Description
1. Commercial Introduction
The PGA Prime multifilament absorbable suture from the Premisut range is the reference surgical closure solution for hospitals, private clinics, purchasing groups and wholesalers requiring a high-performance synthetic absorbable thread with a very competitive price-quality ratio in B2B procurement and tenders.
This presentation combines a ½ circle needle with a cylindrical body and conical point (Taperpoint) of 48 mm (HR 48), USP gauge 0 and a 90 cm violet fluted thread. Each unit is supplied sterile and ready to use, in a box of 12 individual sachets with batch traceability, ensuring continuous and traceable supply for operating theatres and general surgery departments.
2. Technical Explanation
Polyglycolic acid (PGA) is a synthetic absorbable polymer composed 100 % of glycolide units, presented in a braided multifilament structure with a fluted finish and a coating of polycaprolactone and calcium stearate. This configuration provides great flexibility, low shape memory and a contained coefficient of friction that favours the sliding of the thread through tissue, manageability and firm, secure knot tying, with high initial tensile strength.
Absorption occurs by hydrolysis, a predictable mechanism independent of the patient’s enzymatic activity: the polymer degrades into glycolic acid, a substance that the body metabolises without generating relevant tissue reaction. The tension retention profile maintains the necessary support during the critical healing phase (75 % at 2 weeks, 50 % at 3 and 40 % at 4).
The ½ circle needle with cylindrical body and conical Taperpoint point penetrates by separating the fibres without cutting effect, which minimises trauma to soft and friable tissues. The fluted violet colouring ensures optimal visibility on the surgical field. Product sterilised by ethylene oxide, single-use and with no possibility of re-sterilisation.
3. Technical Specifications
3.1. Technical data sheet
| Parameter | Specification |
| Reference (SKU Cahnos) | AMP113HR4810 |
| Factory reference | P113HR4810 |
| Range / Quality | Premisut – Prime (premium range) |
| Thread material | 100 % polyglycolic acid (PGA), synthetic absorbable |
| Thread structure | Braided multifilament with grooved finish |
| Coating | Polycaprolactone and calcium stearate |
| Gauge / size | USP 0 (metric 3.5 – EP), conforming to USP/EP Pharmacopoeia |
| Thread colour | Violet grooved |
| Thread length | 90 cm (1 needle per thread) |
| Needle shape | ½ circle (1/2 circular) |
| Needle body | Cylindrical |
| Needle tip | Conical Taperpoint, atraumatic |
| Needle size | 48 mm (HR 48 Prime) |
| Absorption mechanism | Hydrolysis (60–90 days) |
| Sterilisation | Ethylene oxide (EO); sterile, single-use, non-resterilisable |
| Tissue reaction | Minimal |
| Indications | Tissue approximation, ligatures and general surgery |
| Presentation | Box of 12 sterile units (individual pouches) |
| Minimum order | 30 units |
3.2. Tensile strength retention profile
| Period after implantation | Retained tensile strength |
| 2 weeks | 75 % of initial tensile strength |
| 3 weeks | 50 % of initial tensile strength |
| 4 weeks | 40 % of initial tensile strength |
| Total absorption | 60–90 days, by hydrolysis |
Predictable and controlled absorption, with minimal tissue reaction, excellent tensile strength and optimal knot tying. Absorption range and retention values copied literally from the manufacturer’s documentation, without extension of limits.
4. Equivalences and Cross-References
Braided coated polyglycolic acid (PGA) with medium absorption (60–90 days) is one of the most well-established multifilament synthetic absorbable suture raw materials on the international market (classic Dexon-type family). By polymer composition (PGA homopolymer), braided structure, coating, hydrolysis absorption profile and tension retention of 75/50/40 % at 2/3/4 weeks, this reference is comparable at category level with the polyglactin 910 and PGA lines of the main manufacturers in the sector.
There is no equivalence table validated by the manufacturer that cross-references this specific configuration (PGA multifilament Prime, ½ circle conical Taperpoint needle HR 48 mm, USP gauge 0, 90 cm violet thread) with closed codes from other brands; consult the technical department of Cahnos before using in tenders. As homologous families to verify in their official datasheet before any substitution, the following are cited: Coated Vicryl / Vicryl Plus (Ethicon), Safil (B. Braun), Polysorb (Medtronic), Surgicryl PGA (SMI), DemeSORB PGA (DemeTECH) and AssuCryl PGA (Assut). The closest references located in authorised distribution (e.g. Vicryl CTX conical point 48 mm / 90 cm / violet, series J codes) correspond to USP gauge 1 and to polyglactin 910 copolymer composition, so they are not interchangeable 1:1 with this USP 0 reference in PGA homopolymer. The brands cited belong to their respective manufacturers and are mentioned only as comparative reference; check one by one: material, USP gauge, needle size (48 mm), ½ curvature, point family (conical), configuration (single needle), thread length (90 cm) and colour.
5. Selection tips and variants of interest
USP gauge 0 (metric 3.5) is very versatile: suitable for subcutaneous tissue, superficial muscle planes, approximation of soft tissues and ligation of medium-calibre vessels, with an excellent balance between initial strength and predictable absorption during healing of 2 to 4 weeks. The conical Taperpoint point separates the fibres without cutting them, being the most atraumatic option for soft and friable tissues; if the tissue is very delicate, consider the blunt point (HRN) and for firm and dense structures, the reverse cutting triangular point (HS) or trocar (HRT) from the same family.
Within the Cahnos catalogue, the variants of interest in the PGA Prime HR 48 family are:
- AM113HR4810 – same configuration (PGA, HR 48 mm needle, USP 0, violet, 90 cm) in Normal quality.
- AMP113HR4801 – PGA Prime USP 1, same HR 48 mm needle, violet, 90 cm (larger gauge).
- AMP113HR4802 – PGA Prime USP 2, same HR 48 mm needle, violet, 90 cm (higher strength).
- AMP113HR4010 – PGA Prime USP 0, conical HR 40 mm needle, violet, 90 cm (shorter needle).
- AMP113HRN4810 – PGA Prime USP 0, blunt point HRN 48 mm needle (maximum atraumatic).
- P1L13HR4810 – polyglactin 910 (PGLA) Prime, HR 48 mm needle, USP 0, violet, 90 cm (absorption 56–72 days).
If the protocol requires longer wound support (180–210 days), the polydioxanone (PDO) family is the recommended alternative. As it is sterile single-use material, always check the integrity of the sachet and the expiry date before each procedure, keep the reference together with the batch for traceability in the operating theatre and store the suture in a cool, dry place, protected from sunlight. For special B2B conditions by volume, hospital supply contracts or validation of equivalences with other manufacturers, contact the technical-commercial team of Cahnos.