Description
1. Commercial Introduction
At Cahnos we make available to you the absorbable multifilament Polyglactin 910 (PGLA) suture from the Premisut Prime range, reference P1L12HR2230, the professional reference solution for B2B supply to operating theatres, ophthalmic clinics and purchasing centres. It combines a synthetic copolymer with predictable behaviour, exceptional handling and great knot security in the approximation of soft tissues, ligatures and ophthalmic surgery.
Its ½-circle needle with cylindrical body and taper-point conical tip of 22 mm (HR 22) ensures a clean and atraumatic passage, an essential quality in delicate and friable tissues. The violet thread of 75 cm offers optimal visibility on the surgical field and smooth sliding during suturing. It is a synthetic, absorbable, sterile and single-use medical device, with an excellent price-quality ratio. Commercial presentation in a box of 12 units.
2. Technical explanation of the product and procedure
Polyglactin 910 is a synthetic absorbable copolymer composed of 90 % glycolide and 10 % L-lactide, presented in a braided multifilament structure with a grooved and coated finish. This configuration provides a regular surface that favours the sliding of the thread through the tissue, smooth handling, low memory and a firm and secure knot seating, even in fine gauges such as USP 3/0 (metric 2).
Absorption occurs by hydrolysis, that is, by degradation of the polymer in contact with organic fluids, over a predictable period of 56 to 72 days, independent of the patient’s enzymatic activity. This profile maintains the necessary tensile strength during the first 2 to 4 weeks of healing, while the tissue recovers its own resistance. The ½-circle needle with cylindrical body and taper-point conical tip separates the fibres without a cutting effect, minimising injury to the tissue bed, which is why it is especially indicated in ophthalmic surgery and in soft tissues. The product is supplied sterile by ethylene oxide, single-use and with no possibility of re-sterilisation.
3. Technical Specifications
3.1. General characteristics of reference P1L12HR2230
| Reference (SKU) | P1L12HR2230 |
| Brand / Quality | Premisut – Prime |
| Material | Polyglactin 910 (PGLA), copolymer 90 % glycolide / 10 % L-lactide |
| Thread structure | Braided, coated and fluted multifilament |
| Gauge | USP 3/0 (metric 2) |
| Colour | Violet (also available colourless in the family) |
| Needle shape | ½ circle, cylindrical body |
| Needle tip | Conical taperpoint (HR 22), atraumatic |
| Needle size | 22 mm |
| Thread length | 75 cm |
| Absorption mechanism | Hydrolysis (56-72 days) |
| Sterilisation | Ethylene oxide (single-use) |
| Tissue reaction | Minimal, with excellent tensile strength and knot security |
| Indications | Approximation of soft tissues, ligatures and ophthalmic surgery |
| Presentation | Box of 12 units |
3.2. Absorption profile and tensile strength retention
| Implantation time | Tensile strength retention |
| 2 weeks | 75 % |
| 3 weeks | 50 % |
| 4 weeks | 30 % |
| Complete absorption | 56-72 days by hydrolysis |
Complete absorption by hydrolysis between 56 and 72 days, with minimal tissue reaction, excellent tensile strength and optimal knot tying, with a fully predictable absorption rate that allows tissue support to be planned with great reliability. Price per box of 12 units; consult B2B conditions for volume orders.
4. Equivalences and Cross-References
Polyglactin 910 is the most widely used absorbable multifilament material in surgical practice and presents an absorption profile (56-72 days) and tensile strength retention equivalent to the braided lines of the main international manufacturers:
| Manufacturer | Equivalent line |
| Ethicon (Johnson & Johnson) | Vicryl (coated braided polyglactin 910, violet colour) |
| B. Braun | Safil (braided polyglactin 910, violet colour) |
| Medtronic / Covidien | Polysorb (absorbable braided range of polyglactin 910) |
| SMI | Surgicryl (absorbable polyglactin 910) |
| Assut Sutures | Range of polyglactin 910 with ½ circle cylindrical needles |
| Premisut range (Cahnos) | Multifilament PGLA with ½-circle needle (HR series, ref. P1L12HR2230) |
Within the Cahnos catalogue, its direct equivalent in double-thread format is reference P1L12HR2230.2 (PGLA Prime, HR 22 mm, USP 3/0, violet, 2×75 cm) and, in the polyglycolic acid family, the structurally equivalent alternative is reference AMP112HR2230 (PGA, cylindrical needle HR 22 mm, USP 3/0, violet, 75 cm), recommended when an absorption profile of 60 to 90 days is preferred. The brands cited belong to their respective manufacturers and are mentioned only as a reference of equivalence; for a formal substitution of references, consult our technical-commercial department.
5. Selection tips and variants of interest
To choose the most suitable suture, take into account the wound support time required by the tissue to be sutured and the necessary gauge according to the anatomical area. The USP 3/0 (metric 2) gauge is indicated for tissues with moderate tension: skin, subcutaneous tissue, mucosa, fascia and soft tissue in general. The ½-circle cylindrical needle with conical tip of 22 mm is especially recommended in soft tissues, ligatures and ophthalmic surgery, where the passage of the needle must be clean and atraumatic.
| Reference | Variant of interest |
| P1L12HR2230.2 | PGLA Prime, HR 22 mm, USP 3/0, violet, 2×75 cm (double-thread format) |
| P1L12HR2030 | PGLA Prime, HR 20 mm, USP 3/0, violet, 75 cm (shorter needle for very delicate tissues) |
| P1L12HR2630 | PGLA Prime, HR 26 mm, USP 3/0, violet, 75 cm (larger needle size) |
| P1L12HR2220 | PGLA Prime, HR 22 mm, USP 2/0, violet, 75 cm (larger gauge, greater strength) |
| P1L12HR2240 | PGLA Prime, HR 22 mm, USP 4/0, violet, 75 cm (finer) |
| P1L18HS2230 | PGLA Prime, reverse cutting triangular needle HS 22 mm, USP 3/0, violet, 70 cm (for firm tissues) |
| AMP112HR2230 | Alternative in PGA, HR 22 mm, USP 3/0, violet, 75 cm (absorption 60-90 days) |
If you need a more prolonged strength retention (180-210 days) or a monofilament profile, the polydioxanone (PDO) family is the recommended alternative. All references are supplied in a box of 12 units. As it is sterile single-use material, check the integrity of the packaging before each procedure, respect the expiry date and store it in a cool, dry place, protected from sunlight. For special B2B conditions and volume orders, contact our technical-commercial team.