Description
1. Commercial Introduction
The 15-litre EUROPA B EVO autoclave guarantees the physical and legal safety required by healthcare practice: effective sterilisation of instruments is the first link in patient and professional safety. Designed in accordance with standard EN 13060, this class B steriliser can process complete type B loads: hollow materials (handpieces and turbines) and porous materials (gowns, gauze swabs and textiles), making it the reference solution for dental clinics, medical practices, podiatry, advanced aesthetics, outpatient operating theatres and laboratories.
2. Technical explanation of the equipment
The EUROPA B EVO uses a fractionated vacuum cycle with three vacuum phases, capable of removing air from the chamber and from porous and hollow loads, an essential condition for saturated steam to penetrate and sterilise in depth. The entire process is managed by a state-of-the-art electronic board with a microprocessor, regulated by a PID system (Proportional-Integral-Derivative) that constantly monitors the entire operating system and acts automatically in the event of any anomaly or incorrect parameter.
The unit features internal tanks for pure water and waste water, both with minimum and maximum level detection probes, and preset temperatures of 134 °C and 121 °C: the 134 °C cycles are commonly used for solid and ferrous instruments, while the 121 °C cycles are reserved for thermoplastic and delicate materials. The cycles are fully automatic (vacuum, preheating, exposure time and drying, without human intervention) and incorporate anti-clumping and auto-stop if the operator is not next to the unit when it finishes. A control device constantly monitors each cycle, blocking the process if an anomaly that compromises sterilisation is detected and informing the operator via messages and LED. It also includes periodic test cycles (Helix, Bowie & Dick and vacuum test) to validate the vacuum capacity in hollow bodies and the chamber’s tightness.
3. Technical Specifications
General characteristics
| Parameter | Specification |
|---|---|
| Capacity | 15 litres |
| External dimensions | 510 x 390 x 590 mm (W x H x D) |
| Chamber dimensions | 245 x 320 mm (Ø x depth) |
| Operating voltage | 230 V – 50/60 Hz |
| Maximum power consumption | 1,700 W |
| Weight | 54 kg |
| Integrated printer | Yes |
| Manual closure | Yes |
| Connection for pure water deioniser | Yes |
| PID system | Yes |
| Load classification | Type B (EN 13060) |
Sterilisation cycles
| Cycle | Working pressure min-max | Minimum total duration* |
|---|---|---|
| 121 °C Wrapped | 1.04 – 1.30 | 60 minutes (25+18+17) |
| 121 °C Unwrapped | 1.04 – 1.30 | 49 minutes (25+18+6) |
| 134 °C Flash | 2.02 – 2.38 | 31 minutes (25+4+2) |
| 134 °C Wrapped | 2.02 – 2.38 | 47 minutes (25+5+17) |
| 134 °C Unwrapped | 2.02 – 2.38 | 36 minutes (25+5+6) |
| 134 °C Prion | 2.02 – 2.38 | 62 minutes (25+20+17) |
*Total time includes vacuum, preheating, cycle phase and drying, calculated with the autoclave at full load. All sterilisation cycles have 3 vacuum phases. The vacuum and preheating time may vary between 25 and 35 minutes depending on autoclave conditions.
Product data
| Manufacturer | GIMA |
|---|---|
| Manufacturer reference | 35653 |
| Sales unit | 1 pc |
| Device type | Medical device |
| Device class | II B |
| CND | Z12011305 |
| GMDN | 42162 |
| EAN | 8023279356533 |
Supply contents: autoclave with integrated printer, 1 closed tray rack, 4 aluminium trays, 1 dual-purpose key (tray extraction and door adjustment), water fill and drain hoses and chamber cleaning sponge.
4. Equivalences and Cross-References
The factory reference for this steriliser is 35653 (EAN 8023279356533), marketed by Cahnos under SKU GM35653 within the EUROPA B EVO class B autoclave range. No equivalence table validated by the manufacturer has been found in reliable sources to cross-reference this reference 1:1 with models from other brands; the 15-litre EUROPA B EVO belongs to the class B autoclave family in accordance with standard EN 13060, differing by chamber capacity and equipment (with or without printer) depending on the version. Before any replacement or use in tenders, it is recommended to verify field by field (capacity, chamber dimensions, available voltages, cycles and regulatory classification) and consult the technical-commercial department of Cahnos.
5. Selection tips and variants of interest
Select the 15-litre EUROPA B EVO when you need a class B benchtop autoclave with integrated printer, suitable for sterilising hollow and porous loads in dental clinics, medical practices and advanced aesthetics. Before choosing, assess the daily processing volume: the 15-litre capacity is suitable for medium-to-high activity practices; for higher volumes or centralised sterilisation, consider larger capacity autoclaves from the same class B family. Always check the mains voltage (230 V), the available power (1,700 W) and the installation conditions for the optional pure water deioniser. Use only purified/deionised water to preserve the chamber and steam generators, periodically perform the test cycles (Helix, Bowie & Dick and vacuum) according to the centre’s protocol and record the cycles using the integrated printer for documentary traceability. For large orders you can request a quote: for volume conditions, training, commissioning or technical documentation for tenders, contact the technical-commercial team of Cahnos.