Description
1. Commercial Introduction
The absorbable multifilament suture PREMISUT RAPID (PGAR, rapid absorption polyglycolic acid) is the accelerated resorption reference in our catalogue for the professional who needs a braided, flexible thread with early and fully predictable tensile strength loss, without the need for suture removal. This configuration combines a strand of 100 % polyglycolic acid in USP 3/0 gauge, undyed and 45 cm, with a ⅜ circle needle with reverse cutting triangular body and point of 24 mm (DS24 profile): firm penetration in skin and low-elasticity tissues, with full control of the path and excellent aesthetic result.
Absorbable synthetic medical device, sterile, single-use and non-resterilisable, presented in a box of 12 units in individual sachets with batch traceability. Aimed at general and outpatient surgery operating theatres, emergency departments, purchasing centres, clinics and tenders looking for a high-performance rapid absorption thread with excellent value for money in the B2B channel.
2. Technical explanation
The PGAR suture (rapid profile polyglycolic acid, low molecular weight) is a synthetic polymer 100 % absorbable by hydrolysis, spun into a braided multifilament structure and coated with polycaprolactone and calcium stearate. This coating reduces friction when passing through tissues (less saw effect), improves thread glide and facilitates firm, secure knotting with minimal tissue reaction.
Its degradation profile is designed for short-term healing: it retains 50 % of tensile strength at 5 days and loses virtually all its strength at 14 days, completing total absorption by hydrolysis between 40 and 45 days, in a predictable manner independent of the patient’s enzymatic activity. The DS24 needle features a ⅜ circle curvature (135°) with reverse cutting triangular body and point: the cutting edge is oriented outward from the wound, creating a clean entry hole that preserves the skin edge. The undyed thread offers maximum discretion on the surgical field. Gamma radiation sterilisation ensures final sterility with minimal alteration of mechanical properties.
3. Technical Specifications
3.1. Technical data sheet for reference 221DS2430
| Parameter | Specification |
| Factory reference | 221DS2430 (standard range) |
| Reference Cahnos (SKU) | AM221DS2430 |
| Brand / family | PREMISUT RAPID – PGAR (rapid polyglycolic acid) |
| Thread material | 100 % polyglycolic acid (PGAR), braided multifilament |
| Coating | Polycaprolactone and calcium stearate |
| Gauge / size | USP 3/0 (metric 2 / EP 2), according to USP/EP Pharmacopoeia |
| Approximate thread diameter | 0.20 – 0.249 mm |
| Thread length | 45 cm (single strand, without loop) |
| Thread colour | Colourless |
| Needle shape | ⅜ circle (3/8 circular, 135°) |
| Needle tip | Triangular – reverse cutting, body and point |
| Needle size | DS24 (24 mm), 1 needle per thread |
| Swaged | 1 needle / 1 suture per unit |
| Absorption mechanism | Hydrolysis; complete between 40 and 45 days |
| Sterilisation | Gamma radiation; sterile, single-use, not re-sterilisable |
| Tissue reaction | Minimal; excellent tensile strength and knot tying |
| Indications | Tissue approximation, ligatures and general surgery; skin and firm tissues |
| Commercial presentation | Box of 12 units (sterile individual sachets); batch traceability |
| Regulatory standard | Conforming to USP/EP Pharmacopoeia for synthetic absorbable sutures; CE marking (MDR) |
3.2. Tensile strength retention profile
| Post-implantation time | Retained tensile strength |
| 5 days | 50 % of initial tensile strength |
| 14 days | 0% (total loss of tensile support) |
| Total absorption | 40 – 45 days, by hydrolysis (predictable and uniform rate) |
The cited retention profile corresponds to the rapid absorption formulation (RAPID line); must not be confused with standard absorption polyglycolic acid (PGA line, 60–90 days) nor with standard polyglactin 910 (PGLA, 56–72 days).
3.3. Gauge equivalence according to USP/EP Pharmacopoeia (synthetic absorbable suture)
| USP size | Metric measurement (EP) | Approximate thread diameter |
| 3/0 (this product) | 2 | 0.20 – 0.249 mm |
| 2/0 | 3 | 0.30 – 0.339 mm |
| 0 | 3.5 | 0.35 – 0.399 mm |
| 1 | 4 | 0.40 – 0.499 mm |
The strand diameter complies with the official USP Pharmacopoeia ranges for synthetic absorbable sutures, without extending its limits.
3.4. Decoding of the official reference
Official reference 221DS2430: 221 (45 cm thread) + DS24 (⅜ circle needle with reverse cutting triangular point of 24 mm) + 30 (USP 3/0 gauge). Cahnos markets the official reference without altering characters: no digit should be added, removed or reordered when ordering or tendering.
4. Equivalences and Cross-References
Reference 221DS2430 corresponds to the official nomenclature of the PREMISUT RAPID range (rapid absorption polyglycolic acid, multifilament, USP 3/0, undyed 45 cm thread, ⅜ circle 24 mm needle with reverse cutting triangular point). Rapid absorption polyglycolic acid (PGAR) is one of the most widespread families of short-term braided absorbable suture on the market, comparable at clinical category level with the rapid absorption lines of major manufacturers. Only references whose codes and attributes have been verified in the manufacturer’s official catalogue and authorised distribution are listed:
| Manufacturer / Brand | Family / Model | Reference code / SKU | Equivalence level | Technical differences / Remarks |
| DemeTECH | DemeCRYL™ (polyglactin 910) multifilament, USP 3/0, 24 mm ⅜ circle reverse cutting needle, undyed, 70 cm, box of 12 | G1173024F4P | Comparable by category (not interchangeable 1:1) | The closest in geometry: they match in needle size (24 mm), ⅜ circle curvature, reverse cutting point family, USP 3/0 gauge and undyed colour; they differ in base polymer (polyglactin 910 vs polyglycolic acid), absorption profile (DemeCRYL standard 56–70 days vs PGAR rapid 40–45 days) and strand length (70 cm vs 45 cm). Code confirmed in official catalogue and authorised distribution. |
| B. Braun / Aesculap | Safil Quick (rapid absorption PGA), braided multifilament, USP 3/0, 45 cm, ⅜ circle reverse cutting | C1049630 | Comparable by category (not interchangeable 1:1) | They match in USP 3/0 gauge, thread length (45 cm), ⅜ circle curvature and reverse cutting point family; they differ in needle size (19 mm vs 24 mm of this reference) and braid coating. Code verified in authorised distribution; check technical data sheet/IFU before substituting in a tender. |
The international clinical benchmark in this segment, VICRYL RAPIDE® from Ethicon (Johnson & Johnson MedTech, rapidly absorbing polyglactin 910), is comparable by category in terms of material and absorption profile; its closest documented code in geometry (VR2252, USP 3/0 with a 24 mm FS-1 needle in a ⅜ circle) corresponds to a 75 cm strand presentation, so it is not listed as an interchangeable candidate for this 45 cm strand. The remaining equivalent ranges (Novosyn Quick from B. Braun, Surgicryl Fast Absorb from SMI, Velosorb Fast from Medtronic/Covidien) are pending code confirmation with each manufacturer for the exact configuration USP 3/0 · 45 cm · 24 mm reverse cutting needle and are not published as closed data. Standard VICRYL (56–72 day absorption) is not comparable to this RAPID version. The brands cited belong to their respective manufacturers and are mentioned solely as comparative reference.
There is no equivalence table validated by the manufacturer; consult the technical department before use in tenders. Within the PREMISUT system itself, the PGAR RAPID family offers the same material and absorption profile in other gauges, geometries and lengths (see block 5), which are fully interchangeable with each other provided the gauge and needle geometry are respected.
5. Selection tips and variants of interest
The USP 3/0 gauge (metric 2) offers an ideal balance between initial strength and handling, indicated for tissues under moderate tension: skin, subcutaneous tissue, mucosa and soft tissue in general, as well as for ligation of small-calibre vessels. The DS24 needle, ⅜ circle with a triangular reverse cutting point, is the recommended option when firm penetration is required in tissues with low elasticity, with minimal tearing of the puncture bed and a better aesthetic result. The undyed colour facilitates direct visualisation of the surgical field and is preferred in aesthetic areas or superficial tissue.
| Variant / Reference | Description |
| AM221DS2430 (221DS2430) | PGAR RAPID – DS24 reverse cutting 3/8 – USP 3/0 undyed – 45 cm (this product) |
| AMP221DS2430 (P221DS2430) | PGAR RAPID Prime – DS24 – USP 3/0 undyed – 45 cm (same configuration, premium range) |
| AM222DS2430 (222DS2430) | PGAR RAPID – DS24 – USP 3/0 undyed – 75 cm (same geometry, longer strand) |
| AMP222DS2430 (P222DS2430) | PGAR RAPID Prime – DS24 – USP 3/0 undyed – 75 cm (Prime range, 75 cm strand) |
| AM222DS2420 (222DS2420) | PGAR RAPID – DS24 – USP 2/0 undyed – 75 cm (same needle, larger gauge for more tension) |
| AM222DS2630 (222DS2630) | PGAR RAPID – DS26 – USP 3/0 undyed – 75 cm (needle 2 mm longer, medium-depth planes) |
If the protocol requires longer tensile support, consider the PGA families (standard polyglycolic acid, 60–90 days), PGLA standard polyglactin 910 (56–72 days), PGCL monofilament (90–120 days) or PDO polydioxanone (180–210 days), depending on the patient’s healing kinetics. Sterility and storage: sterile product by gamma radiation and single-use; check the integrity of the pouch and the expiry date before each use, keep reference 221DS2430 together with the batch number for traceability in the operating theatre and store below 30 °C, away from moisture and direct sunlight. For B2B volume conditions, hospital supply agreements or validation of equivalents from other commercial houses, contact the technical-commercial team at Cahnos.