Description
OPTILINE double-bottle PLUS set: advanced and safe chest drainage
The OPTILINE® double-bottle PLUS set is the reference solution at Cahnos for chest and pleural drainage in demanding clinical environments. Designed to controllably evacuate air, secretions and blood from the pleural or mediastinal cavity, this system separates the fluid collection bottle from the water seal, ensuring clear monitoring and constant vacuum throughout the care process. A robust, transparent and easy-to-use device, designed for thoracic surgery, emergency, intensive care and hospital ward services, backed by the modular OPTILINE range.
Technical operation of the device
In the double-bottle configuration, the first bottle acts as an independent collection chamber that collects drained fluids, while the second functions as a water seal: the distal end of the tube remains submerged under a fluid column that prevents air from re-entering the pleural cavity and maintains the negative pressure required for lung re-expansion. The evacuated air is directed to the seal through a short tube with a lip valve, which maintains a constant water column level and low, continuous resistance towards the cavity, regardless of the volume of fluid drained.
The PLUS version incorporates a suction control tube with a graduated scale and a pressure lock system that limits maximum suction to 21 cmH2O: if the external source exceeds this level, the control tube opens automatically to prevent excessive pressure on the patient. The lip valve also prevents fluid reflux and splashing during temporary pressure increases (coughing, manoeuvres or supine transport), also acting as a leak or fistula indicator for early detection of complications.
Technical specifications
| Characteristic | Specification |
|---|---|
| Reference | 27109 |
| Product type | Double-bottle chest drainage set PLUS (OPTILINE range) |
| Configuration | Double bottle: separate fluid collector and water seal bottle |
| Seal fill volume | 925 ml of sterile water or saline solution |
| Maximum suction | 21 cmH2O (equivalent to maximum immersion depth) |
| Pressure lock | Automatic opening of the control tube on excessive external suction |
| Suction control | Control tube with graduated scale for precise adjustment |
| Non-return valve | Prevents splashing and reflux during temporary pressure increases |
| Lip valve | Prevents reflux and acts as a leak indicator towards the collector |
| Operating mode | Drainage by gravity or with external suction |
| Material | High-quality transparent bottles for visual monitoring of drainage |
| Recommended use | Operating theatre, emergency, ICU, ward and thoracic surgery |
| Presentation | Box of 25 units |
Equivalences and cross-references
This set is part of the OPTILINE modular chest drainage range from Cahnos, a complete system with interchangeable components. Reference 27109 is the PLUS double-bottle configuration; its standard double-bottle variant is reference 27107, and the PLUS double-bottle tubing system with lip valve (reference 27108) constitutes the connection component of this same configuration. The circuit is completed by the single-bottle set (ref. 27105), the tubing systems (refs. 27104 and 27106), the 1,000 and 3,000 ml bottles (refs. 27100 and 27101), the replacement airtight lid (ref. 27102) and the mounting brackets (refs. 27112, 27113 and 27115).
As a standard water seal system with controlled suction, the set is interchangeable with the classic gravity and two-chamber suction pleural drainage configurations commonly used in hospital practice by other manufacturers in the sector, facilitating its integration into existing thoracic surgery and intensive care protocols.
Selection tips and variants of interest
Choose the PLUS double-bottle configuration (27109) when you need maximum safety in suction control, anti-reflux protection and early leak detection, especially in patients with mixed outputs or during transport. If suction control is not a priority, the standard double-bottle version (27107) covers basic needs; for low-volume drainage, the single-bottle set (27105) may be considered. If you work with external suction, check the tightness of connections before each use, keep spare lids (27102) for high-turnover shifts and accompany the system with the appropriate support (27113 for double bottle). Our technical-commercial team at Cahnos can validate the compatibility of these components with your current equipment and advise you on order volumes for purchasing or tendering projects.