Description
1. Commercial Introduction
We make available to operating theatres, general and digestive surgery departments, plastic and reconstructive surgery units, surgical dermatology and purchasing centres the absorbable multifilament PGAR RÁPIDO suture, factory reference P228XDS2430, marketed under SKU AMP228XDS2430. This presentation combines a colourless braided thread of 70 cm in USP 3/0 gauge with a ⅜ circle precision needle of 24 mm (XDS24 profile), reverse cutting triangular point and square body: a format designed for firm, non-distensible tissues where clean penetration, minimal trauma and impeccable aesthetic results are required. Each unit is supplied sterile, ready for immediate use, in a box of 12 individual sachets.
2. Technical explanation of the material and procedure
The PGAR RÁPIDO suture is a synthetic braided multifilament of rapid-absorption polyglycolic acid, coated with polycaprolactone and calcium stearate. The coating reduces friction when passing through tissue (less saw effect), facilitates thread sliding and ensures firm, reproducible knot tying, with minimal tissue reaction. Absorption occurs by controlled hydrolysis, a predictable mechanism independent of enzymatic activity, over a period of 40 to 45 days: the thread retains approximately 50% of its tensile strength at 5 days and loses practically all its resistance at 14 days, a profile designed to support only the initial phase of healing of rapidly consolidating tissues.
The XDS24 needle (24 mm, ⅜ circle curvature) with precision reverse cutting triangular point and square body concentrates the cutting edge on the outer face of the curvature: it passes through skin, dense subcutaneous tissue and aponeurosis with minimal resistance, prevents the cut from widening in the puncture bed and improves the aesthetic result, hence its common use in plastic and cosmetic procedures. Sterilisation by gamma radiation; single-use product, not re-sterilisable.
3. Technical Specifications
3.1. Technical data sheet for reference P228XDS2430
| Parameter | Specification |
|---|---|
| Brand / Quality | PREMISUT RÁPIDO (standard quality; XDS24 precision needle) |
| Factory reference | P228XDS2430 |
| Reference Cahnos (SKU) | AMP228XDS2430 |
| Thread material | 100% rapid-absorption polyglycolic acid (PGAR) |
| Thread structure | Synthetic absorbable braided multifilament |
| Coating | Polycaprolactone and calcium stearate |
| Gauge / size | USP 3/0 (metric EP 2), conforming to USP and European Pharmacopoeia |
| Approximate thread diameter | 0.20 – 0.249 mm |
| Thread length | 70 cm (1 strand / 1 needle per unit) |
| Thread colour | Colourless |
| Needle shape | 3/8 circle (⅜ circular) |
| Tip type | Precision reverse cutting triangular (Reverse Extra Cutting) with square body |
| Needle size | XDS 24 (24 mm) |
| Swaged | 1 needle / 1 thread of 70 cm per unit |
| Absorption mechanism | Hydrolysis |
| Absorption period | 40 – 45 days (rapid absorption) |
| Sterilisation | Gamma radiation; sterile, single-use, not re-sterilisable |
| Tissue reaction | Minimal |
| Indications | Tissue approximation, ligatures, skin closure and general surgery; commonly used in plastic and cosmetic surgery |
| Commercial presentation | Box of 12 units (sterile individual sachets); batch traceability |
| Regulatory standard | Conforms to USP and European Pharmacopoeia (EP) gauges |
| Storage | Below 30 °C, away from moisture and direct sunlight |
3.2. Tensile strength retention profile
| Period after implantation | Tensile strength retention |
|---|---|
| 5 days | 50 % of initial tensile strength |
| 14 days | 0% of the initial tensile strength |
| Complete absorption | 40 – 45 days, by hydrolysis |
3.3. Gauge equivalence according to USP / EP Pharmacopoeia (absorbable suture)
| USP size | Metric measurement (EP) | Approximate thread diameter |
|---|---|---|
| 3/0 (this product) | 2 | 0.20 – 0.249 mm |
| 2/0 | 3 | 0.30 – 0.339 mm |
| 0 | 3.5 | 0.35 – 0.399 mm |
| 1 | 4 | 0.40 – 0.499 mm |
4. Equivalences and Cross-References
Reference P228XDS2430 corresponds to the official nomenclature of the PREMISUT RÁPIDO range (XDS24 profile: ⅜ circle needle of 24 mm with precision reverse cutting triangular point and square body, USP 3/0 gauge, colourless 70 cm thread) and Cahnos markets it without altering characters: no digit should be added, removed or reordered when ordering or bidding. The rapid-absorption braided multifilament category (rapid PGA/polyglactin profile, 40–45 days) has in the VICRYL RAPIDE® line from Ethicon (Johnson & Johnson MedTech) its international clinical reference. The following correspondences have been verified against official catalogue or authorised distribution, both comparable by category and not interchangeable 1:1:
| Manufacturer / Brand | Family / Model | Reference code / SKU | Equivalence level | Technical differences / Remarks |
|---|---|---|---|---|
| Ethicon (J&J MedTech) | Coated VICRYL RAPIDE® (polyglactin 910) colourless multifilament, USP 3/0, 70 cm, ⅜ circle reverse cutting needle | NW2732 | Comparable by category (not interchangeable 1:1) | Confirmed in authorised distribution: multifilament structure, polymer family, USP 3/0 gauge, colourless, length 70 cm, ⅜ curvature and reverse cutting point coincide. The needle profile differs (standard reverse cutting versus precision XDS24 with square body and extra-long cutting edge). Validate technical data sheet/IFU before substituting in tender. |
| Ethicon (J&J MedTech) | Coated VICRYL® (polyglactin 910) colourless multifilament, USP 3/0, 70 cm, PS-2 needle of 19 mm, ⅜ circle, reverse cutting | J427H | Comparable by category (not interchangeable 1:1) | Confirmed in authorised distribution: material, USP 3/0 gauge and length coincide; the absorption profile differs (standard VICRYL® retains tensile strength longer than rapid PGAR), the needle (PS-2 of 19 mm versus XDS24) and the coating. |
| DemeTECH | DemeCRYL™ (polyglactin 910) multifilament, USP 3/0, 24 mm ⅜ circle reverse cutting needle, undyed, 70 cm, box of 12 | G1173024F4P | Comparable by category (not interchangeable 1:1) | The closest in geometry: same needle size (24 mm, ⅜ reverse cutting), gauge and length; the absorption profile differs (standard DemeCRYL 56–70 days versus rapid PGAR 40–45 days) and the needle body. |
There is no equivalence table validated by the manufacturer that matches this specific configuration (PGAR rapid multifilament, XDS24 24 mm precision reverse cutting needle with square body, USP gauge 3/0, colourless 70 cm thread, ref. P228XDS2430) with references from other brands in the exact configuration; consult with the technical department before use in tenders. The homologous families cited (VICRYL RAPIDE® and VICRYL® from Ethicon, Safil Quick from B. Braun and rapid PGA lines from DemeTECH, SMI and Dolphin Sutures) must be checked property by property (material, gauge, curvature, point family, length, number of strands, colour, coating and current CE/MDR regulatory classification) before any substitution. The factory reference of this item does not appear in public regulatory databases (GUDID/EUDAMED) and must be confirmed with the supplier before use in tender documents. The cited brands belong to their respective manufacturers and are mentioned for comparative reference only.
5. Selection tips and variants of interest
Tissue to suture: the USP gauge 3/0 (metric EP 2) with a 24 mm needle is the balance point between fineness and strength for approximation of skin, subcutaneous tissue, mucous membranes and ligation of medium-calibre vessels. The precision reverse cutting tip with square body penetrates firm tissues without tearing the puncture bed, making it especially suitable for plastic and cosmetic surgery: better aesthetic outcome and less trauma. Do not use in tissues requiring prolonged support (fascia, abdominal wall under tension), given its fast absorption profile.
Variants within the PGAR rapid range: same XDS24 configuration in 45 cm thread (AMP221XDS2430) and in USP gauge 2/0 with 70 cm (AMP228XDS2420); if you prefer the standard DS24 needle profile with triangular body and point, AMP222DS2430 is available (USP 3/0, 75 cm, Prime quality). For the same format in polyglactin 910 RAPID Prime, see AMP2L28XDS2430; for polyglecaprone 25 monofilament, AMP328XDS2430.
Alternative absorption profile: if the procedure requires longer wound support, consider standard polyglactin 910 or PGA (56–90 days), PGCL monofilament (90–120 days) or polydioxanone PDO (180–210 days), according to the patient’s healing kinetics.
Sterility verification: product sterile by gamma radiation and single-use; check the integrity of the blister and the expiry date before each procedure, keep the reference P228XDS2430 and lot number for traceability in the operating theatre, and store below 30 °C, away from moisture and direct sunlight. For special B2B volume conditions, hospital supply agreements or validation of equivalents from other commercial brands, contact the technical-commercial team at Cahnos.