Description
1. Product Presentation
The absorbable multifilament polyglactin 910 (PGLA) suture from the Prime range, reference P1L18HR2640VB, is the reference technical solution for professionals who demand secure closure, reliable knotting and fully predictable absorption. It features a ½ circle needle with a cylindrical body and conical point (taperpoint) of 26 mm in VisiBlack finish, violet thread of 70 cm and USP gauge 4/0 (metric 1.5). It is supplied in a box of 12 units, with a minimum order of 30 boxes, and is a preferred option for the approximation of soft tissues, ligatures and ophthalmic surgery. Synthetic absorbable medical device, sterile and single-use, with an excellent quality-price ratio for B2B supply to operating theatres, ophthalmic clinics and purchasing centres.
2. Technical explanation
Polyglactin 910 is a synthetic absorbable copolymer composed of 90% glycolide and 10% L-lactide, presented in a braided multifilament structure without grooves. This configuration provides a regular surface that favours the sliding of the thread through the tissue, smooth handling and firm, secure knot tying, even in fine gauges such as USP 4/0.
Absorption occurs by hydrolysis, i.e., by degradation of the polymer in contact with organic fluids, over a predictable period of 56 to 72 days. This profile maintains the necessary tensile strength during the first weeks of healing (75% at 2 weeks, 50% at 3 and 30% at 4), the precise time for the tissue to recover its own strength.
The ½ circle needle with cylindrical body and conical point (taperpoint geometry) with VisiBlack finish combines an atraumatic point with maximum visibility on the surgical field, allowing it to pass through soft tissues of moderate density with the least possible tissue tearing. It is the usual configuration in ophthalmic surgery and in the approximation of delicate tissues. The product is sterilised by ethylene oxide, is single-use and does not allow re-sterilisation.
3. Technical Specifications
3.1. General characteristics of reference P1L18HR2640VB
| Characteristic | Specification |
| Material | Polyglactin 910 (PGLA), copolymer 90 % glycolide / 10 % L-lactide |
| Thread structure | Non-grooved braided multifilament |
| Quality | Prime (Premisut range) |
| Gauge | USP 4/0 (metric 1.5) |
| Thread colour | Violet |
| Needle shape | 1/2 circle |
| Needle body | Cylindrical |
| Tip type | Taperpoint |
| Needle size | 26 mm (HR 26) with VisiBlack finish |
| Thread length | 70 cm |
| Absorption mechanism | Hydrolysis (56–72 days) |
| Sterilisation | Ethylene oxide (EO), single-use |
| Tissue reaction | Minimal |
| Indications | Tissue approximation, ligatures and ophthalmic surgery |
| Presentation | Box of 12 units |
| Minimum order | 30 boxes |
| Reference (SKU) | P1L18HR2640VB |
3.2. Tensile strength retention profile
| Period | Tensile strength retention |
| 2 weeks | 75 % |
| 3 weeks | 50 % |
| 4 weeks | 30 % |
Complete absorption by hydrolysis between 56 and 72 days. Minimal tissue reaction, excellent tensile strength and optimal knot tying, with a fully predictable absorption rate. Price per box of 12 units.
4. Equivalences and Cross-References
This multifilament polyglactin 910 suture is clinically positioned in the same family as the reference synthetic absorbable braided sutures on the international market, commonly used in general surgery, gynaecology, urology and specialised surgery:
| Manufacturer / Company | Equivalent line |
| Ethicon (Johnson & Johnson) | Vicryl (coated braided polyglactin 910, violet colour) |
| B. Braun | Safil (braided polyglactin 910, violet colour) |
| Medtronic (Covidien) | Polysorb (absorbable braided range of polyglactin 910) |
| SMI | Surgicryl (braided polyglactin 910) |
| Assut | Assucryl 910 (polyglactin 910) |
| DemeTECH | Vitacryl (coated braided polyglactin 910) |
Within the Cahnos catalogue, its closest equivalent is the standard quality reference 1L18HR2640 (same HR 26 mm needle configuration, gauge 4/0 and 70 cm thread, without VisiBlack needle), and in the polyglycolic acid (PGA) family the reference AMP118HR2640VB, with an absorption profile of 60 to 90 days.
5. Selection tips and variants of interest
To choose the most appropriate suture, consider the wound support time required by the tissue to be sutured and the necessary gauge according to the anatomical area. USP gauge 4/0 (metric 1.5) is indicated for fine, low-tension tissues: skin, mucosa, subcutaneous tissue and delicate planes where aesthetic healing and minimal trauma are sought. The 26 mm conical point needle is especially recommended for delicate tissues, such as ophthalmic surgery, and its VisiBlack finish improves contrast on the surgical field.
Variants of interest in the Cahnos catalogue:
- P1L18HR2630VB: same HR 26 mm VisiBlack needle, USP gauge 3/0, violet, 70 cm (larger gauge for moderate tension).
- P1L18HR2620VB: same HR 26 mm VisiBlack needle, USP gauge 2/0, violet, 70 cm (greater strength).
- P1L18HR2240VB: same gauge 4/0 with HR 22 mm VisiBlack needle (shorter needle for areas with limited access).
- P1L18HRT2630: PGLA Prime trocar point (tapercut) HRT 26 mm, USP 3/0 (slightly cutting point for denser tissues).
If you need a longer retention profile (180–210 days) or a monofilament thread, the polydioxanone (PDO) family is the recommended alternative. All references are supplied in boxes of 12 units; the minimum order is 30 boxes to ensure continuous supply to your healthcare facility. Keep the original packaging in a dry, cool place, respect the expiry date and handle the thread with gloves to avoid altering its mechanical properties. For special B2B conditions and volume orders, contact our technical-commercial team.