Description
1. Introduction
The Contiplex® S Ultra 360 NRFit® set from B. Braun is the continuous regional anaesthesia solution for anaesthesiology departments, pain units and enhanced recovery protocols (ERAS). Its catheter-through-needle design with a 20° bevel allows locating plexuses and peripheral nerves using dual guidance (ultrasound imaging and neurostimulation) and leaving a stable infusion line for local anaesthetic for prolonged postoperative analgesia, with lower doses and reduced opioid consumption.
It is the variant of the Contiplex® S Ultra 360 family with NRFit® safety connectors (ISO 80369-6) along the entire infusion line, ideal for hospitals working with NRFit® syringes, three-way stopcocks and infusion circuits.
2. Technical explanation of the equipment and procedure
The set is based on a catheter-through-needle system: the Contiplex® 18 G puncture needle (outer Ø 1.30 mm), with a 20° facet bevel and insulating coating, allows optimal puncture and sliding. Its exclusive echogenic design, with a pattern of reflector marks arranged in 360°, guarantees clear visualisation of the needle and its tip from any position during ultrasound control. The safety code facilitates immediate identification of the tip, and the ergonomic hub provides excellent tactile sensitivity, with depth marks (graduation every 1 cm) offering direct visual feedback.
Once the target structure is located, the echogenic Contiplex® Ultra catheter is advanced through the needle. The adaptable side port allows simultaneous injection/suction of the local anaesthetic with catheter placement without manipulating the assembly. The system is completed with the adhesive label identifying the catheter location, the Perifix® catheter connector, the Perifix® 0.2 µm filter, the Perifix® PinPad (adhesive fixation of the filter) and the extension line, all with NRFit® connection.
3. Technical Specifications
| Characteristic | Specification |
| Product type | Catheter-through-needle set (puncture needle) |
| Application | cPNB – continuous anaesthetic block of plexuses and peripheral nerves |
| Puncture needle | Contiplex® 18 G, outer Ø 1.30 mm, 20° bevel, insulating coating, graduated every 1 cm |
| Ultrasound visibility | Reflective marks at 360° (Ultra 360) and tip safety code |
| Connection | NRFit® according to ISO 80369-6 (needle, syringe, connector, filter and extension line) |
| Catheter | Echogenic Contiplex® Ultra, 20 G, outer Ø 0.85 mm, inner Ø 0.45 mm, 6 side holes |
| Catheter material | PUR, PA (polyurethane, polyamide) |
| Filter | Perifix® 0.2 µm with NRFit® connection; priming volume 0.45 ml |
| Maximum pressure | 7.0 bar |
| Syringe included | 1 x Omnifix 5 ml NRFit – Lock |
| Catheter connector included | Yes (Perifix® with NRFit® connection) |
| PinPad included | Yes (Perifix® PinPad) |
| Packaging | Sterile blister |
| Contains PVC / latex / DEHP | PVC: yes / latex: no / DEHP: no |
| Use | Single-use |
| Sterility / sterilisation | Sterile / ethylene oxide (EO) |
| Presentation | Box of 10 units |
Available references (sold in boxes of 10 units). The references differ by needle length and associated catheter length:
| Reference | Description | Gauge | Needle length | Catheter | Presentation |
| 4898610NR-27 | SET CONTIPLEX S ULTRA 360 1.3 x 100 mm N | 18 G (Ø 1.30 mm) | 100 mm (4″) | 20 G – 1010 mm (outer Ø 0.85 mm / inner Ø 0.45 mm) | Box of 10 units |
| 4898650NR-27 | SET CONTIPLEX S ULTRA 360 1.3 x 50 mm NR | 18 G (Ø 1.30 mm) | 50 mm (2″) | 20 G – 400 mm (outer Ø 0.85 mm / inner Ø 0.45 mm) | Box of 10 units |
| 4898615NR-27 | SET CONTIPLEX S ULTRA 360 1.3 x 150 mm N | 18 G (Ø 1.30 mm) | 150 mm (6″) | 20 G – 1010 mm (outer Ø 0.85 mm / inner Ø 0.45 mm) | Box of 10 units |
4. Equivalences and Cross-References
Within the Contiplex® family from B. Braun, the NRFit® version (this set) incorporates NRFit® safety connectors (ISO 80369-6) along the entire infusion line and is not 1:1 interchangeable with standard Luer versions.
| Manufacturer / Brand | Family / Model | Code / SKU | Equivalence level | Observations |
| B. Braun | Contiplex® S Ultra 360 (Luer version, without NRFit®) | — | Comparable by category (not interchangeable 1:1) | Same catheter-through-needle system and 20° bevel, but with Luer connection: incompatible with NRFit® syringes and systems ISO 80369-6. |
| B. Braun | Contiplex® Tuohy Ultra 360 NRFit® | 4898705NR-01 / 4898704NR-01 / 4898710NR-01 / 4898715NR-01 (BBR…) | Comparable by category (not interchangeable 1:1) | Same NRFit® family with 360° echogenic needle; Tuohy bevel instead of 20° bevel. |
| B. Braun | Contiplex® C NRFit® | 4898115NR / BBR4898115NR | Comparable by category (not interchangeable 1:1) | Catheter-over-needle system (25 G) with NRFit® connection; different placement technique. |
| Teleflex (Arrow) | StimuCath™ Continuous Peripheral Nerve Block Set | AB-19608-S | Comparable by category (not interchangeable 1:1) | Complete continuous block system with stimulating catheter and introducer; verify configuration, gauges and connection before substitution. |
There is no equivalence table officially validated by the manufacturer for cross-referencing these references with other brands; the correspondences indicated are offered as guidance by category. Before any substitution in tender or for alternative brand, consult the technical department to verify gauges, lengths, bevel and connector compatibility.
5. Selection tips and variants of interest
- Choice of needle length: the 50 mm reference (4898650NR-27) covers superficial approaches such as brachial plexus, femoral or peripheral nerves in patients with thin tissue; the 100 mm (4898610NR-27) and 150 mm (4898615NR-27) are reserved for deep blocks or patients with greater adipose tissue.
- NRFit® connection: choose this set if your hospital works with NRFit® syringes, three-way stopcocks and infusion circuits (ISO 80369-6). If your equipment park is Luer, consider the standard Contiplex® S Ultra 360 version.
- Guidance technique: the set is prepared for dual guidance (ultrasound + neurostimulation) thanks to the 360° echogenic needle and direct connection to the neurostimulator; the 20° bevel facilitates puncture and catheter advancement.
- House variants: for continuous block with Tuohy bevel consider the Contiplex® Tuohy Ultra 360 NRFit® set; for catheter-over-needle system of 25 G, consider the Contiplex® C NRFit® set.
- Compatibility: for optimal and safe connection between the neurostimulator and the needle, the manufacturer recommends combining products from the same brand; consult the instructions for use (IFU).
- Single use: sterile single-use product; check package integrity and expiry date before each procedure. Supplied in boxes of 10 units.







Reviews