Description
1. Introduction
The 200 mm x 70 m mixed Tyvek® Esterivec roll (Ref. 310819) is the sterile barrier packaging solution that your sterile processing department needs to protect medical devices processed in low-temperature gas plasma and hydrogen peroxide (H2O2) cycles. Manufactured within the Esterivec range from Pergut Medical, this roll combines a porous polyolefin fibre face with a Tyvek® base and a transparent plastic film face, allowing visual verification of the contents and ensuring a reliable sterile barrier until the moment of use. Its presentation in a box of 4 units is designed for continuous supply in sterile processing departments, operating theatres and health services.
2. Technical explanation
Sterilisation by gas plasma and hydrogen peroxide (H2O2) is a low-temperature process indicated for heat-sensitive medical instruments and devices that cannot tolerate steam autoclave (endoscopes, electronics, optical material, among others). In these cycles, the sterile barrier system (SBS) must allow the entry and evacuation of the gaseous sterilising agent and, at the same time, prevent the passage of microorganisms once the process is complete.
The Esterivec roll fulfils this dual function: the microporous polyolefin sheet (Tyvek® base) acts as a microbial filter that allows vaporised H2O2 and gas plasma to diffuse, while the transparent film allows visual inspection of the device without opening the packaging. It incorporates a Type e1 chemical process indicator (ISO 11140-1) printed under the seals, which changes from red to yellow upon confirmation of exposure to the agent, facilitating verification of each cycle. The material is cut to size with the sealer and closed by heat sealing, forming pouches adapted to the length of each device.
3. Technical Specifications
| Characteristic | Specification |
|---|---|
| Material | Microporous polyolefin fibre (Tyvek® base) and transparent plastic film |
| Dimensions | 200 mm wide x 70 m long |
| Compatibility | Hydrogen peroxide (H2O2) and plasma gas |
| Process indicator | Chemical indicator Type e1 (ISO 11140-1) – changes from red to yellow |
| Application | Packaging of medical devices for low-temperature chemical sterilisation |
| Format | Flat mixed roll or reel |
| Presentation | Box of 4 units |
| Regulatory standard | ISO 11607-1/2:2019; EN 868-9:2019 |
| CE marking | Yes – Medical device Class I (RD 192/2023 · MDR (EU) 2017/745) |
| Manufacturer | Pergut Medical (Esterivec range) |
Material properties (Pergut Medical technical data sheet)
| Property | Standard | Value |
|---|---|---|
| Polyethylene fibre sheet (Tyvek® base) | ISO 536 | 60.5 g/m² |
| Polyester – polyethylene film (12/50 µm) | ISO 536 | 64 g/m² |
| Film delamination | EN 868-5 Annex A | Exempt |
| Seal width | EN 868-5 annex C | > 9 mm |
| H2O2 – Plasma process indicator (Type e1) | ISO 11140-1 | Red – yellow |
| Dry seal strength | ISO 1924-2 | > 2.5 N/15 mm |
| Wet seal strength | ISO 1924-2 | > 2.2 N/15 mm |
| Peelability (fibre tear) | EN 868-5 annex C | < 10 mm |
| Dimensional tolerance (width) | EN 868-5 annex E | ± 3 mm |
| Dimensional tolerance (length) | EN 868-5 annex E | ± 10 mm |
| Sealing temperature | EN 868-5 · 4.7.a | 115–126 °C |
| Expiry | — | 5 years from date of manufacture |
| Recommended storage | — | 10–30 °C · 30–60 % relative humidity |
4. Equivalences and Cross-References
Reference equivalence within the same range: this roll belongs to the Esterivec range of flat mixed rolls from Pergut Medical (Sterimed Group), compatible with the main plasma and hydrogen peroxide sterilisers on the market (STERRAD, V-PRO, LTSF and other low-temperature systems). The other widths in the same family (75, 100, 150, 250, 300, 350, 400 and 500 mm, all 70 m) share the same construction and materials, differing only in size; for the 200 mm width (SKU Cahnos 310819), the official Pergut manufacturer code is pending confirmation and no unverified code is indicated. For reference, in neighbouring formats verified in the Pergut Medical technical data sheet, the 300 mm roll corresponds to references 135310821 (standard programme) and 18G2EST018 (new factory), and the 350 mm to 135310824 and 18G2EST021.
Comparable at category level (not 1:1 interchangeable): due to its Tyvek®/film composition and its indicator for H2O2 and plasma, this roll is comparable to other barrier systems for hydrogen peroxide vapour (vH2O2) based on Tyvek 1073B with film, such as the Tyvek ranges from Bolsaplast (e.g. MRT1176 and MRT1177) or the Tyvek rolls from Igaltex. The comparison is functional: exact dimensions, Tyvek grammage and manufacturer validation must be confirmed before substituting one reference for another. There is no official equivalence table validated by Pergut Medical for 1:1 cross-referencing between manufacturers; it is recommended to consult the technical department before use in tenders or in the homogenisation of a packaging portfolio.
Do not confuse with other agents: the Esterivec range is specific to low-temperature processes (plasma and hydrogen peroxide). For steam, ethylene oxide or formaldehyde cycles, the mixed paper and film rolls from the BOP TI series of Pergut Medical (G2T references) are used, with chemical indicators specific to each agent.
5. Selection tips and variants of interest
Select the roll width according to the device size: the packaging must exceed the size of the instrument leaving a minimum margin of 10 mm per side, so that the seal is not under tension. The 200 mm x 70 m format is suitable for scissors, long forceps, guides, semi-rigid laparoscopes, endoscopes and medium-sized instrument sets, allowing pouches to be made to measure and optimising material consumption. For small instruments, the 75 to 150 mm formats are more cost-effective; for bulky trays and kits, the 250 to 500 mm widths.
Respect the sealing parameters (temperature 115–126 °C) and check that the seal width is greater than 9 mm to guarantee the integrity of the sterile barrier. Verify the change of the process indicator (red to yellow) after each cycle, as required by process control according to ISO 11140-1, and store the roll in a cool, dry place (10–30 °C; 30–60 % relative humidity), in its original packaging and away from sterilising agents until use.
Variants of the same range of Esterivec rolls (70 m length): Ref. 310816 (75 mm, box of 8 units), Ref. 310817 (100 mm, box of 8 units), Ref. 310818 (150 mm, box of 4 units), Ref. 310820 (250 mm, box of 2 units), Ref. 310821 (300 mm, box of 2 units), Ref. 310824 (350 mm, box of 2 units), Ref. 310822 (400 mm, box of 2 units) and Ref. 310823 (500 mm, box of 2 units). There are also preformed flat mixed Esterivec bags (refs. 310806 to 310813) for direct packaging without a sealer.
Recommended accessories: CT40 indicator tape (20 mm x 50 m) for sealing and external control of plasma and hydrogen peroxide packages, CD40 multivariable chemical indicator (Type 4, ISO 11140-1) for internal control of each package, and gas plasma biological controls for process validation. For high volumes, framework agreements or process approval needs, please consult our technical-commercial team: we can advise you on sealing validation, labelling and supply in accordance with regulations.




Reviews
There are no reviews yet.