Description
1. Product Presentation
The multifilament absorbable suture PREMISUT RAPID Prime (PGAR) is the reference surgical closure solution when the procedure requires short-term wound support, fully predictable absorption and excellent tissue handling. Made from 100 % high-purity polyglactic acid of Prime quality, it offers firm penetration, a secure knot and minimal tissue reaction.
Its half-circle needle with cylindrical body and tapercut point of 26 mm (HRT 26 series) combines the atraumatic penetration of a round body with a sharpened tip that passes through firm or inelastic tissues without tearing or enlarging the exit hole. It is presented in B2B format in a box of 12 units, designed for continuous supply to operating theatres, general surgery departments, clinics and purchasing centres.
2. Technical Explanation
Polyglactic acid (PGAR) is an absorbable synthetic polymer presented in a braided multifilament structure, with a coating of polycaprolactone and calcium stearate that provides smooth passage through tissues, optimal thread glide during knotting and a firm, secure tie.
In its RAPID version (low molecular weight), absorption occurs by hydrolysis, i.e., by chemical degradation in contact with tissue fluids, without depending on the patient’s enzymatic activity, over a period of 40 to 45 days. It retains approximately 50 % of its initial tensile strength at 5 days and loses it completely at 14 days, making it ideal for tissues with rapid healing, tissue approximation, ligatures and general surgery where minimising residual foreign material is desired.
The half-circle needle (160°) with cylindrical body and tapercut point (HRT 26 profile, 26 mm) pierces dense tissues without aggressive cutting effect, avoiding enlargement of the exit hole. The undyed suture offers maximum discretion on the surgical field, a quality especially valued in superficial layers. Sterilisation by gamma radiation guarantees final sterility with minimal alteration of the material’s mechanical properties. Single-use product, not re-sterilisable.
3. Technical Specifications
3.1. Technical data sheet of the reference
| Parameter | Specification |
| Reference (SKU) | AMP228HRT2610 (box of 12 units) |
| Brand / Quality | PREMISUT RAPID – PGAR Prime |
| Material | 100% polyglycolic acid (PGAR) absorbable by hydrolysis |
| Thread structure | Braided multifilament, undyed |
| Gauge | USP 0 (metric 3.5; approximate diameter 0.35–0.399 mm) |
| Thread colour | Colourless |
| Thread length | 75 cm |
| Needle shape | Half circle (1/2, 160°) |
| Needle body | Cylindrical |
| Needle tip | Trocar / Tapercut |
| Needle size | 26 mm (HRT 26 Prime) |
| Absorption mechanism | Hydrolysis (40–45 days) |
| Sterilisation | Gamma radiation – sterile, single-use, not resterilisable |
| Tissue reaction | Minimal; excellent tensile strength and knot security |
| Indications | Tissue approximation, ligatures and general surgery |
| Commercial format | Box of 12 units |
3.2. Tensile strength retention profile
| Period after implantation | Retained tensile strength |
| 5 days | 50 % of initial tensile strength |
| 14 days | 0% of the initial tensile strength |
| Total absorption | 40–45 days, by hydrolysis (predictable and uniform rate) |
3.3. Size equivalence
| USP size | Metric size | Approximate diameter |
| 0 | 3.5 | 0.35 – 0.399 mm |
4. Equivalences and Cross-References
Rapid-absorption polyglactic acid is a standardised category whose international clinical reference is the VICRYL RAPIDE line from Ethicon (J&J MedTech), with which it shares low molecular weight polyglycolic material, accelerated strength loss profile (50 % at 5 days / 0 % at 14 days), half-circle curvature, 26 mm needle, USP size 0 and undyed suture. After verification in public manufacturer sources (Ethicon EMEA catalogue) and authorised distributors, the following cross-reference is included:
| Manufacturer / Brand | Family / Model | Reference code | Equivalence level | Technical differences / Remarks |
| Ethicon (J&J MedTech) | VICRYL RAPIDE, CT-2 needle 26 mm, half circle, taper point, undyed suture, USP 0 | V56H | Comparable by category (not interchangeable 1:1) | They match in material, rapid absorption (≈42 days), USP size 0 (metric 3.5), half-circle curvature, 26 mm needle and undyed braided suture (70 cm). They differ in needle point: taper point (CT-2) versus tapercut point (HRT 26) of this reference, the braid coating and the packaging format (36 units in Ethicon versus 12 units). Code confirmed in Ethicon EMEA catalogue (VICRYL RAPIDE, USP 0, CT-2, undyed). |
Among the homologous families of rapid-absorption polyglycolic without a verified cross-code for this exact configuration are Safil Quick (B. Braun), Polysorb (Medtronic/Covidien), Surgicryl 910 (SMI), Assucryl 910 (Assut) and DemeCRYL (DemeTECH); in all cases, size, point geometry, needle measurement and suture length must be confirmed before any substitution. Standard VICRYL (absorption 56–72 days) is not comparable to the RAPID profile in terms of tensile strength retention.
There is no equivalence table validated by the manufacturer; consult with the technical department before use in a tender.
5. Selection tips and variants of interest
USP size 0 (metric 3.5) offers an optimal balance between strength and handling, indicated for tissues with moderate tension: skin, subcutaneous tissue, mucosa, soft tissue approximation and ligatures of small and medium-calibre vessels. The tapercut point needle is especially recommended when it is necessary to pass through tissues somewhat firmer than with the conventional taper point, without the cutting effect of triangular needles.
| Reference | Variant |
| AM228HRT2610 | PGAR RAPID USP 0, HRT 26 needle, undyed, 75 cm (standard quality, same configuration) |
| AMP2L28HRT2610 | PGLAR RAPID Prime USP 0, HRT 26 needle, undyed, 70 cm (polyglactin 910, same needle geometry) |
| AMP222HRT2620 | PGAR RAPID Prime USP 2/0, HRT 26 needle, undyed, 75 cm (finer size) |
| AMP2L22HRT2630 | PGLAR RAPID Prime USP 3/0, HRT 26 needle, undyed, 75 cm (polyglactin 910, finer size) |
| AM2L23HRT3620 | PGLAR RAPID USP 2/0, HRT 36 needle, undyed, 90 cm (larger needle) |
If the protocol prefers even more atraumatic penetration in lax tissues, consider the taper point series (HR / HRP taperpoint) from the same range; if a wider working field is required, consult HRT 36/40 needles. The undyed version offers maximum discretion in the field; if greater visibility is needed, consult the same configuration in dyed suture on request.
As this is a sterile single-use product, check the packaging integrity and expiry date before each procedure and store the suture in a cool, dry place, protected from sunlight. For special B2B conditions, volume orders or enquiries about additional sizes, lengths and colours, contact the technical-commercial team at Cahnos.





Reviews
There are no reviews yet.