Description
1. Commercial Introduction
The absorbable multifilament suture of Polyglactin 910 (PGLA) reference 1L11DS1940 from the PREMISUT range is the reference technical solution for the professional seeking a surgical thread with predictable behaviour, excellent handling and great knot security in tissue approximation, ligatures and ophthalmic surgery. Its ⅜ circular needle with reverse cutting triangular point of 19 mm (DS 19) offers firm and controlled penetration, ideal for firm tissues and delicate structures, while the violet braided thread of 45 cm ensures optimal visibility on the surgical field and smooth sliding during suturing.
It is a synthetic absorbable medical device, sterile and single-use, with a very competitive price-quality ratio for B2B supply to operating theatres, ophthalmology practices, purchasing centres and distributors. Commercial presentation in a box of 12 units.
2. Technical Explanation
Polyglactin 910 is a synthetic absorbable copolymer composed of 90% glycolide and 10% L-lactide, presented in a braided multifilament structure with a violet grooved finish. This configuration provides a regular surface that facilitates the passage of the thread through tissue, smooth handling and firm, secure knot tying, even in fine gauges such as USP 4/0.
Absorption occurs by hydrolysis, i.e., by degradation of the polymer in contact with organic fluids, over a predictable period of 56 to 72 days. This profile makes the suture ideal for tissues that require wound support during the first weeks of healing, maintaining the necessary tensile strength while the tissue recovers its own resistance.
The ⅜ circle needle with reverse cutting triangular point (inverted triangular) of 19 mm penetrates firm tissues with controlled trauma, avoiding tears, being especially indicated in ophthalmic surgery procedures, tissue approximation and ligatures. The product is supplied sterile by ethylene oxide, single-use and without the possibility of re-sterilisation.
3. Technical Specifications
3.1. General characteristics of reference 1L11DS1940
| Characteristic | Specification |
|---|---|
| Reference (SKU) | 1L11DS1940 (PREMISUT range) |
| Thread structure | Braided coated multifilament, fluted finish |
| Composition | Polyglactin 910 (PGLA), copolymer of 90 % glycolide and 10 % L-lactide |
| Thread size | USP 4/0 (metric 1.5) |
| Thread length | 45 cm |
| Needle shape | ⅜ circle |
| Needle size | 19 mm (DS 19) |
| Tip type | Reverse cutting triangular (inverted triangular) |
| Colour and finish | Violet grooved (also available colourless) |
| Absorption | By hydrolysis between 56 and 72 days |
| Sterilisation | Ethylene oxide (single-use, not re-sterilisable) |
| Tissue reaction | Minimal, with excellent tensile strength and optimal knot tying |
| Indications | Tissue approximation, ligatures and ophthalmic surgery |
| Presentation | Box of 12 units |
3.2. Tensile strength retention profile
| Time | Tensile strength retention |
|---|---|
| 2 weeks | 75 % |
| 3 weeks | 50 % |
| 4 weeks | 30 % |
| Complete absorption | 56-72 days (hydrolysis) |
Complete absorption by hydrolysis between 56 and 72 days. Minimal tissue reaction, excellent tensile strength and optimal knot tying, with a fully predictable absorption rate. Price per box of 12 units.
4. Equivalences and Cross-References
The polyglactin 910 from PREMISUT is clinically equivalent to the absorbable braided polyglactin 910 sutures from the main international manufacturers, which facilitates its incorporation into purchasing centres, tenders and operating theatre protocols:
| Manufacturer / commercial house | Equivalent line |
|---|---|
| Ethicon (Johnson & Johnson) | Vicryl (coated braided polyglactin 910, violet colour) |
| B. Braun | Safil (braided and coated polyglactin 910, equivalent retention and absorption profile) |
| Medtronic | Absorbable braided polyglactin 910 range |
| Laboratory Aragó | Range of polyglactin 910 with ⅜ circle needles (DS series) |
Within the Cahnos catalogue, its direct equivalent in configuration (PGLA 4/0, DS 19 needle) is reference 1L12DS1940 in violet with 75 cm thread, and 1L22DS1940 in colourless grooved colour. The variants with 16 mm needle, gauges 3/0, 5/0 and 6/0, and the polyglycolic acid (PGA) family complete the rest of ophthalmic and soft tissue needs.
5. Selection tips and variants of interest
The USP 4/0 gauge is ideal for skin, subcutaneous tissue, low-tension soft tissues, ligatures and superficial ophthalmic sutures. The ⅜ needle with reverse cutting triangular point is indicated when firm tissues need to be penetrated without tearing, with an atraumatic passage highly valued in ophthalmology. The violet colour favours the visibility of the thread on the surgical field; if minimising visibility is preferred, the colourless version is available. For tissues with longer healing, consider the PGA family with absorption between 60 and 90 days.
| Reference | Variant |
|---|---|
| 1L11DS1940 | PGLA USP 4/0, ⅜ triangular needle DS 19 mm, violet, 45 cm (current product) |
| 1L12DS1940 | PGLA USP 4/0, same DS 19 mm needle, violet, 75 cm (longer thread) |
| 1L22DS1940 | PGLA USP 4/0, same DS 19 mm needle, colourless grooved, 75 cm (colourless variant) |
| 1L11DS1950 | PGLA USP 5/0, triangular needle DS 19 mm, violet, 45 cm (finer gauge) |
| 1L11DS1650 | PGLA USP 5/0, triangular needle DS 16 mm, violet, 45 cm (shorter needle) |
| 1L12DS1640 | PGLA USP 4/0, triangular needle DS 16 mm, violet, 75 cm (shorter needle) |
| AM111DS1940 | PGA USP 4/0, ⅜ triangular needle DS 19 mm, violet, 45 cm (absorption of 60 to 90 days) |
| AM112DS1940 | PGA USP 4/0, ⅜ triangular needle DS 19 mm, violet, 75 cm (absorption of 60 to 90 days) |
All references are supplied in boxes of 12 units. For special B2B conditions and volume orders, contact our technical-commercial team at Cahnos.





Reviews