Description
1. Commercial Introduction
The absorbable multifilament polyglycolic acid (PGA) suture from the PREMISUT range, reference 122DS1640 (SKU Cahnos AM122DS1640), is the reference solution for operating theatre services, purchasing centres and distributors seeking a synthetic absorbable thread with predictable behaviour, excellent handling and maximum knot security in tissue approximation, ligatures and general surgery.
This presentation incorporates a ⅜ circle needle with a reverse cutting triangular point (triangular inverted / reverse cutting) of 16 mm (DS16), USP gauge 4/0 (metric 1.5) and a 75 cm colourless fluted thread. Sterile medical device, single-use, ready to use, in a box of 12 units. Tariff reference 122DS1640 · SKU Cahnos AM122DS1640.
2. Technical Explanation
Polyglycolic acid (PGA) is a synthetic absorbable polymer widely used in surgical sutures, presented here in a braided multifilament structure with a fluted finish and a coating of polycaprolactone and calcium stearate. This configuration facilitates the passage of the thread through tissue, reduces the sawing effect, offers smooth sliding and guarantees excellent knot security, even in fine gauges such as USP 4/0 (approximate diameter of 0.15 to 0.199 mm according to USP/EP).
Absorption occurs by hydrolysis, a predictable mechanism independent of the patient’s enzymatic activity: the polymer degrades into glycolic acid, a substance that the body metabolises without generating significant tissue reaction. The tensile strength retention profile (75% at 2 weeks, 50% at 3 weeks and 40% at 4 weeks) maintains the necessary support during the critical phase of healing, while the tissue recovers its own resistance. Complete absorption occurs between 60 and 90 days.
The ⅜ circle needle with a 16 mm reverse cutting triangular point penetrates firm, fibrous and non-compliant tissues (skin, fascia, aponeurosis) with controlled trauma and without tearing, making it especially suitable for skin closures, soft tissue approximation, ligatures and reduced surgical fields. The colourless colour minimises visual interference on the surgical field. Sterilisation by ethylene oxide (EO); does not allow re-sterilisation.
3. Technical Specifications
3.1. General characteristics of reference 122DS1640
| Parameter | Specification |
| Brand / Range | PREMISUT (standard quality) |
| Catalogue reference | 122DS1640 |
| SKU Cahnos | AM122DS1640 |
| Composition | 100 % Polyglycolic Acid (PGA), absorbable by hydrolysis |
| Thread structure | Braided multifilament with grooved finish |
| Coating | Polycaprolactone and calcium stearate |
| Gauge | USP 4/0 (metric 1.5) |
| Approximate diameter | 0.15 – 0.199 mm (indicative according to USP/EP) |
| Thread colour | Colourless (also available in violet in the range) |
| Thread length | 75 cm |
| Needle shape | 3/8 circle (approx. 135°) |
| Tip type | Triangular with reverse cutting (inverted triangular / reverse cutting), body and point |
| Needle size | DS16 (16 mm), 1 needle per thread |
| Absorption mechanism | By hydrolysis, between 60 and 90 days |
| Sterilisation | Ethylene oxide (EO), single-use, not re-sterilisable |
| Tissue reaction | Minimal |
| Indications | Tissue approximation, ligatures and general surgery |
| Presentation | Box of 12 units (individual sterile sachets) |
| Minimum order | 30 boxes |
| Regulatory standard | USP / European Pharmacopoeia (EP) |
3.2. Tensile strength retention profile after implantation
| Time point | Tensile strength retention |
| 2 weeks | 75% of initial strength |
| 3 weeks | 50% of initial strength |
| 4 weeks | 40% of initial strength |
| Complete absorption | 60-90 days by hydrolysis |
Excellent initial tensile strength, optimal knot tying and fully predictable absorption rate, with minimal tissue reaction.
3.3. Size equivalence
| USP size | Metric size | Approximate diameter |
| 4/0 | 1.5 | 0.15 – 0.199 mm (indicative according to USP/EP) |
4. Equivalences and Cross-References
The coated braided multifilament polyglycolic acid (PGA) with medium absorption (60-90 days) corresponds to the homologous families of the main manufacturers on the market: pure PGA is the material of the Safil (B. Braun), Surgicryl (SMI) and DemeSORB (DemeTECH) lines; the Vicryl (Ethicon, polyglactin 910) and Polysorb (Medtronic, Lactomer) ranges are copolymers based on glycolide, comparable by clinical category but not interchangeable 1:1 with the PGA homopolymer. The references cited below have been verified in official catalogues or authorised stores of each manufacturer.
| Manufacturer / Brand | Family / Model | Reference code / SKU | Equivalence level | Technical differences / Remarks |
| SMI (Surgical Materials Industries, Belgium) | Surgicryl® PGA (polyglycolic acid, braided multifilament coated with polycaprolactone and calcium stearate) | 11151516 (DS16, 16 mm) | Comparable by category (not interchangeable 1:1) | Surgicryl PGA USP 4/0 (metric 1.5) with DS16 needle of ⅜ circle and reverse cutting of 16 mm and 75 cm thread; matches in material (100% PGA), coating, gauge, needle and length; differs in colour (violet versus colourless). Configuration confirmed in authorised SMI distributors (presentation in box of 12); verify needle geometry and availability of colourless version before substituting. |
| DemeTECH (USA) | DemeSORB® PGA (glycolide homopolymer, braided coated) | PGA284016F4P (DFS-3, 16 mm) | Comparable by category (not interchangeable 1:1) | DemeSORB PGA USP 4/0, 75 cm (30″) thread and 3/8 circle reverse cutting needle of 16 mm; matches in material (PGA coated with polycaprolactone and calcium stearate), gauge, length and needle size; differ in colour (violet versus colourless), needle designation (DFS-3 versus DS16) and published retention profile (approx. 84% at 2 weeks / 23% at 4 weeks). Configuration confirmed in the official DemeTECH store. |
There is no equivalence table validated by the manufacturer for this specific configuration (multifilament PGA, ⅜ circle DS16 needle of 16 mm, reverse cutting triangular point, USP gauge 4/0, colourless 75 cm thread); consult the technical department of Cahnos before using in tenders. Before substituting this reference in a specification, check one by one: material, USP gauge, needle size (16 mm), ⅜ curvature, point family (reverse cutting), thread length (75 cm) and colour. The copolymer ranges of polyglactin 910 (PGLA), such as Vicryl® from Ethicon, are chemically different from the PGA homopolymer and are not interchangeable 1:1 without validation from the technical department. The brands cited belong to their respective manufacturers and are mentioned only as a reference for equivalence.
5. Selection tips and variants of interest
The USP gauge 4/0 (metric 1.5) is indicated for delicate tissues and moderate tension: soft tissue approximation, skin closures, subcutaneous planes and ligatures of small-calibre vessels. The 16 mm DS16 needle with reverse cutting triangular point is recommended when it is necessary to penetrate firm tissues without tearing in surgical fields of reduced size, paediatric surgery or areas of limited access. The colourless thread offers maximum aesthetic discretion on the surgical field; if maximum visibility is preferred, a violet version of the same configuration is available.
| Reference (SKU) | Variant of interest |
| AM122DS1640 | PGA USP 4/0, ⅜ triangular DS16 needle, undyed, 75 cm (current product) |
| AM112DS1640 | PGA USP 4/0, same DS16 needle, violet, 75 cm (maximum visibility in the field) |
| AMP122DS1640 | PGA Prime quality USP 4/0, DS16 needle, undyed, 75 cm (superior quality) |
| AM122DS1650 | PGA USP 5/0, DS16 needle, undyed, 75 cm (finer gauge) |
| AM122DS1940 | PGA USP 4/0, DS19 needle, undyed, 75 cm (larger needle size) |
| AM121DS1640 | PGA USP 4/0, DS16 needle, undyed, 45 cm (shorter thread length) |
| 1L22DS1640 | PGLA (polyglactin 910) USP 4/0, DS16 needle, undyed, 75 cm (absorption 56-72 days) |
| AM222DS1640 | PGAR rapid USP 4/0, DS16 needle, colourless, 75 cm (absorption 40-45 days) |
All references are supplied in boxes of 12 units; the minimum order is 30 boxes to guarantee continuous supply to your healthcare centre. As this is a sterile single-use product, always check the integrity of the packaging before each procedure, store it in a cool, dry place, protected from sunlight, and respect the expiry date. For special B2B conditions, volume orders or enquiries about additional gauges, lengths and colours made to order, contact the technical-commercial team of Cahnos.