Description
1. Commercial Introduction
The Branson MH range is the reference choice for ultrasound cleaning in laboratories, clinics, precision workshops and sterilisation rooms. These units combine in a single solution a 60-minute mechanical timer, continuous operation and a heating system, enabling deep, fast and safe cleaning of instruments and parts: their power is sufficient to remove heavy oils and waxes, penetrates circuit boards and, at the same time, is gentle enough to clean small precision assemblies. The manufacturer has selected Branson technology for its long experience in the ultrasonic cleaning method, which results in easy-to-use equipment with a higher-than-average frequency for deeper cleaning, lower noise levels and a longer service life.
2. Technical explanation of the equipment
Ultrasonic cleaning is based on the phenomenon of cavitation: ultrasonic sound waves travelling through the cleaning solution generate rapid formation and implosion of micro-bubbles that dislodge dirt even in the most difficult-to-reach areas. Heat enhances this process, so the deep cleaning action of ultrasound also removes the most persistent dirt and contamination.
Branson MH units operate at a frequency of 40 kHz, higher than the market average, resulting in a deeper and more effective effect. They incorporate a frequency sweep mode (sweep frequency) that distributes ultrasonic energy evenly throughout the tank and high/low power control: 100 % power for normal applications and 70 % for safe cleaning of delicate components. The heating control and 60-minute mechanical timer allow the work cycle to be adjusted simply and reliably, always with the option of continuous operation.
3. Technical Specifications
Features of the MH series (mechanical with heating)
| Timer | 60 minutes |
| Tank lid | Included |
| Continuous mode | Yes |
| Frequency sweep | Yes (sweep frequency) |
| Heating | Yes |
| Working frequency | 40 kHz |
| Power supply | 230-240 V |
| Power control | 100 % normal / 70 % delicate applications |
Technical details by model
GM35502 · Branson 2800 MH · 2.8 l
| Tank dimensions | 230 x 140 x 100 mm |
| External dimensions | 337 x 304 x 302 mm |
| Tank capacity | 2.8 l |
| Drain and valve kit | No |
| Weight | 4.1 kg |
GM35511 · Branson 3800 MH · 5.7 l
| Tank dimensions | 290 x 150 x 150 mm |
| External dimensions | 397 x 317 x 377 mm |
| Tank capacity | 5.7 l |
| Drain and valve kit | Included |
| Weight | 5.4 kg |
GM35521 · Branson 5800 MH · 9.5 l
| Tank dimensions | 290 x 240 x 150 mm |
| External dimensions | 397 x 401 x 378 mm |
| Tank capacity | 9.5 l |
| Drain and valve kit | Included |
| Weight | 6.4 kg |
GM35532 · Branson 8800 MH · 20.8 l
| Tank dimensions | 495 x 290 x 150 mm |
| External dimensions | 596 x 466 x 391 mm |
| Tank capacity | 20.8 l |
| Drain and valve kit | Included |
| Weight | 11.8 kg |
Available references
| Reference | Model | Sales unit |
|---|---|---|
| GM35502 | 2800 MH ULTRASONIC CLEANER 2.8 l | 1 pc |
| GM35511 | 3800 MH ULTRASONIC CLEANER 5.7 l | 1 pc |
| GM35521 | 5800 MH ULTRASONIC CLEANER 9.5 l | 1 pc |
| GM35532 | 8800 MH ULTRASONIC CLEANER 20.8 l | 1 pc |
Product data
| Manufacturer | GIMA |
|---|---|
| Sales unit | 1 pc |
| Device type | No medical device |
All Branson units are supplied with a tank lid and a multilingual manual (GB, FR, IT, ES, DE, PT, SE, NL, RU). Their use with the optional perforated basket is recommended.
4. Equivalences and Cross-References
The references GM35502, GM35511, GM35521 and GM35532 correspond to the Branson 2800 MH, 3800 MH, 5800 MH and 8800 MH models, whose manufacturer catalogue codes are 35502, 35511, 35521 and 35532. There is no equivalence table validated by the manufacturer that cross-references these codes with those of other brands; the products listed below are classified as comparable at category level and are not interchangeable 1:1 without prior verification of the technical data sheet of each unit.
| Manufacturer / Brand | Family / Model | Reference SKU | Equivalence level | Technical differences / Remarks |
|---|---|---|---|---|
| Euronda Pro System | Eurosonic Energy | 113025 | Comparable by category (not interchangeable 1:1) | 2.7 l tank with Sweep Mode technology and mechanical timer, but without heating; operating frequency of 30-32 kHz compared to the 40 kHz of the Branson MH series. |
| Euronda Pro System | Eurosonic 3D | 113102 | Comparable by category (not interchangeable 1:1) | Digital 2.7 l tank with constant temperature of 60 °C and electronic timer from 5 to 30 minutes; digital control compared to the mechanical timer of the Branson MH series. |
| Euronda Pro System | Eurosonic 4D | 113152 | Comparable by category (not interchangeable 1:1) | Digital 3.8 l tank with 6 preconfigured cycles and LED control; lower capacity than the Branson 2800 models at the top of the range. |
| Euronda Pro System | Eurosonic Micro | 113007 | Comparable by category (not interchangeable 1:1) | Compact 0.5 l tank with a single 10-minute cycle, designed for small instruments and burs. |
It is recommended to consult the technical department of Cahnos before proposing any substitution or use in tenders, since capacities, frequencies and control systems differ between manufacturers.
5. Selection tips and variants of interest
- Working volume: choose the capacity according to the volume of instruments to be processed: 2.8 l (GM35502) for small loads, 5.7 l (GM35511) and 9.5 l (GM35521) for daily use in practices or laboratories, and 20.8 l (GM35532) for high volumes or larger parts.
- Heating and continuous mode: the MH series includes heating and continuous operation, which accelerates the cleaning action and allows prolonged work cycles without interruption.
- Delicate parts: use the power control at 70% for sensitive components and 100% for general cleaning of instruments and parts with embedded dirt.
- Optional perforated basket: use with the optional perforated basket is recommended to keep parts in suspension and optimise the cavitation effect.
- Process validation: to periodically check the cavitation performance, an efficacy test in accordance with standard EN ISO 15883-5 can be used.
- Cleaning liquid: use detergents specifically for ultrasound and demineralised or distilled water, and respect the fill level indicated by the manufacturer to protect the transducer.














Reviews