Description
1. Introduction
The Contiplex® Tuohy Ultra 360 Set from B. Braun is the continuous regional anaesthesia solution for anaesthesiology departments and pain units, as well as for enhanced recovery protocols (ERAS). Its catheter-through-needle system with Tuohy bevel allows locating plexuses and peripheral nerves using dual guidance (ultrasound imaging and neurostimulation) and leaving a stable infusion line for local anaesthetic in place for prolonged postoperative analgesia, with lower doses and reduced opioid consumption.
2. Technical explanation of the equipment and procedure
The set is based on a catheter-through-needle system: the Contiplex® 18 G puncture needle, with Tuohy bevel and insulating coating, ensures optimal puncture and sliding and facilitates catheter direction during placement. Its exclusive echogenic design, with a pattern of reflective marks arranged at 360°, guarantees clear visualisation of the needle and its tip from any position during ultrasound control; the safety code facilitates immediate identification of the tip, and the ergonomic hub provides excellent tactile sensitivity, with depth marks (graduation every 1 cm) offering direct visual feedback.
Once the target structure has been located using neurostimulation and/or ultrasound imaging, the echogenic Contiplex® Ultra catheter is advanced through the needle. The adaptable side port allows simultaneous injection/suction of the local anaesthetic and catheter placement without manipulating the assembly. The system is completed with the adhesive label identifying the catheter location, the Perifix® catheter connector, the Perifix® 0.2-micron filter, the Perifix® PinPad (adhesive fixation of the filter) and the infusion extension line.
3. Technical Specifications
| Product type | Catheter-through-needle set (puncture needle) |
| Application | cPNB – continuous anaesthetic block of plexuses and peripheral nerves |
| Puncture needle | Contiplex® 18 G, outer Ø 1.30 mm, Tuohy bevel, insulating coating, graduated every 1 cm |
| Ultrasound visibility | Reflective marks at 360° (Ultra 360) and tip safety code |
| Catheter | Echogenic Contiplex® Ultra, 20 G, outer Ø 0.85 mm, inner Ø 0.45 mm, 6 side holes |
| Catheter material | PUR, PA (polyurethane, polyamide) |
| Filter | Perifix® 0.2 µm; priming volume 0.45 ml |
| Maximum pressure | 7.0 bar |
| Syringe included | 1 x Omnifix 5 ml – LL (Luer-Lock) |
| Catheter connector included | Yes (Perifix®) |
| PinPad included | Yes (Perifix® PinPad) |
| Packaging | Sterile blister |
| Contains PVC / latex / DEHP | PVC: yes / latex: no / DEHP: no |
| Use | Single-use |
| Sterility / sterilisation | Sterile / ethylene oxide (EO) |
| Presentation | Box of 10 units |
Available references
The references differ in the length of the needle and that of the associated catheter:
| Reference | Description | Gauge | Needle length | Catheter | Presentation |
| 4898705-01 | CONTIPLEX TUOHY ULTRA 360 G18 1.3X50MM | 18 G (Ø 1.30 mm) | 50 mm (2″) | 20 G – 400 mm (outer Ø 0.85 mm / inner Ø 0.45 mm) | Box of 10 units |
| 4898704-01 | CONTIPLEX TUOHY ULTRA 360 G18 1.3X40MM | 18 G (Ø 1.30 mm) | 40 mm (1 1/2″) | 20 G – 400 mm (outer Ø 0.85 mm / inner Ø 0.45 mm) | Box of 10 units |
| 4898710-01 | CONTIPLEX TUOHY ULTRA 360 G18 1.3X100MM | 18 G (Ø 1.30 mm) | 100 mm (4″) | 20 G – 1010 mm (outer Ø 0.85 mm / inner Ø 0.45 mm) | Box of 10 units |
| 4898715-01 | CONTIPLEX TUOHY ULTRA 360 G18 1.3X150MM | 18 G (Ø 1.30 mm) | 150 mm (6″) | 20 G – 1010 mm (outer Ø 0.85 mm / inner Ø 0.45 mm) | Box of 10 units |
4. Equivalences and Cross-References
Within the Contiplex® family from B. Braun, the standard version with Luer connection (this set) differs from the Contiplex® Tuohy Ultra 360 NRFit® version exclusively in the type of connection: the NRFit® variant incorporates safety connectors according to ISO 80369-6 throughout the infusion line (needle, syringe, connector, filter and extension line) and is not interchangeable 1:1 with Luer systems.
| Manufacturer / Brand | Family / Model | Code / SKU | Equivalence level | Observations |
| B. Braun | Contiplex® Tuohy Ultra 360 NRFit® | 4898705NR-01 / 4898715NR-01 / 4898704NR-01 / 4898710NR-01 | Comparable by category (not interchangeable 1:1) | Same catheter-through-needle system and Tuohy bevel, but with NRFit® connection (ISO 80369-6): incompatible with Luer syringes and systems. |
| B. Braun | Contiplex® S Ultra 360 (NRFit®) | — | Comparable by category (not interchangeable 1:1) | Same family with 360° echogenic needle; 20° bevel instead of Tuohy. |
| B. Braun | Contiplex® C (NRFit®) | 4898115NR / BBR4898115NR | Comparable by category (not interchangeable 1:1) | Catheter-over-needle system (25 G) with NRFit® connection; different placement technique. |
| Teleflex (Arrow) | StimuCath™ Continuous Peripheral Nerve Block Set | AB-19608-S | Comparable by category (not interchangeable 1:1) | Complete continuous block system with stimulating catheter and introducer; verify configuration, gauges and connection before replacement. |
There is no equivalence table officially validated by the manufacturer for cross-referencing these references with other brands; the indicated correspondences are offered as guidance by category. Before any substitution in tenders or for an alternative brand, consult the technical department to verify gauges, lengths, bevel and connector compatibility.
5. Selection tips and variants of interest
- Choice of needle length: the 40 and 50 mm references (4898704-01 and 4898705-01) cover superficial approaches such as brachial plexus, femoral or peripheral nerves in patients with thin tissue; the 100 mm (4898710-01) and 150 mm (4898715-01) ones are reserved for deep blocks or patients with greater adipose tissue.
- Luer connection: this set is supplied with a Luer-Lock connection. If your hospital works with NRFit® syringes, three-way stopcocks and infusion circuits (ISO 80369-6), consider the Contiplex® Tuohy Ultra 360 NRFit® version.
- Guidance technique: prepared for dual guidance (ultrasound imaging + neurostimulation) thanks to the 360° echogenic needle and direct connection to the neurostimulator; the Tuohy bevel facilitates catheter direction in through-the-needle approaches.
- Manufacturer variants: for continuous block with a 20° bevel, consider the Contiplex® S Ultra 360 set; for a 25 G catheter-over-needle system, consider the Contiplex® C set.
- Compatibility: for an optimal and safe connection between the neurostimulator and the needle, the manufacturer recommends combining products from the same brand; consult the instructions for use (IFU).
- Single-use: sterile single-use product; check the integrity of the packaging and the expiry date before each procedure. It is marketed in boxes of 10 units.








Reviews