPad.
- Infusion extension line.
3. Technical Specifications
| Characteristic | Specification |
| System | Catheter-over-needle for continuous blockade of plexuses and peripheral nerves (cPNB) |
| Product type | Needle |
| Application | cPNB |
| Bevel type | Facet |
| Bevel angle | 15 ° |
| Graduation | Graduated every 1 cm |
| Needle gauge | 25 G |
| Needle outer diameter | 0.53 mm |
| Needle length | 190 mm (7 1/2″) |
| Catheter calibre | 19 G |
| Catheter outer diameter | 1.05 mm |
| Catheter inner diameter | 0.60 mm |
| Catheter length | 187 mm |
| Catheter material | PA, PUR |
| Catheter holes | Central opening |
| Guidance | Neurostimulation and/or ultrasound imaging (dual guidance) |
| Filter | Perifix® 0.2 µm (pore size) |
| Filter priming volume | 0.45 ml |
| Maximum pressure | 7.0 bar |
| Catheter connector included | Yes (Perifix®) |
| PinPad included | Yes (Perifix® PinPad) |
| Packaging | Blister |
| Contains PVC | Yes |
| Contains latex | No |
| Contains DEHP | No |
| Use | Single-use |
| Sterility | Sterile |
| Sterilisation | Ethylene oxide (EO) |
Available references
| Reference | Description | Presentation |
| 4898115 | SET CONTIPLEX C | Box of 5 units |
Products are sold in boxes of 5 units.
4. Equivalences and Cross-References
Within the Contiplex® family itself from B. Braun, reference 4898115 (also coded as BBR4898115) corresponds to the standard version of the Contiplex® C set with Luer connection. The version with NRFit® safety connectors (ISO 80369-6) is marketed under reference 4898115NR / BBR4898115NR: it performs the same clinical function (same catheter-over-needle system with 25 G needle and 19 G echogenic catheter), but is not 1:1 interchangeable with standard Luer connection systems, so connector compatibility must be verified by the department before substitution.
| Manufacturer / Brand | Family / Model | Code / SKU | Equivalence level | Observations |
| B. Braun | Contiplex® C NRFit® | 4898115NR / BBR4898115NR | Comparable by category (not interchangeable 1:1) | Same catheter-over-needle system, but with NRFit® connection (ISO 80369-6) throughout the infusion line. |
| B. Braun | Contiplex® C (Luer variant) | 4898130 | Comparable by category (not interchangeable 1:1) | Variant of the same family with Luer connection; verify sizes and components before substitution. |
| B. Braun | Contiplex® S Ultra 360 (NRFit®) | — | Comparable by category (not interchangeable 1:1) | Same family with catheter-through-needle (18 G) and 360° echogenic needle. |
| B. Braun | Contiplex® Tuohy Ultra 360 (NRFit®) | — | Comparable by category (not interchangeable 1:1) | Set with Tuohy bevel for continuous catheterisation. |
| Teleflex (Arrow) | StimuCath™ Continuous Peripheral Nerve Block Set | AB-19608-S | Comparable by category (not interchangeable 1:1) | Complete continuous blockade system; verify configuration and references before substitution. |
There is no equivalence table officially validated by the manufacturer for cross-referencing this reference with other brands; the correspondences are offered as guidance by category. Before any substitution in tenders or for an alternative brand, consult the technical department to verify gauges, lengths and connector compatibility.
5. Selection tips and variants of interest
- Connection: this standard version incorporates a Luer connection. If your hospital works with NRFit® safety systems (ISO 80369-6), consider the Contiplex® C NRFit® version (ref. 4898115NR).
- Guidance technique: the set is prepared for dual guidance (ultrasound + neurostimulation); the 15° faceted bevel of the 25 G needle offers fine, atraumatic sliding, suitable for superficial approaches and patients with delicate tissue.
- System advantage: by advancing needle and catheter together, placement is performed in a single step, reducing manoeuvres and the risk of catheter displacement compared with conventional catheterisation systems.
- House variants: for continuous blockade with a needle-through-catheter of 18 G, consider the Contiplex® S Ultra 360 or Contiplex® Tuohy Ultra 360 sets, which share echogenic technology and Perifix® components.
- Compatibility: for an optimal connection between needle and neurostimulation cable, the manufacturer recommends combining products from the same brand; consult the instructions for use (IFU).
- Single-use: sterile single-use product; check the integrity of the packaging and the expiry date before each procedure.






Reviews