Description
1. Product Presentation
The multifilament absorbable suture of Polyglactin 910 (PGLA) from Cahnos, from the PREMISUT range in Prime quality, is the synthetic absorbable solution designed for the professional who demands predictable performance, excellent tensile strength and great knot security. Its 1/2 circle needle with cylindrical body and Taperpoint conical point of 10 mm (HR10) ensures smooth and controlled penetration in soft and delicate tissues, minimising tissue trauma.
It is the ideal choice for tissue approximation, ligatures and ophthalmic surgery, in a professional box format of 12 units, ideal for stock management in operating theatres, ophthalmic clinics and purchasing centres, with special B2B conditions for hospitals and volume buyers.
2. Technical explanation of the material and procedure
Polyglactin 910 (PGLA) is a synthetic absorbable copolymer composed of 90% glycolide and 10% L-lactide, presented in a braided multifilament structure that provides great flexibility, smooth passage through tissue and excellent knotting behaviour. Its absorption occurs by hydrolysis, a uniform process independent of tissue enzymes, which guarantees a predictable loss of mass between 56 and 72 days and minimal tissue reaction.
The tensile strength retention profile is maintained at 75% at 2 weeks, 50% at 3 weeks and 30% at 4 weeks, a sufficient period for the healing of most soft tissues. The 1/2 circle needle with cylindrical body and conical point (Taperpoint geometry) combines an atraumatic point, which separates fibres without cutting effect, with smooth passage through tissue: it is the usual configuration in ophthalmic surgery and in the approximation of delicate tissues. The violet colour of the thread ensures optimal visibility on the surgical field, highly valued in deep planes and in microsurgery. The product is sterilised by ethylene oxide (EO), is single-use and does not allow re-sterilisation.
3. Technical Specifications
Technical data sheet for reference P1L18HR1050:
| Parameter | Specification |
|---|---|
| Reference (SKU) | P1L18HR1050 |
| Brand / Quality | PREMISUT · Prime |
| Suture type | Synthetic absorbable multifilament |
| Material | Polyglactin 910 (PGLA), copolymer 90 % glycolide / 10 % L-lactide |
| Thread structure | Braided multifilament |
| Gauge | USP 5/0 (metric 1) |
| Thread length | 70 cm |
| Thread colour | Violet (colourless version available on request) |
| Needle shape | 1/2 circle |
| Needle body | Cylindrical |
| Needle tip | Atraumatic conical (Taperpoint) |
| Needle size | 10 mm (HR10 Prime) |
| Absorption type | By hydrolysis, between 56 and 72 days |
| Sterilisation | Ethylene oxide (EO), sterile, single-use |
| Tissue reaction | Minimal |
| Indications | Tissue approximation, ligatures and ophthalmic surgery |
| Presentation | Box of 12 units |
Tensile strength retention and absorption profile:
| Period after implantation | Retained tensile strength |
|---|---|
| 2 weeks | 75 % of initial tensile strength |
| 3 weeks | 50 % of initial tensile strength |
| 4 weeks | 30% of initial tensile strength |
| Total absorption | 56-72 days, by hydrolysis |
4. Equivalences and Cross-References
Polyglactin 910 is one of the most well-established multifilament synthetic absorbable suture raw materials on the market. This Prime reference is clinically equivalent, in chemical composition, absorption profile (56-72 days) and tensile strength retention, to the main international lines:
| Manufacturer / commercial house | Equivalent line |
|---|---|
| Ethicon (Johnson & Johnson) | Vicryl / Vicryl Plus (braided polyglactin 910, violet colour) |
| B. Braun | Safil (braided polyglactin 910, violet colour) |
| Medtronic / Covidien | Polysorb (absorbable braided range of polyglactin 910) |
| SMI | Surgicryl 910 (braided polyglactin 910) |
| Assut | Assucryl 910 (polyglactin 910) |
| DemeTECH | Vitacryl / DemeCRYL (coated braided polyglactin 910) |
The 1/2 circle needle with cylindrical body and 10 mm conical point corresponds to the classic pattern of small-format taper-type needles used in ophthalmic surgery and delicate tissues. Within the Cahnos catalogue, its variant with Ultra Plus trocar conical point of identical dimensions and gauge is reference P1L18HRP1050; in polyglycolic acid (PGA) material, reference AMP118HR1050 offers similar performance with 60-90 day absorption. The equivalences indicated are for guidance; for a formal substitution of references in your healthcare centre, please consult our technical-commercial department.
5. Selection tips and variants of interest
The USP 5/0 gauge (metric 1) is indicated for very fine and low-tension tissues: mucosa, subcutaneous tissue, delicate planes and precision approximations, with special application in ophthalmic surgery, where aesthetic healing and minimal trauma are sought. The 1/2 circle needle with 10 mm conical point (HR10) is the most atraumatic option for these tissues and smaller-radius accesses.
- P1L18HRP1050: variant with Ultra Plus trocar conical point of 10 mm, USP 5/0, violet, 70 cm (greater penetration capacity in firmer tissues).
- P1L18HR1350: same Taperpoint conical point, 13 mm needle, USP 5/0, violet, 70 cm (larger needle arc).
- P1L18HR1360: HR 13 mm needle, USP 6/0 gauge, violet, 70 cm (even finer diameter for microsurgery).
- P1L18HRP1750: Ultra Plus conical point of 17 mm, USP 5/0, violet, 70 cm (longer needle for medium-consistency tissues).
- P1L12HR1350: same HR 13 mm configuration, USP 5/0, with 75 cm thread.
- AMP118HR1050: polyglycolic acid (PGA) family with similar performance and 60-90 day absorption, for comparing suppliers and unifying stocks.
All references are supplied in boxes of 12 units. As this is sterile single-use material, check the integrity of the packaging before each procedure, respect the expiry date and store it in a cool, dry place, protected from sunlight. For special B2B volume conditions, hospital supply contracts or equivalences from other commercial houses, contact our technical-commercial team.





Reviews