Description
1. Commercial Introduction
The Prime Absorbable Multifilament PGLA (Polyglactin 910) Suture from the PREMISUT range is the reference solution for operating theatre services, ophthalmic clinics, purchasing centres and distributors seeking a high-performance synthetic absorbable thread with exceptional handling and fully predictable clinical behaviour. This presentation incorporates a ⅜ circle needle with a 30 mm reverse cutting triangular point (DS30), USP 2/0 gauge and a 75 cm colourless grooved thread, an ideal combination for tissue approximation, ligatures and ophthalmic surgery. It is supplied in a box of 12 units, sterile and ready for immediate use in hospitals, private clinics and surgical material purchasing centres.
2. Technical Explanation
Polyglactin 910 (PGLA) is a synthetic absorbable copolymer composed of 90% glycolide and 10% L-lactide, configured in a braided multifilament structure coated with a grooved finish. This architecture favours smooth thread sliding through tissue, offers excellent handling and guarantees great knot security, even in working gauges such as USP 2/0.
Its absorption occurs by hydrolysis, a predictable mechanism independent of the patient’s enzymatic activity: the polymer degrades into glycolic acid and lactic acid, substances that the body metabolises without causing significant tissue reaction. The tensile strength retention profile maintains the necessary support during the critical phase of healing. The ⅜ circle needle with reverse cutting triangular point (reverse cutting) penetrates firm, non-compliant tissues with controlled trauma and no tearing, making it especially suitable for ophthalmic surgery, ligatures and tissue approximation. The colourless colour minimises the visibility of the strand on the surgical bed, highly valued in procedures where aesthetic discretion is sought. The product is supplied sterile by ethylene oxide, is single-use and cannot be re-sterilised.
3. Technical Specifications
3.1. General characteristics
| Brand / Range | PREMISUT (Prime quality) |
| Material | Polyglactin 910 (PGLA), copolymer of 90 % glycolide and 10 % L-lactide |
| Thread structure | Braided and coated multifilament (grooved) |
| Gauge | USP 2/0 (metric 3) |
| Thread colour | Colourless (violet on request) |
| Thread length | 75 cm |
| Needle shape | 3/8 circle |
| Tip type | Reverse cutting triangular (reverse cutting) |
| Needle size | DS 30 mm (Prime) |
| Sterilisation | Ethylene oxide (EtO), single-use |
| Presentation | Box of 12 units |
| Reference (SKU) | P1L22DS3020 |
3.2. Absorption profile and tensile strength retention
| Period | Tensile strength retention |
| 2 weeks | 75 % |
| 3 weeks | 50 % |
| 4 weeks | 30 % |
| Complete absorption | 56 – 72 days by hydrolysis |
Minimal tissue reaction, excellent initial tensile strength and optimal knot tying, with a fully predictable absorption rate. Indicated for tissue approximation, ligatures and ophthalmic surgery.
3.3. Size equivalence
| USP size | Metric size | Approximate diameter |
| 2/0 | 3 | 0.30 – 0.349 mm (indicative according to Pharmacopoeia) |
4. Equivalences and Cross-References
Polyglactin 910 is the most widely used medium-duration multifilament absorbable in the surgical market. This Prime PGLA suture is equivalent in chemical composition, absorption profile (56-72 days) and tensile strength retention to the main international commercial ranges:
| Manufacturer / commercial house | Equivalent line |
| Ethicon (Johnson & Johnson) | Vicryl / Vicryl Plus (braided coated polyglactin 910) |
| B. Braun | Safil (braided polyglactin 910) |
| Medtronic / Covidien | Polysorb (absorbable braided range of polyglactin 910) |
| SMI | Surgicryl (braided polyglactin 910) |
| Laboratory Aragó | Polyglactin 910 range with ⅜ circle needles (DS series) |
Within the Cahnos catalogue, the violet version of this same configuration (DS 30 mm needle, USP 2/0, 75 cm) corresponds to reference P1L12DS3020, and the double strand 2×75 cm presentation in violet to P1L12RDS3020.2. The indicated equivalences are for guidance; for a formal substitution of references, please consult our technical-commercial department.
5. Selection tips and variants of interest
The USP 2/0 gauge (metric 3) offers an excellent balance between initial strength and handling for soft tissues of medium tension: skin, subcutaneous tissue, mucosa, fascia and ligatures. The ⅜ circle needle with 30 mm reverse cutting triangular point is especially recommended for firm, non-compliant tissues, where a clean and controlled needle passage is required, including microsurgery and ophthalmic surgery. The colourless thread provides maximum aesthetic discretion; if you prefer greater visibility on the surgical field, request the violet version.
Variants of interest in the same DS30 family:
- P1L12DS3020: PGLA Prime, DS30, USP 2/0, violet, 75 cm (field-detectable strand).
- P1L12RDS3020.2: PGLA Prime, DS30, USP 2/0, violet, double strand 2×75 cm.
- P1L12DS3030: PGLA Prime, DS30, USP 3/0, violet, 75 cm (finer gauge).
- P1L28DS3010: PGLA Prime, DS30, USP 0, colourless, 75 cm.
- P1L12DS3620: PGLA Prime, DS36, USP 2/0, violet, 75 cm (longer needle).
- AMP112DS3010: alternative in polyglycolic acid (PGA) with 60-90 day absorption.
If you need a longer retention profile (180-210 days) or a monofilament thread, consider the polydioxanone (PDO) family. All references are supplied in boxes of 12 units; consult the Cahnos technical-commercial team for minimum order and special B2B conditions by volume. As this is sterile single-use material, check the integrity of the packaging before each procedure and store it in a cool, dry place, protected from sunlight.