Description
1. Commercial Introduction
The OPTILINE® double-bottle set (ref. 27107) is Cahnos’ reference solution for thoracic and pleural drainage in demanding clinical settings. Designed to evacuate air, secretions and blood from the thoracic cavity in a controlled manner, this system separates the fluid collection bottle from the water seal, ensuring clear monitoring and constant vacuum throughout the care process. A robust, transparent and easy-to-use device, intended for thoracic surgery, emergency, intensive care and hospital ward services, backed by the modular OPTILINE range.
Its standard double-bottle configuration combines the safety of the water seal with the anti-reflux protection of the lip valve, making it the usual choice for postoperative drainage with mixed liquid and air output.
2. Technical explanation of the equipment
In the double-bottle configuration, the first bottle acts as an independent collection chamber that collects drained fluids, while the second functions as a water seal: the distal end of the tube remains submerged under a column of liquid that prevents air from re-entering the pleural cavity, maintaining the negative pressure required for lung re-expansion. The evacuated air is directed to the seal through a short tube with a lip valve, which maintains a constant water column level and low, continuous resistance towards the cavity, regardless of the volume of fluid drained.
The lip valve prevents fluid reflux from the water seal towards the patient even during temporary loss of the seal (coughing, manoeuvres, transport in supine position) and acts as a leak or fistula indicator for early detection of complications. The transparent body of the bottles allows immediate assessment of drainage rate, bubbling and appearance of the exudate, and the integrated recording label facilitates continuous clinical supervision. The system works by gravity or connected to an external suction source via the corresponding connection hose from the range.
3. Technical Specifications
| Characteristic | Specification |
| Reference | 27107 |
| Product type | Double-bottle chest drainage set (OPTILINE range, standard version) |
| Configuration | Double bottle: separate fluid collection bottle + water seal bottle |
| Bottle capacity | 2 bottles of 1,000 ml (standard configuration; compatible with 3,000 ml bottle, ref. 27101, for high output) |
| Seal mechanism | Water seal with constant vacuum maintenance and low, continuous resistance |
| Valve | Lip valve with non-return function and leak/fistula indicator |
| Operating mode | Drainage by gravity or with external suction |
| Visual control | Transparent bottles + integrated recording label |
| Material | High-quality transparent medical materials |
| Recommended use | Operating theatre, emergency department, ICU, hospital ward and thoracic surgery |
| Presentation | Box of 25 units |
Components and compatible accessories from the OPTILINE range:
| Reference | Component |
| 27100 | 1,000 ml chest drainage bottle |
| 27101 | 3,000 ml chest drainage bottle |
| 27102 | Replacement airtight cap |
| 27104 | OPTILINE hose system for single bottle |
| 27105 | OPTILINE single-bottle set |
| 27106 | Double-bottle hose system with lip valve |
| 27108 | Double-bottle PLUS hose system with lip valve |
| 27109 | OPTILINE double-bottle PLUS set (with graduated suction control and pressure lock) |
| 27112 | Stand for 1,000 ml bottle (single-bottle set) |
| 27113 | Stand for double-bottle set (2 × 1,000 ml) |
| 27115 | Stand for 3,000 ml bottle |
| 27116 | Heimlich valve for unidirectional drainage |
| 27121 | Chest drainage bag |
4. Equivalences and Cross-References
Within the modular OPTILINE range, the PLUS double-bottle variant (ref. 27109) is the direct equivalent of this configuration with advanced features (suction control tube with graduated scale, 925 ml seal charge and maximum suction limited to 21 cmH2O). The other references in the family behave as interchangeable circuit components (bottles, hoses, caps and stands).
Cross-references verified with other manufacturers in the sector, at the functional category level (double-chamber water seal systems for pleural drainage):
| Manufacturer / Brand | Family / Model | Reference code / SKU | Equivalence level | Technical differences / Remarks |
| Rocket Medical (United Kingdom) | Rocket® Chest Drainage – double-chamber bottle with water seal | R54509 | Comparable by category (not interchangeable 1:1) | Integrated double-chamber configuration (collector + seal) versus the two separate bottles of the OPTILINE set; total capacity 1,800 ml; graduations of 5 and 10 ml; sterile single-use. Code verified in GUDID/FDA (Rocket Medical PLC, class I). |
| Cardinal Health (formerly Covidien) | Aqua-Seal™ – thoracic drainage unit with wet suction, double drainage | 8888571406 | Comparable by category (not interchangeable 1:1) | Wet double drainage system without OPTILINE-type lip valve; verified in public declaration of conformity (ANMAT, Argentina). |
| Cardinal Health (formerly Covidien) | Argyle™ – tubing set for two bottles (neck 38 mm; tube 7.15 mm) | 8884729600 | Comparable by category (not interchangeable 1:1) | Connection tubing between bottles only, not a complete set; useful as a connector reference in two-bottle configurations. |
Notable homologous brands and families in the sector for the same clinical indication (pleural drainage by gravity and suction with water seal) include Teleflex (Pleur-evac™, dry suction with wet seal), Redax (Drentech Variant, three- and four-chamber configurations) and Vygon (thoracic drainage systems), among others. There is no manufacturer-validated equivalence table allowing 1:1 interchange with these brands; consult the technical department before use in tenders.
5. Selection tips and variants of interest
- Suction control requirement: if your centre requires limiting external suction with a graduated control tube and pressure lock, consider the PLUS variant (ref. 27109); the standard version 27107 covers basic double-bottle needs with anti-reflux protection.
- Output and volume: for patients with high-volume drainage or prolonged air leaks, combine the system with the 3,000 ml bottle (ref. 27101); the 1,000 ml one is sufficient for conventional postoperative drainage.
- External suction: if you work with suction, complete the circuit with the appropriate hose system (ref. 27106 or 27108) and secure the bottles with the double-bottle support (ref. 27113).
- Continuity of care: have spare airtight caps (ref. 27102) available for high-rotation shifts and check the tightness of connections and the integrity of the circuit before each use.
- Alternative configurations: for low-volume drainage, the single-bottle set (ref. 27105) may be considered; for simple unidirectional drainage, the Heimlich valve (ref. 27116) or the drainage bag (ref. 27121).
Our technical-commercial team at Cahnos can validate the compatibility of these components with your current equipment and advise you on order volumes for purchase or tender projects.