Description
The EUROPA B EVO autoclave (ref. 35657) is a class B steam steriliser with a 24-litre chamber, designed for dental clinics, medical practices, prosthetic laboratories and services that need to sterilise instruments with certified physical and legal safety features. As a class B autoclave in accordance with standard EN 13060, it reliably sterilises hollow materials such as handpieces and turbines, porous materials (gowns, gauze swabs) and solid loads, with a printed record of each cycle.
Technical explanation of the equipment
The EUROPA B EVO series uses a P.I.D. (Proportional Integral Derivative) system, the ultimate reference in closed-loop control: it constantly monitors and regulates the entire operating system of the autoclave and acts automatically in the event of any anomaly or incorrect technical or electronic parameter. The unit has internal tanks for pure water and used water, each with minimum and maximum level detection probes. The preconfigured working temperatures are 134 °C (for solid and ferrous materials) and 121 °C (for thermoplastic and delicate materials).
Cycle selection is carried out with a single button, which allows scrolling through all sterilisation cycles and choosing the most suitable one on the screen, and another button starts the complete cycle; it is impossible to select an inappropriate cycle by mistake. All cycles are fully automatic (vacuum, preheating, exposure time and drying managed by a control board with a powerful microprocessor) and include an anti-condensation process and automatic shutdown if the operator is not next to the unit when the cycle finishes. The autoclave also incorporates test cycles (Helix, Bowie & Dick and vacuum) for periodic control through physical and functional tests, continuous monitoring of cycles with immediate lockout in the event of any anomaly that compromises sterilisation, and guaranteed sterilisation of type B loads in all cycles thanks to the fractional vacuum system and parameter control.
It includes an integrated osmosis system (reverse osmosis) that removes even the smallest contaminating particles to guarantee the bacteriological purity of the water through an osmotic membrane with molecular-level filtration, and an integrated printer for the documentary record of the cycles.
Technical specifications
General characteristics
| Characteristic | Specification |
|---|---|
| External dimensions | 510 x 390 x 730 mm (width x height x depth) |
| Chamber dimensions | 245 x 500 mm (Ø x depth) |
| Operating voltage | 230 V – 50/60 Hz |
| Maximum power consumption | 2,200 W |
| Weight | 58 kg |
| Integrated printer | Yes |
| Manual closure | Yes |
| Connection for pure deioniser | Yes |
| P.I.D. system | Yes |
| Internal tanks | Pure water and used water with level probes |
| Integrated osmosis system | Yes (reverse osmosis) |
| Supply | 1 closed tray holder, 4 aluminium trays, 1 dual-purpose key for tray extraction and door adjustment, water fill and drain hoses, chamber cleaning sponge and integrated osmosis system |
Sterilisation cycles
| Cycle | Min-max working pressure | Minimum total cycle duration* |
|---|---|---|
| 121 °C Wrapped | 1.04 – 1.30 | 56 min (25+18+13) |
| 121 °C Unwrapped | 1.04 – 1.30 | 55 min (25+18+12) |
| 134 °C Flash | 2.02 – 2.38 | 33 min (25+4+4) |
| 134 °C Wrapped | 2.02 – 2.38 | 43 min (25+5+13) |
| 134 °C Unwrapped | 2.02 – 2.38 | 42 min (25+5+12) |
| 134 °C Prions | 2.02 – 2.38 | 58 min (25+20+13) |
*The total time includes vacuum, preheating time, cycle and drying phase; the times are calculated for a fully loaded autoclave. All sterilisation cycles have 3 vacuum stages. The vacuum and preheating time may vary between 25 and 35 minutes depending on the autoclave conditions.
Product data
| Manufacturer | GIMA |
|---|---|
| Sales unit | 1 unit |
| Device type | Medical device (medical device) |
| Device class | II B |
| CND | Z12011305 |
| GMDN | 42162 |
| EAN | 8023279356571 |
Equivalences and cross-references
This autoclave corresponds to the manufacturer’s reference 35657. (EUROPA B EVO series, 24 litres, 230 V), marketed by Cahnos with SKU GM35657 and EAN 8023279356571. No validated 1:1 equivalence table with other brands is available from reliable sources for this model: before approving an autoclave from another manufacturer, the load class (B), chamber capacity (24 l), chamber dimensions, voltage, presence of integrated printer and osmosis system, available cycles (121 °C and 134 °C) and regulatory classification (class II B medical device) must be verified. The CND (Z12011305) and GMDN (42162) codes allow the equipment to be located in the European medical device classifications.
Selection tips and variants of interest
This 24-litre autoclave is intended for centres with a medium workload that need to sterilise hollow instruments (handpieces and turbines), porous instruments and solid loads with Class B guarantee. When selecting the equipment, check the chamber capacity against the volume of instruments processed per day, the available supply voltage (230 V – 50/60 Hz) and the required power (2,200 W). Also consider the need for an integrated printer for document traceability, the integrated osmosis system to ensure water purity, and the included test cycles (Helix, Bowie & Dick and vacuum) for periodic equipment checks. For B2B conditions, complete equipment projects, advice on the EUROPA B EVO family or support in tenders, contact the technical-commercial department of Cahnos.