Description
1. Product Presentation
The multifilament absorbable suture PREMISUT RAPID Prime Polyglactin 910 (PGLAR) is the professional choice for ophthalmic surgery services, general surgery and procedures that require a secure, manageable closure with controlled and accelerated absorption. It combines a ½ circle needle with cylindrical body and trocar point (tapercut) of 17 mm (HRT 17 series) with a colourless thread of 70 cm in USP 3/0 gauge.
It is presented in B2B format in a box of 12 units, designed for continuous supply to operating theatres, ophthalmology clinics, distributors and purchasing centres. It is a synthetic, absorbable, sterile, single-use medical device, with no possibility of re-sterilisation, offering an excellent quality-price ratio.
2. Technical explanation of the material and procedure
Polyglactin 910 is a synthetic absorbable copolymer composed of 90 % glycolide and 10 % L-lactide, presented in a braided multifilament structure that provides great flexibility, smooth passage through tissue and excellent knotting behaviour. The RAPID version is characterised by an accelerated loss of strength: it retains 50 % of its tensile strength at 5 days and loses it completely at 14 days, with total absorption by hydrolysis between 40 and 45 days, a predictable process independent of the patient’s enzymatic activity.
The ½ circle needle with cylindrical body and trocar point (tapercut, HRT series) combines the atraumatic penetration of a round body with a sharp cutting tip, allowing it to pass through firm or inelastic tissues without tearing or enlarging the exit hole. The USP 3/0 gauge (metric 2) provides an optimal balance between strength and handling for fine to medium thickness sutures. The colourless colour of the thread offers maximum discretion on the surgical field.
3. Technical Specifications
3.1. Technical data sheet of the reference
| Parameter | Specification |
| Reference (SKU) | AMP2L28HRT1730 (manufacturer ref. P2L28HRT1730) |
| Brand / Quality | PREMISUT RAPID – PGLAR PRIME |
| Material | Polyglactin 910 (copolymer 90 % glycolide / 10 % L-lactide) |
| Thread structure | Braided multifilament, undyed |
| Gauge | USP 3/0 (metric 2; approximate diameter 0.20–0.249 mm) |
| Thread colour | Colourless |
| Thread length | 70 cm |
| Needle shape | ½ circle (1/2) |
| Needle body | Cylindrical |
| Needle tip | Trocar (tapercut) |
| Needle size | 17 mm (HRT 17) |
| Absorption mechanism | Hydrolysis (40–45 days) |
| Sterilisation | Gamma radiation – sterile, single-use |
| Tissue reaction | Minimal; excellent tensile strength and knot security |
| Indications | Tissue approximation, ligatures and ophthalmic surgery |
| Presentation | Box of 12 units (minimum order 30 boxes) |
3.2. Tensile strength retention profile
| Period after implantation | Retained tensile strength |
| 5 days | 50 % of initial tensile strength |
| 14 days | 0% of the initial tensile strength |
| Total absorption | 40–45 days, by hydrolysis |
4. Equivalences and Cross-References
Rapid absorption polyglactin 910 is a standardised category whose international clinical reference is the VICRYL RAPIDE line from Ethicon (J&J MedTech), with which it shares composition (copolymer 90 % glycolide / 10 % low molecular weight L-lactide), absorption profile and accelerated loss of strength. After verification in public sources (official Ethicon EMEA catalogue and authorised distributors), the following cross-reference is included:
| Manufacturer / Brand | Family / Model | Reference code | Equivalence level | Technical differences / Remarks |
| Ethicon (J&J MedTech) | VICRYL RAPIDE, V-4 needle | VR2280 | Comparable by category (not interchangeable 1:1) | Same material (low molecular weight polyglactin 910), USP 3/0 gauge, colourless thread of 75 cm and tapercut point (V-4); the needle geometry differs: 3/8 circle of 17 mm versus ½ circle of 17 mm (HRT 17) of this reference, as well as thread length (75 cm versus 70 cm). Code confirmed in Ethicon EMEA and authorised distributors. |
Among the homologous families of rapid absorption polyglycolic without a verified cross-code for this exact configuration are Safil Quick (B. Braun), Polysorb (Medtronic / Covidien), Surgicryl 910 (SMI), Assucryl 910 (Assut) and DemeCRYL (DemeTECH); in all cases gauge, point geometry and thread length must be confirmed before any substitution. Standard VICRYL (absorption 56–72 days) is not comparable to this RAPID version.
There is no equivalence table validated by the manufacturer; consult the technical department before using in tenders.
5. Selection tips and variants of interest
The trocar point (tapercut) is indicated for firm or inelastic tissues where penetration firmness is required without enlarging the exit hole; if the tissue is soft and easy to penetrate, consider the conical point series (HRP, taperpoint ultra plus) from the same range. Always verify the USP gauge, needle size and thread length from the surgical report before placing the order.
| Reference | Variant |
| AMP2L28HRT2610 | PGLAR RAPID Prime USP 0, needle HRT 26, colourless, 70 cm (larger needle and gauge) |
| AMP2L22HRT2630 | PGLAR RAPID USP 3/0, needle HRT 26, colourless, 75 cm (larger needle, same configuration) |
| AMP2L22HRT3620 | PGLAR RAPID Prime USP 2/0, HRT 36 needle, colourless, 75 cm (larger needle and gauge) |
| AMP2L28HRP1730 | PGLAR RAPID Prime USP 3/0, needle ½ circle conical point Ultra Plus HRP 17, colourless, 70 cm (different point for soft tissue) |
| AMP2L22DRT1630 | PGLAR RAPID Prime USP 3/0, 3/8 circle trocar point needle DRT 16, colourless, 75 cm (different needle geometry) |
| AMP228HRT1730 | PGAR RAPID Prime USP 3/0, needle HRT 17, colourless, 70 cm (polyglactic acid, same design) |
As it is a sterile single-use product, check the integrity of the packaging before each procedure, respect the expiry date and store the suture in a cool, dry place, protected from sunlight. For special B2B conditions by volume or equivalents from other manufacturers, contact our technical-commercial team.