Description
1. Commercial presentation
The absorbable multifilament polyglactin 910 (PGLA) suture from the Prime range, reference P1L12HR2610, is the professional reference solution for B2B supply to operating theatres, ophthalmology clinics, general surgery departments and purchasing centres. It combines a synthetic copolymer with predictable behaviour, exceptional thread handling and great knot security in the approximation of soft tissues, ligatures and ophthalmic surgery.
Its ½ circle needle with cylindrical body and conical point (taperpoint) of 26 mm ensures an atraumatic passage through the tissue, while the violet thread of 75 cm in USP 0 gauge (metric 3.5) offers optimal visibility on the surgical field and smooth sliding during suturing. It is a synthetic, absorbable, sterile and single-use medical device, with an excellent quality-price ratio for continuous use.
2. Technical explanation of the product and procedure
Polyglactin 910 (PGLA) is an absorbable synthetic copolymer composed of 90% glycolide and 10% L-lactide, presented in a braided multifilament structure with a grooved finish. This configuration provides a regular surface that facilitates the passage of the thread through the tissue, excellent flexibility, firm and secure knot tying and smooth traction, essential qualities in soft and friable tissues.
Absorption occurs by hydrolysis, i.e. by degradation of the polymer in contact with organic fluids, over a predictable period of 56 to 72 days, independent of the patient’s enzymatic activity. This profile maintains the necessary tensile strength during the first weeks of healing, while the tissue recovers its own resistance. The ½ circular needle with cylindrical body and conical point (taperpoint) HR of 26 mm allows controlled and atraumatic penetration, separating the fibres without cutting effect and minimising injury to the tissue bed, which is why it is particularly indicated in ophthalmic surgery and delicate tissues. The product is supplied sterile by ethylene oxide, single-use and cannot be re-sterilised.
3. Technical Specifications
3.1. General characteristics of reference P1L12HR2610
| Characteristic | Specification |
| Material | Polyglactin 910 (PGLA), copolymer 90 % glycolide / 10 % L-lactide |
| Thread structure | Braided multifilament with grooved finish |
| Quality / range | Prime (Premisut range) |
| Gauge | USP 0 (metric 3.5) |
| Thread colour | Violet (also available colourless in the family) |
| Needle shape | ½ circle |
| Tip type | Conical point (taperpoint), cylindrical body |
| Needle size | 26 mm (HR 26 Prime) |
| Thread length | 75 cm |
| Absorption mechanism | Hydrolysis (56-72 days) |
| Sterilisation | Ethylene oxide (EO), sterile, single-use |
| Tissue reaction | Minimal, with excellent tensile strength and knot security |
| Indications | Tissue approximation, ligatures and ophthalmic surgery |
| Presentation | Box of 12 units |
| Minimum order | 30 boxes |
| Reference (SKU) | P1L12HR2610 |
3.2. Tensile strength retention profile
| Time after implantation | Retained tensile strength |
| 2 weeks | 75 % |
| 3 weeks | 50 % |
| 4 weeks | 30 % |
| Total absorption | 56-72 days by hydrolysis |
Complete absorption by hydrolysis between 56 and 72 days, with minimal tissue reaction, excellent tensile strength and optimal knot tying, with a fully predictable absorption rate.
4. Equivalences and Cross-References
This multifilament polyglactin 910 suture is clinically positioned in the same family as the reference synthetic absorbable braided sutures on the international market, commonly used in general surgery, ophthalmology, gynaecology and urology:
| Manufacturer or commercial house | Equivalent line |
| Ethicon (Johnson & Johnson) | Vicryl (coated braided polyglactin 910, violet colour) |
| B. Braun | Safil (braided polyglactin 910, violet colour) |
| Medtronic / Covidien | Polysorb (absorbable braided range of polyglactin 910) |
| SMI | Surgicryl (absorbable polyglactin 910) |
| Assut Sutures | Assucryl 910 (polyglactin 910) |
| DemeTECH | Vitacryl (coated braided polyglactin 910) |
Within the Cahnos catalogue, its direct equivalent in structure and needle within the polyglycolic acid (PGA) family is reference AMP112HR2610 (cylindrical HR 26 mm needle, USP 0, violet, 75 cm, absorption 60 to 90 days), recommended when a somewhat longer resorption profile is preferred. The colourless version of the same configuration corresponds to reference P1L28HR2610 (colourless thread of 70 cm). *The brands cited belong to their respective manufacturers and are mentioned only as a reference for equivalence.
5. Selection tips and variants of interest
To choose the most appropriate suture, consider the wound support time required by the tissue and the necessary gauge according to the anatomical area. USP 0 gauge (metric 3.5) is indicated for tissues with moderate tension: musculature, fascia, subcutaneous tissue, vascular ligatures and precision ophthalmic surgery. The HR 26 mm conical point needle is the most atraumatic option for soft and friable tissues; if you need greater penetration in firm tissues, consider reverse cutting or trocar points.
| Reference | Variant |
| P1L12HR2610 | PGLA Prime USP 0, cylindrical HR 26 mm needle, violet, 75 cm (current product) |
| P1L12HR2630 | PGLA Prime USP 3/0, HR 26 mm needle, violet, 75 cm (finer) |
| P1L12HR2620 | PGLA Prime USP 2/0, HR 26 mm needle, violet, 75 cm (smaller gauge) |
| P1L12HR2601 | PGLA Prime USP 1, HR 26 mm needle, violet, 75 cm (greater strength) |
| P1L12HR2602 | PGLA Prime USP 2, HR 26 mm needle, violet, 75 cm (high strength) |
| P1L12HR2620.2 | PGLA Prime USP 2/0, HR 26 mm needle, violet, 2×75 cm (double format) |
| P1L18HS2610 | PGLA Prime USP 0, triangular reverse cutting needle HS 26 mm, violet, 70 cm (firm tissues) |
| AMP112HR2610 | PGA Prime USP 0, HR 26 mm needle, violet, 75 cm (absorption 60-90 days) |
If you need a longer strength retention (180-210 days) or a monofilament profile, consider the polydioxanone (PDO) family from Cahnos. All references are supplied in boxes of 12 units, with a minimum order of 30 boxes to guarantee continuous supply to your healthcare centre. As it is a sterile single-use product, check the integrity of the packaging before each procedure, respect the expiry date and store it in a cool, dry place, protected from sunlight. For special B2B conditions by volume or equivalences from other commercial houses, contact our technical-commercial team.





Reviews