Description
1. Introduction
The Prime absorbable multifilament polyglactin 910 (PGLA) suture from the Premisut range is the reference solution in the operating theatre for professionals seeking reliable, predictable, high-quality performance in soft tissue approximation, ligatures and ophthalmic surgery. Its ½ circle needle with reverse cutting triangular body and point (HS) of 40 mm provides precise penetration even in firm or resistant tissues, while the 70 cm violet braided thread ensures excellent handling and optimal visibility of the surgical field.
Manufactured under strict quality standards, it combines minimal tissue reaction with excellent tensile strength and optimal knot security, ensuring consistency and clinical results in every procedure. Commercial presentation in a box of 12 units, with a minimum order of 30 boxes to ensure continuous supply for your healthcare centre.
2. Technical Explanation
Polyglactin 910 is a synthetic absorbable copolymer composed of 90 % glycolide and 10 % L-lactide, configured in a braided multifilament structure with a ribbed finish. This structure provides a regular surface that favours thread sliding through tissue, excellent handling and firm, secure knot tying, with high initial tensile strength.
Absorption occurs by hydrolysis, i.e., by degradation of the polymer in contact with organic fluids, over a predictable period of 56 to 72 days. This tensile strength retention profile makes it ideal for tissues requiring wound support during the first weeks of healing. The ½ circle needle with reverse cutting triangular point allows penetration of firm tissues with controlled trauma, being especially recommended for tissue approximation, ligatures and ophthalmic surgery. The product is supplied sterile by ethylene oxide.
3. Technical Specifications
| Characteristic | Specification |
| Material | Polyglactin 910 (PGLA), copolymer 90 % glycolide / 10 % L-lactide |
| Thread structure | Braided multifilament with grooved finish |
| Gauge | USP 2/0 |
| Colour | Violet (also available colourless) |
| Needle | ½ circle, reverse cutting triangular body and point (HS) |
| Needle size | 40 mm |
| Thread length | 70 cm |
| Absorption mechanism | Hydrolysis (56-72 days) |
| Sterilisation | Ethylene oxide |
| Presentation | Box of 12 units |
| Reference (SKU) | P1L18HS4020 |
Tensile strength retention profile:
| Time | Tensile strength retention |
| 2 weeks | 75 % |
| 3 weeks | 50 % |
| 4 weeks | 30 % |
Complete absorption by hydrolysis between 56 and 72 days. Minimal tissue reaction, excellent tensile strength and optimal knot tying, with a predictable absorption rate. Indicated for tissue approximation, ligatures and ophthalmic surgery.
4. Equivalences and Cross-References
This multifilament polyglactin 910 suture is clinically in the same family as the reference synthetic absorbable braided sutures on the international market (polyglactin 910 type material, equivalent to lines such as Vicryl from Ethicon or Safil from B. Braun), commonly used in general surgery, gynaecology, urology and specialised surgery.
Within the Cahnos catalogue, its direct equivalent in PGA (polyglycolic acid) material, with the same 40 mm reverse cutting triangular needle, gauge 2/0 and 70 cm length, is reference AMP118HS4020, ideal when an absorption profile of 60 to 90 days is preferred. To configure the complete operating theatre range, it is recommended to combine this reference with the following variants of the same family:
| Reference | Variant |
| P1L18HS4020 | PGLA Prime USP 2/0, HS needle 40 mm reverse cutting triangular, violet, 70 cm (current product) |
| P1L13HS4002 | PGLA Prime USP 2, HS 40 mm triangular reverse cutting needle, violet, 90 cm |
| P1L18HR4030 | PGLA Prime USP 3/0, HR needle 40 mm cylindrical, violet, 70 cm |
| P1L18FR4020 | PGLA Prime USP 2/0, FR needle 40 mm cylindrical 5/8, violet, 70 cm |
| AMP118HS4020 | PGA Prime USP 2/0, HS needle 40 mm reverse cutting triangular, violet, 70 cm |
5. Selection tips and variants of interest
To choose the most suitable suture, consider the wound support time required by the tissue to be sutured and the necessary gauge according to the anatomical area. Polyglactin 910 offers an excellent balance between initial strength and predictable absorption for soft tissues with healing of 2 to 4 weeks. The reverse cutting triangular point needle is indicated when firm tissues need to be penetrated without tearing, while the cylindrical (HR) variant is more atraumatic in delicate tissues.
If you need longer strength retention or a monofilament profile, consider the polydioxanone (PDO) family from Cahnos, with absorption of 180 to 210 days. For tissues with faster absorption needs, consult the rapid-absorption polyglactin 910 range or polyglycolic acid (PGA). All references are supplied in a box of 12 units, with a minimum order of 30 boxes to ensure continuous supply for your healthcare centre.





Reviews