Description
1. Commercial Introduction
The multifilament absorbable suture PREMISUT RÁPIDO (rapid-profile polyglycolic acid, PGAR) is the accelerated absorption reference in our catalogue for the professional who needs a resorbable thread with predictable handling, high knot security and complete disappearance from the body within a few weeks, without the need for removal. This configuration combines a braided strand of 100 % polyglycolic acid in USP 3/0 gauge, colourless and 75 cm, with a ⅜ circle needle with reverse cutting triangular body and point of 24 mm (DS24 profile), the ideal combination for penetrating firm, inelastic tissues with total control of the trajectory and excellent aesthetic result.
Synthetic absorbable medical device, sterile, single-use and non-resterilisable, supplied in a box of 12 units in individual sachets with batch traceability. Intended for general and ambulatory surgery operating theatres, emergency departments, purchasing centres, clinics and tenders seeking a high-performance rapid-absorption thread with an excellent price-quality ratio in the B2B channel.
2. Technical explanation
The PGAR suture (rapid-profile, low-molecular-weight polyglycolic acid) is a synthetic polymer 100 % absorbable by hydrolysis, spun into a braided multifilament structure and coated with polycaprolactone and calcium stearate. This coating reduces friction when passing through tissues (less saw effect), improves thread sliding and facilitates firm, secure knotting, with minimal tissue reaction.
Its degradation profile is designed for short-duration healing: it retains 50 % of tensile strength at 5 days and loses virtually all its resistance at 14 days, completing total absorption by hydrolysis between 40 and 45 days, predictably and independently of the patient’s enzymatic activity. The DS24 needle has a ⅜ circle curvature (135°) with triangular reverse cutting body and point (inverted triangle / reverse cutting): the cutting edge is oriented away from the wound, creating a clean entry hole that preserves the skin edge, with firm penetration into skin, mucosa, superficial fascia and dense subcutaneous tissue. The colourless thread offers maximum discretion on the surgical field, preferred in aesthetic areas or superficial tissue. Gamma radiation sterilisation guarantees final sterility with minimal alteration of mechanical properties.
3. Technical Specifications
3.1. Technical data sheet for reference 222DS2430
| Parameter | Specification |
| Factory reference | 222DS2430 (standard range) |
| Reference Cahnos (SKU) | AM222DS2430 |
| Brand / family | PREMISUT RAPID – PGAR (rapid polyglycolic acid) |
| Thread material | 100 % polyglycolic acid (PGAR), braided multifilament |
| Thread structure | Coated braided multifilament, smooth finish |
| Coating | Polycaprolactone and calcium stearate |
| Gauge / size | USP 3/0 (metric 2 / EP 2), according to USP/EP Pharmacopoeia |
| Approximate thread diameter | 0.20 – 0.249 mm |
| Thread length | 75 cm (single strand, no loop) |
| Thread colour | Colourless |
| Needle shape | ⅜ circle (3/8 circular, 135°) |
| Needle tip | Triangular – reverse cutting, body and point |
| Needle size | DS24 (24 mm), 1 needle per thread |
| Swaged | 1 needle / 1 suture per unit |
| Absorption mechanism | Hydrolysis; complete between 40 and 45 days |
| Residual tensile strength | 50% at 5 days; 0% at 14 days |
| Sterilisation | Gamma radiation; sterile, single-use, not re-sterilisable |
| Tissue reaction | Minimal; excellent tensile strength and knot tying |
| Indications | Tissue approximation, ligatures and general surgery; skin and firm tissues |
| Commercial presentation | Box of 12 units (sterile individual sachets); batch traceability |
| Regulatory standard | Conforming to USP/EP Pharmacopoeia for synthetic absorbable sutures; CE marking (MDR) |
3.2. Tensile strength retention profile
| Post-implantation time | Retained tensile strength |
| 5 days | 50 % of initial tensile strength |
| 14 days | 0% (total loss of tensile support) |
| Total absorption | 40 – 45 days, by hydrolysis (predictable and uniform rate) |
The cited retention profile corresponds to the rapid absorption formulation (RÁPIDO line); it must not be confused with standard absorption polyglycolic acid (PGA line, 60–90 days) or with standard polyglactin 910 (PGLA, 56–72 days).
3.3. Gauge equivalence according to USP/EP Pharmacopoeia (synthetic absorbable suture)
| USP size | Metric measurement (EP) | Approximate thread diameter |
| 3/0 (this product) | 2 | 0.20 – 0.249 mm |
| 2/0 | 3 | 0.30 – 0.339 mm |
| 0 | 3.5 | 0.35 – 0.399 mm |
| 1 | 4 | 0.40 – 0.499 mm |
The strand diameter complies with the official USP Pharmacopoeia ranges for synthetic absorbable sutures, without exceeding their limits.
3.4. Decoding of the official reference
Official reference 222DS2430: 222 (75 cm thread) + DS24 (⅜ circle needle with 24 mm reverse cutting triangular point) + 30 (USP 3/0 gauge). Cahnos markets the official reference without altering characters: no digit may be added, removed or reordered when ordering or tendering.
4. Equivalences and Cross-References
Reference 222DS2430 corresponds to the official nomenclature of the PREMISUT RÁPIDO range (rapid absorption polyglycolic acid, multifilament, USP 3/0, colourless 75 cm thread, ⅜ circle needle of 24 mm with reverse cutting triangular point). Rapid absorption polyglycolic acid (PGAR) is one of the most widespread short-duration braided absorbable suture families on the market, comparable at clinical category level with the rapid absorption lines of major manufacturers. Only references whose codes and attributes have been verified in the manufacturer’s official catalogue and authorised distribution are listed:
| Manufacturer / Brand | Family / Model | Reference code / SKU | Equivalence level | Technical differences / Remarks |
| Ethicon (Johnson & Johnson MedTech) | VICRYL RAPIDE (polyglactin 910, braided, colourless) | VR2252 – USP 3/0, 75 cm thread, FS-1 24 mm needle in ⅜ circle, reverse cutting point, colourless | Comparable by category (not interchangeable 1:1) | They match in USP 3/0 gauge, thread length (75 cm), ⅜ circle curvature, reverse cutting point family and needle size (24 mm); they differ in base polymer (polyglactin 910 90/10 versus polyglycolic acid), coating and needle nomenclature (FS-1 versus DS24). Code confirmed in Ethicon EMEA catalogue and authorised distribution. Validate technical data sheet/IFU before substituting in a tender. |
| DemeTECH | DemeCRYL (polyglactin 910) multifilament, USP 3/0, 24 mm ⅜ circle reverse cutting needle, colourless, 70 cm, box of 12 | G1173024F4P | Comparable by category (not interchangeable 1:1) | The closest in geometry: they match in needle size (24 mm), ⅜ circle curvature, reverse cutting point family, USP 3/0 gauge and colourless colour; they differ in base polymer (polyglactin 910 versus polyglycolic acid), absorption profile (DemeCRYL standard 56–70 days versus rapid PGAR 40–45 days) and strand length (70 cm versus 75 cm). Code confirmed in official catalogue and authorised distribution. |
The remaining homologous ranges in the rapid absorption segment (Safil Quick and Novosyn Quick from B. Braun, Surgicryl Fast Absorb from SMI, Velosorb Fast from Medtronic/Covidien) are comparable by category in terms of material and absorption profile, but their codes for this exact configuration (USP 3/0 · 75 cm · 24 mm reverse cutting needle) are pending confirmation with each manufacturer and are not published as closed data. Standard VICRYL (absorption 56–72 days) is not comparable to this RÁPIDO version. The cited brands belong to their respective manufacturers and are mentioned only as comparative reference.
There is no equivalence table validated by the manufacturer; consult the technical department before using in a tender. Within the PREMISUT system itself, the PGAR RÁPIDO family offers the same material and absorption profile in other gauges, geometries and lengths (see block 5), which are indeed fully interchangeable with each other provided the gauge and needle geometry are respected.
5. Selection tips and variants of interest
The USP 3/0 (metric 2) gauge offers an ideal balance between initial strength and manageability, indicated for tissues under moderate tension: skin, subcutaneous tissue, mucosa and soft tissue in general, as well as for ligation of small-calibre vessels. The DS24 needle of ⅜ circle with a reverse cutting triangular point is the recommended option when firm penetration is required in tissues with low elasticity, with minimal tearing of the puncture bed and a better aesthetic result. The colourless nature facilitates direct visualisation of the surgical field and is preferred in aesthetic areas or superficial tissue.
| Variant / Reference | Description |
| AM222DS2430 (222DS2430) | PGAR RAPID – DS24 reverse cutting 3/8 – USP 3/0 undyed – 75 cm (this product) |
| AMP222DS2430 (P222DS2430) | PGAR RAPID Prime – DS24 – USP 3/0 undyed – 75 cm (premium range, same geometry) |
| AM221DS2430 (221DS2430) | PGAR RAPID – DS24 – USP 3/0 undyed – 45 cm (same configuration, shorter thread) |
| AM222DS2420 (222DS2420) | PGAR RAPID – DS24 – USP 2/0 undyed – 75 cm (same needle, larger gauge for more tension) |
| AM222DS2630 (222DS2630) | PGAR RAPID – DS26 – USP 3/0 undyed – 75 cm (needle 2 mm longer, medium-depth planes) |
| AM2L22DS2430 (2L22DS2430) | PGLAR polyglactin 910 RAPID – DS24 – USP 3/0 undyed – 75 cm (same geometry, copolymer material) |
If the protocol requires a longer tensile support, consider the PGA families (standard polyglycolic acid, 60–90 days), PGLA standard polyglactin 910 (56–72 days), PGCL monofilament (90–120 days) or PDO polydioxanone (180–210 days), depending on the patient’s healing kinetics. Sterility and storage: product sterile by gamma radiation and single-use; check the integrity of the pouch and the expiry date before each use, keep the reference 222DS2430 together with the batch number for traceability in the operating theatre and store below 30 °C, away from humidity and direct sunlight. For B2B volume conditions, hospital supply agreements or validation of equivalents from other commercial houses, contact the technical-commercial team of Cahnos.