Description
1. Introduction
The TOSI®-FlexiCheck efficacy test for endoscope washer-disinfectors is the professional reference solution for the validation and routine control of cleaning and disinfection processes for flexible endoscopes. Designed for sterilisation centres, endoscopy units and reprocessing services, this device quickly and objectively verifies that the washer-disinfector and its detergents act correctly on flexible cannulated instruments of medium and long length, ensuring patient safety and regulatory compliance.
2. Technical explanation of the equipment and procedure
Flexible endoscopes are frequently contaminated with mucus, blood and organic residues that are difficult to remove inside their channels. The TOSI®-FlexiCheck system simulates a flexible endoscope channel and connects directly to the irrigation system of the washer-disinfector (WD-E) via its Luer-lock connector, undergoing the same washing cycle as the real material.
The test plate combines two standardised soils on a single strip: one correlated with coagulated human blood (fibrin) and a polysaccharide mixture that simulates mucous secretions. At the end of the cycle, visual inspection of the strip allows the result to be interpreted: a completely clean plate confirms the effectiveness of the wash; fibrin residues indicate a limited chemical capacity of the detergent, polysaccharide residues point to rinsing or water quality problems, and persistent soiling in both areas suggests an obstructed irrigation system. It is a ready-to-use, easy-to-interpret, repeatable test that complies with standard EN ISO 15883-5 (Annex A).
3. Technical Specifications
| Property | Specification |
|---|---|
| Device type | Efficacy test for endoscope washer-disinfectors (flexible endoscopes) |
| Compatibility | Flexible cannulated instruments of medium and long length |
| Connector | Luer-lock for direct attachment to the rack and to the washer irrigation system |
| Dimensions | 55 x 5 mm |
| Test soil | Double: standardised blood (fibrin) and polysaccharide mixture |
| Interpretation | Immediate visual inspection at the end of the wash cycle |
| Regulatory standard | Validation and control in accordance with EN ISO 15883-5 (Annex A) |
| Presentation | Box of 25 units |
4. Equivalences and Cross-References
The TOSI®-FlexiCheck system belongs to the TOSI® washing indicator family from the manufacturer Pergut Medical, the most widely used washing test family worldwide. The cross-references of this product in the main European and American suppliers are:
- stericlin: devices and strips under references 3FRIN555120 (test device), 3FRIN555125 (spare strips, box of 25), 3FRIN555121 (adapter A, Wassenburg) and 3FRIN555122 (adapter C, Olympus).
- Healthmark Industries: TOSI FlexiCheck monitor in a box of 25 units, reference FLEXCHK-101.
- Pergut Medical (official catalogue): TOSI-FlexiCheck efficacy test, reference 2024025, box of 25 units.
This test shares the standard and control philosophy with the rest of the instrument cleaning family available at Cahnos: the TOSI® test for general surgical instruments (ref. 11100120), the TOSI®-LumCheck for rigid cannulated instruments (ref. 20100250), the HemoCheck-S for detection of blood residues on surfaces (ref. 6706712) and the SonoCheck® for ultrasound washers (ref. 30550300).
5. Selection tips and variants of interest
- For the validation and routine control of flexible endoscope washer-disinfectors, choose the TOSI®-FlexiCheck (SKU 2024025); it is the specific test for flexible lumens of medium and long length.
- If your instrument inventory is mostly rigid cannulated (laparoscopic forceps, cannulas), choose the TOSI®-LumCheck (SKU 20100250).
- For wash control of general surgical instruments (scissors, needle holders, articulated instruments), use the TOSI® test (SKU 11100120).
- Check the compatibility of the connection adapter with your washer: there are specific versions for Wassenburg systems (adapter A) and Olympus (adapter C), among others.
- The recommended frequency of use is defined according to the profile of each unit (process validation, qualification and periodic control), following quality assurance protocols based on EN ISO 15883-5.
- Complete the programme with the HemoCheck-S (SKU 6706712) to confirm the absence of blood residues on surfaces and manually washed material before sterilisation.