Description
1. Commercial Introduction
The multifilament absorbable suture made of polyglycolic acid (PGA), reference AM113HRN4810, is the surgical closure solution your facility needs when a reliable, flexible synthetic absorbable thread with fully predictable absorption behaviour is required. It combines a ½ circle cylindrical-bodied needle with a blunt point of 48 mm (HRN 48) and a 90 cm violet fluted braided thread in USP 0 gauge, designed for the approximation of soft tissues, ligatures and medium-tension closures in general surgery.
It is a synthetic absorbable medical device, sterile and single-use, presented in B2B format in a box of 12 units with a minimum order of 30 boxes, ensuring continuous and traceable supply for hospitals, clinics, operating theatres, distributors and purchasing centres.
2. Technical Explanation
Polyglycolic acid (PGA) is a synthetic absorbable polymer composed of 100% glycolide units, presented in a braided multifilament structure with a fluted finish and a coating of polycaprolactone and calcium stearate. This configuration provides great flexibility, low shape memory and a contained friction coefficient, which favours the sliding of the thread through tissue, manageability and a firm, secure knot tying, with high initial tensile strength.
Absorption occurs by hydrolysis, i.e. by degradation of the polymer in contact with organic fluids, over a predictable period of 60 to 90 days, independent of the patient’s enzymatic activity and with minimal tissue reaction: the polymer degrades into glycolic acid, which the body metabolises without generating a relevant response. The thread retains 75% of tensile strength at 2 weeks, 50% at 3 and 40% at 4, an ideal profile for tissues requiring support during 2 to 4 weeks of healing.
The ½ circle needle with cylindrical body and blunt point of 48 mm passes through tissues separating the fibres without cutting effect, minimising trauma to delicate and friable structures. The violet colour of the thread ensures optimal visibility on the surgical field. Sterilisation by ethylene oxide; sterile single-use product, with no possibility of re-sterilisation.
3. Technical Specifications
3.1. Technical data sheet for reference AM113HRN4810
| Reference (SKU) | AM113HRN4810 |
| Range / Quality | Premisut, Normal quality |
| Material | 100% polyglycolic acid (PGA), absorbable by hydrolysis |
| Thread structure | Braided multifilament with grooved finish |
| Coating | Polycaprolactone and calcium stearate |
| Gauge | USP 0 (metric 3.5), conforming to USP/EP Pharmacopoeia |
| Thread colour | Grooved violet (also available undyed) |
| Thread length | 90 cm, 1 needle per thread |
| Needle shape | ½ circle (1/2 circular) |
| Needle body | Cylindrical |
| Needle tip | Roma (blunt point), atraumatic |
| Needle size | HRN 48 mm |
| Absorption mechanism | Hydrolysis, completed between 60 and 90 days |
| Sterilisation | Ethylene oxide (EO); sterile, single-use, non-resterilisable |
| Tissue reaction | Minimal |
| Indications | Tissue approximation, ligatures and general surgery |
| Presentation | Box of 12 units (individual sterile sachets); minimum order of 30 boxes |
| Regulatory standard | Conforms to USP/EP Pharmacopoeia for synthetic absorbable sutures |
3.2. Tensile strength retention profile
| Period after implantation | Retained tensile strength |
| 2 weeks | 75 % of initial tensile strength |
| 3 weeks | 50 % of initial tensile strength |
| 4 weeks | 40 % of initial tensile strength |
| Total absorption | 60-90 days, by hydrolysis |
Excellent resistance during the critical healing period, smooth passage through tissues, uniform and atraumatic surface, secure knotting and comfortable handling. Absorption range and retention values copied literally from the manufacturer’s documentation, without extension of limits.
3.3. Gauge equivalence according to USP/EP Pharmacopoeia
| USP size | Metric measurement (EP) | Approximate thread diameter |
| 2/0 | 3 | 0.30 – 0.349 mm |
| 0 (this product) | 3.5 | 0.35 – 0.399 mm |
| 1 | 4 | 0.40 – 0.499 mm |
| 2 | 5 | 0.50 – 0.599 mm |
The thread diameter complies with the official USP Pharmacopoeia ranges for synthetic absorbable sutures, without exceeding its limits.
4. Equivalences and Cross-References
Braided polyglycolic acid (PGA) with 60-90 days of absorption is one of the most well-established medium-absorption synthetic multifilament absorbable suture raw materials on the international market. No closed and confirmed code from another manufacturer has been found in official sources for the exact configuration of this reference (½ circle needle, cylindrical body, blunt point HRN 48 mm, USP 0 gauge, 90 cm violet thread).
At family level, the homologous reference lines in the sector are those of 100% coated braided PGA —Safil (B. Braun), Dexon II (Medtronic/Covidien), Surgicryl PGA (SMI) and DemeSORB PGA (DemeTECH)— and those of poliglactin 910 copolymer —Coated Vicryl / Vicryl Plus (Ethicon), Polysorb (Medtronic) and Novosyn (B. Braun)—. All are comparable at category level (same type of base polymer and absorption profile) but not interchangeable 1:1: for any formal substitution, the polymer and coating, USP 0 gauge, the configuration (single needle), the ½ circle curvature, the point family (blunt, blunt point), the 48 mm needle size and the 90 cm thread length must be verified one by one, since small variations in radius or point profile prevent direct equivalence. The cited brands belong to their respective manufacturers and are mentioned only as comparative reference.
There is no equivalence table validated by the manufacturer; consult the technical department before use in tenders.
5. Selection tips and variants of interest
USP 0 gauge (metric 3.5) is very versatile: suitable for subcutaneous tissue, superficial muscle planes, approximation of soft tissues and ligatures of medium-calibre vessels, with an excellent balance between initial resistance and predictable absorption during a 2 to 4 week healing period. The blunt point (HRN) is the most atraumatic option for delicate and friable tissues; if you need greater ease of penetration in somewhat denser tissues, consider the conical point version (HR 48) from the same PGA family, and for fibrous structures the trocar point (HRT).
| Reference (SKU) | Variant of interest |
| AM113HRN4810 | PGA USP 0, blunt point HRN 48 mm, violet, 90 cm (current product) |
| AMP113HRN4810 | PGA Prime USP 0, same HRN 48 mm needle, violet, 90 cm (superior quality) |
| AM113HRN4801 | PGA USP 1, same HRN 48 mm needle, violet, 90 cm (higher resistance) |
| AM113HR4810 | PGA USP 0, conical point HR 48 mm, violet, 90 cm (soft tissues) |
| AM113HRN4010H | PGA USP 0, blunt point HRN 40 mm Heavy, violet, 90 cm (reinforced smaller radius) |
| 1L13HRN4810 | Polyglactin 910 USP 0, same HRN 48 mm needle, violet, 90 cm (56-72 day absorption) |
All references are supplied in boxes of 12 units; the minimum order is 30 boxes to ensure continuous supply to your healthcare facility. As this is sterile single-use material, always check the integrity of the packaging before each procedure, respect the expiry date and store in a cool, dry place, protected from sunlight. If the procedure requires longer wound support (180-210 days), the polydioxanone (PDO) family is the recommended alternative. For special B2B volume conditions, hospital supply contracts or confirmation of equivalences for tenders, contact the technical-commercial team at Cahnos.





Reviews