Description
1. Commercial Introduction
The multifilament absorbable suture PREMISUT RAPID Prime (polyglactic acid PGAR) is the accelerated-absorption closure solution for general and ambulatory surgery theatres, emergency departments, digestive surgery, gynaecology and B2B purchasing centres that require a braided, flexible thread with early and fully predictable loss of tensile strength. This reference combines a 75 cm colourless strand in USP 3/0 gauge with a ⅜ circle needle with reverse cutting triangular point, 26 mm (DS26 Prime profile): clean penetration through skin and firm tissues with excellent aesthetic results and minimal tissue reaction. Sterile, ready-to-use product, supplied in a box of 12 units in individual sachets with batch traceability; intended for hospital replenishment, clinics, distribution and tenders.
2. Technical Explanation
Polyglactic acid (PGAR) is a 100% absorbable synthetic polymer, spun into a braided multifilament structure and coated with polycaprolactone and calcium stearate: this coating reduces friction when passing through tissues, improves thread glide and facilitates firm, secure knotting, with minimal tissue reaction. In its RAPID variant (low molecular weight), absorption occurs by hydrolysis —a predictable process independent of enzymatic activity— over a period of 40 to 45 days: it retains approximately 50% of the initial tensile strength at 5 days and loses it completely at 14 days, making it ideal for rapidly healing tissues.
The DS26 needle features a ⅜ circle curvature with a reverse cutting triangular point: the cutting edge is oriented towards the outer face of the curvature, so it cuts in the opposite direction to the wound bed, preventing the cut from widening and improving the aesthetic outcome of the closure. Its 26 mm length is suitable for medium-depth planes, skin and subcutaneous tissue. Gamma radiation sterilisation ensures final sterility with minimal alteration of mechanical properties; single-use product, not resterilisable.
3. Technical Specifications
3.1. Technical data sheet of the reference
| Parameter | Specification |
|---|---|
| Factory reference | P222DS2630 (Prime range) |
| Reference Cahnos (SKU) | AMP222DS2630 |
| Brand / quality | PREMISUT RAPID – PGAR Prime |
| Thread material | 100 % polyglactic acid (PGAR), rapid absorption |
| Thread structure | Braided multifilament, smooth |
| Coating | Polycaprolactone and calcium stearate |
| Gauge / size | USP 3/0 (metric 2 / EP 2), USP/EP Pharmacopoeia |
| Approximate thread diameter | 0.20 – 0.249 mm |
| Thread length | 75 cm (free end, without loop) |
| Thread colour | Colourless (violet on demand) |
| Needle shape | ⅜ circle (3/8 circular, 135°) |
| Needle tip | Triangular reverse cutting (Reverse Cutting) |
| Needle size | DS 26 mm (Prime), 1 needle per thread |
| Swaged | 1 needle / 1 suture per unit |
| Absorption | Hydrolysis; complete between 40 and 45 days |
| Sterilisation | Gamma radiation; sterile, single-use, not re-sterilisable |
| Tissue reaction | Minimal; excellent tensile strength and knot tying |
| Indications | Tissue approximation, ligatures and general surgery |
| Commercial format | Box of 12 units (individual sterile sachets); minimum order 30 boxes |
| Regulatory standard | Conforming to USP/EP Pharmacopoeia for synthetic absorbable sutures; CE marking (MDR) |
3.2. Decoding of the official reference
The official reference P222DS2630 breaks down as follows: P (Prime range) + 222 (75 cm thread) + DS26 (⅜ circle needle with 26 mm reverse cutting triangular point) + 30 (USP 3/0 gauge). Cahnos markets the official reference without altering characters: no digit may be added, removed or reordered when ordering or tendering.
3.3. Tensile strength retention profile
| Post-implantation time | Retained tensile strength |
|---|---|
| 5 days | 50 % of initial tensile strength |
| 14 days | 0% (total loss of tensile support) |
| Total absorption | 40 – 45 days, by hydrolysis (predictable and uniform rate) |
The retention profile corresponds to the rapid-absorption formulation (RAPID line); it must not be confused with standard-absorption polyglactic acid (60–90 days) or with standard polyglactin 910 (56–72 days), also available in our catalogue.
3.4. Gauge equivalence according to USP/EP Pharmacopoeia (synthetic absorbable suture)
| USP size | Metric measurement (EP) | Approximate thread diameter |
|---|---|---|
| 3/0 (this product) | 2 | 0.20 – 0.249 mm |
| 2/0 | 3 | 0.30 – 0.339 mm |
| 0 | 3.5 | 0.35 – 0.399 mm |
The strand diameter complies with the official USP Pharmacopoeia ranges for synthetic absorbable sutures, without extending their limits.
4. Equivalences and Cross-References
Reference P222DS2630 corresponds to the official nomenclature of the PREMISUT RAPID range (rapid-absorption polyglactic acid, multifilament, USP 3/0, 75 cm colourless thread, ⅜ circle needle with 26 mm reverse cutting triangular point). Rapid-absorption polyglactic acid (PGAR) belongs to the worldwide family of short-duration braided multifilament absorbable sutures, whose international clinical benchmark is the VICRYL RAPIDE line (polyglactin 910). Only the reference whose full code and attributes have been verified against the manufacturer’s official catalogue (Ethicon EMEA) and authorised distribution is included:
| Manufacturer / Brand | Family / Model | Reference code / SKU | Equivalence level | Technical differences / Remarks |
|---|---|---|---|---|
| Ethicon (Johnson & Johnson MedTech) | VICRYL RAPIDE (polyglactin 910, braided, colourless) | W9932 – USP 3/0, 75 cm, colourless, Prime reverse cutting needle of 26 mm, ⅜ circle, box of 12 | Comparable by category (not interchangeable 1:1) | They match: short-duration absorbable material (≈42 days), USP 3/0 gauge, colourless, 75 cm length, ⅜ circle geometry, 26 mm reverse cutting needle and box format of 12 units. They differ in the base polymer (polyglactin 910 vs. polyglactic acid), the exact braid coating and the needle nomenclature (PS vs. DS26). Code confirmed in Ethicon’s EMEA catalogue and authorised distributors; verify the absorption profile in the official datasheet before substituting in a tender. |
The remaining equivalent lines in the rapid-absorption segment —Safil Quick and NOVOSYN QUICK (B. Braun / Aesculap), DemeCRYL Fast / DemeQUICK (DemeTECH) and Surgicryl Fast Absorb (SMI)— are comparable by category in terms of material and absorption profile, but must be reviewed property by property (material, USP 3/0 gauge, multifilament structure, ⅜ circle curvature, reverse cutting point family, 26 mm needle length, 75 cm strand length, colourless colour and CE/MDR regulatory classification) before any substitution; their codes for this exact configuration are pending confirmation with each manufacturer and are not published as closed data. Standard VICRYL (56–72 day absorption) is not comparable to this RAPID version. The brands cited belong to their respective manufacturers and are mentioned solely as comparative reference. There is no manufacturer-validated equivalence table; consult the technical department before use in a tender. Within the PREMISUT system itself, the PGAR RAPID family offers the same material and absorption profile in other gauges, geometries and lengths (see block 5), which are fully interchangeable with each other provided gauge and needle geometry are respected.
5. Selection tips and variants of interest
USP 3/0 gauge (metric 2) offers an optimal balance between strength and fineness, indicated for tissues under moderate tension: skin, subcutaneous tissue, mucosa and soft tissue in general, as well as for ligation of small-calibre vessels, with an early loss of strength that in many indications avoids the need for suture removal. The DS26 needle with ⅜ circle and reverse cutting point is the recommended option when firm penetration is required with preservation of the skin edge and better aesthetic results, particularly valued in general surgery and skin closures.
| SKU | Variant |
|---|---|
| AM222DS2630 (222DS2630) | PGAR RAPID DS26 – USP 3/0 colourless – 75 cm (same configuration, standard quality; direct equivalent within the same manufacturer) |
| AMP222DS2620 (P222DS2620) | PGAR RAPID Prime DS26 – USP 2/0 undyed – 75 cm (larger gauge for more tension) |
| AMP222DS2430 (P222DS2430) | PGAR RAPID Prime DS24 reverse cutting – USP 3/0 undyed – 75 cm (needle 2 mm shorter) |
| AMP222DS1630 (P222DS1630) | PGAR RAPID Prime DS16 reverse cutting – USP 3/0 undyed – 75 cm (short needle, superficial planes) |
| AMP2L22DS2630 (P2L22DS2630) | PGLAR Polyglactin 910 RAPID Prime DS26 – USP 3/0 undyed – 75 cm (same configuration in polyglactin 910) |
If the protocol prefers even more atraumatic penetration in lax tissues, consider the conical point series (HR / taperpoint) from the same range; for deep planes with a wider field, consult the HRT36/HRT40 needles. The undyed version offers maximum aesthetic discretion on the surgical field; if greater visibility is required, request the same configuration in dyed thread (violet) on demand. Sterility and storage: product sterile by gamma radiation and single-use; check the integrity of the pouch and the expiry date before each use, keep the reference P222DS2630 together with the batch number for traceability in the operating theatre and store below 30 °C, away from moisture and direct sunlight. For special B2B conditions by volume or hospital supply agreements, contact the technical-commercial team at Cahnos.





Reviews
There are no reviews yet.