Description
1. Introduction
The Tyvek® Esterivec mixed pouch of 250 x 480 mm (SKU 310812) is the sterile barrier packaging solution that your sterile services department needs for medium-to-large medical devices. Manufactured by Pergut Medical within its Esterivec range for low-temperature processes, it combines a polyolefin fibre side with a Tyvek® base and a transparent plastic film side: reliable microbial barrier, mechanical strength and full visibility of the contents until the moment of use. Single-use, heat-sealable pouch, designed for the safe packaging of instrument sets in chemical sterilisation by plasma and hydrogen peroxide.
2. Technical explanation of the procedure
Sterilisation by Gas Plasma and Hydrogen Peroxide (H2O2) is a low-temperature process indicated for heat-sensitive instruments and medical devices that cannot tolerate steam autoclaving (endoscopes, electronics, optical materials, among others). In these cycles, the packaging material must allow the diffusion of the sterilising agent and, at the same time, act as a sterile barrier once the process is complete.
The Esterivec pouch fulfils this dual function: the porous polyolefin fibre side with a Tyvek® base allows hydrogen peroxide and gas plasma to penetrate and be evacuated, while the transparent film allows visual inspection of the contents without opening the pouch. It incorporates a chemical process indicator for H2O2 and Gas Plasma (Type e1, ISO 11140-1) that changes from red to yellow upon confirmation of exposure to the cycle, facilitating verification of each pack. Sealing is carried out by heat sealing with an approved sealer, respecting the temperature parameters (115–126 °C) indicated in the manufacturer’s technical data sheet. Latex-free product.
3. Technical Specifications
| Characteristic | Specification |
| Material | Polyolefin fibre side (Tyvek® base) and transparent plastic film side |
| Measurements | 250 x 480 mm |
| Compatibility | Hydrogen Peroxide (H2O2) and Gas Plasma |
| Process indicator | Chemical indicator for H2O2 and Gas Plasma (Type e1, red – yellow colour change) |
| Application | Packaging of medical devices for low-temperature chemical sterilisation |
| Format | Preformed flat mixed pouch |
| Presentation | Box of 500 units |
| Regulatory standard | ISO 11607-1 and 2:2019; EN 868-9:2019 |
| CE marking | Yes – Class I medical device (RD 192/2023) in accordance with Regulation (EU) 2017/745 (MDR) |
| Manufacturer | Pergut Medical (Esterivec range) |
Technical properties of the Esterivec packaging range (Pergut Medical technical data sheet)
| Property | Standard | Value |
| Polyethylene fibre sheet (Tyvek® base) | ISO 536 | 60.5 g/m² |
| Polyester – polyethylene film (12/50 µm) | ISO 536 | 64 g/m² |
| Film delamination | EN 868-5 Annex A | Exempt |
| Seal width | EN 868-5 annex C | > 9 mm |
| H2O2 – Plasma process indicator (Type e1) | ISO 11140-1 | Red – yellow |
| Dry seal strength | ISO 1924-2 | > 2.5 N/15 mm |
| Wet seal strength | ISO 1924-2 | > 2.2 N/15 mm |
| Peelability (fibre tear) | EN 868-5 annex C | < 10 mm |
| Sealing temperature | EN 868-5 · 4.7.a | 115–126 °C |
| Expiry | — | 5 years from date of manufacture |
| Recommended storage | — | 10–30 °C · 30–60 % relative humidity |
Esterivec family of flat mixed pouches (box of 500 units)
| Reference | Measurements | Indicator |
| 310806 | 75 x 200 mm | H2O2 – Plasma |
| 310807 | 100 x 260 mm | H2O2 – Plasma |
| 310809 | 150 x 320 mm | H2O2 – Plasma |
| 310810 | 150 x 420 mm | H2O2 – Plasma |
| 310811 | 200 x 400 mm | H2O2 – Plasma |
| 310812 | 250 x 480 mm | H2O2 – Plasma |
| 310813 | 320 x 560 mm | H2O2 – Plasma |
| 310814 | 300 x 650 mm | H2O2 – Plasma |
4. Equivalences and Cross-References
Exact reference equivalence (same range): this pouch belongs to the Esterivec range of flat mixed pouches from Pergut Medical (Sterimed Group), compatible with the main plasma and hydrogen peroxide sterilizers on the market (STERRAD, V-PRO, LTSF and other low-temperature systems). The manufacturer’s official reference for this format follows the documented range scheme: 134310812 (Pergut program code); the factory code for this specific format is pending confirmation in the manufacturer’s official data sheet before use in tenders. The sales reference on Cahnos is SKU 310812.
Comparable at category level (not interchangeable 1:1): due to its Tyvek®/film composition and its indicator for H2O2 and plasma, this pouch is comparable to other commercial barrier ranges for hydrogen peroxide vapour (vH2O2) based on Tyvek 1073B with film, such as Tyvek-Bolsaplast pouches (e.g. MRT1176 and MRT1177), Igaltex Tyvek pouches, STERIKING pouches or Tuttnauer Tyvek® pouches. The comparison is established at a functional level: exact dimensions, Tyvek grammage (60.5 g/m² in Esterivec) and manufacturer validation must be confirmed before replacing one reference with another.
There is no equivalence table validated by the manufacturer for 1:1 cross-referencing with other brands; it is recommended to consult the technical department before use in tenders or in the standardisation of a packaging fleet. The Esterivec range is specific to low-temperature processes; for steam, ethylene oxide or formaldehyde, the mixed paper and film pouches and rolls from the Viewpack or BOP series by Pergut Medical are recommended.
5. Selection tips and variants of interest
Format selection: the pouch must exceed the size of the device leaving a minimum margin of 10 mm on each side, so that the seal is not under tension. The 250 x 480 mm pouch is ideal for instrument sets, laparoscopes, long forceps and scissors, guides and medium-to-large devices.
Variants of the same family (Esterivec flat mixed pouches): for smaller parts consider refs. 310806 (75 x 200 mm), 310807 (100 x 260 mm), 310809 (150 x 320 mm) and 310810 (150 x 420 mm); for longer instruments, refs. 310811 (200 x 400 mm) and 310813 (320 x 560 mm), all in boxes of 500 units. For high volumes or custom pouch manufacturing, consider the Esterivec mixed rolls of 70 m (refs. 310816 to 310824).
Recommended accessories: CT40 indicator tape (20 mm x 50 m) for sealing and external control of plasma and hydrogen peroxide packs, CD40 multivariable chemical indicator (Type 4, ISO 11140-1) for internal control of each pack, and gas plasma biological controls for process validation.
Correct use: seal with an approved heat-sealing machine, respecting the temperature and pressure indicated by the manufacturer, check that the weld width is greater than 9 mm, verify the indicator colour change (red to yellow) after each cycle, and store the bags in a cool, dry place (10–30 °C, 30–60 % relative humidity), in their original packaging and away from sterilising agents until use.
Professional B2B purchasing: for high volumes, framework agreements or process approval needs, please consult our technical-commercial team: we can advise you on seal validation, labelling and supply in accordance with regulations.