Description
1. Commercial Introduction
The Cleantex® mixed bag of 420 x 1180 mm from Pergut Medical Sterimed is the packaging solution your sterilisation department needs when the medical device is so large, heavy or elongated that standard medical paper falls short. Its preformed sterile barrier system (SBS) format combines a 60 g/m² non-woven side with a high-strength transparent film, offering an effective microbial barrier, great packaging flexibility and clean opening without film curling. A professional, reliable and standards-compliant answer for operating theatres, sterilisation departments and instrumentation services that handle instruments of great length and volume.
2. Technical Explanation
Cleantex® is the first preformed sterile barrier system (SBS) that replaces paper with a wet-laid non-woven side of 60 g/m², combined with a multilayer film of polyester (12 µm) and polypropylene (36–40 µm) printed in the interlayer. This construction is designed for heavy and/or bulky medical devices that generate mechanical stresses during handling, transport and the sterilisation cycle itself.
The porous side acts as a microbial filter: it allows the entry and evacuation of the sterilising agent (steam, ethylene oxide or formaldehyde) but prevents the passage of microorganisms, preserving the sterility of the contents during storage. The film side allows visual inspection of the contents and, as the chemical indicator is printed in the film interlayer, there is no need to open the package to verify exposure to the agent. The film does not curl or splinter when opened, facilitating aseptic opening with controlled peelability. It contains no natural rubber latex and is compliant with TSE/BSE (EMEA/410/01).
3. Technical Specifications
Data verified in the official SPS Medical technical data sheet (document FT-SCX_01/EN, March 2022) and in the online catalogue of the Cleantex® range:
| Property | Specification |
| Product | Preformed SBS sterilisation mixed pouch |
| Manufacturer reference | SCX4201180 (SPS Medical catalogue / Cleantex® range) |
| Reference Cahnos (SKU) | 482S042118 |
| Nominal dimensions | 420 x 1180 mm (tolerance ±3 mm in width and ±10 mm in length, according to EN 868-5) |
| Porous side (barrier) | Wet-laid nonwoven of 60 g/m² (cellulose with synthetic fibres and binders), validated according to EN 868-2 |
| Film side (visibility) | Transparent film with a lilac tint, laminated from 12 µm polyester and 36–40 µm polypropylene, printed intrafilm; does not curl when opened |
| Process indicator | Steam, Class 1 according to ISO 11140-1, printed intrafilm (colour change pink → brown) |
| Sterilisation methods | Steam, ethylene oxide and formaldehyde (according to SPS/Pergut commercial documentation). Not suitable for gamma radiation or hydrogen peroxide plasma (according to the official technical data sheet) |
| Seal strength | ≥ 3 N/15 mm (EN 868-5) |
| Sealing temperature | 170 – 185 °C; sealing pressure 90 – 120 N/mm² |
| Conformity | EN ISO 11607-1 and EN 868-5 |
| CE classification | Medical device Class I (accessory) according to Regulation (EU) 2017/745 (MDR) |
| Expiry | 5 years from date of manufacture |
| Recommended storage | 10 – 30 °C; 30 – 60 % relative humidity, in original packaging |
| Presentation | Box of 200 units (see note) |
Note: the SPS Medical France catalogue lists SCX4201180 at 300 units per box; the Cahnos presentation (and the Pergut Portugal website for the same SKU 482S042118) is 200 units. Verify the presentation when placing the order.
Cleantex® range: size programme and cross-references Cahnos / manufacturer
| Manufacturer reference | Dimensions (mm) | SKU Cahnos | Units/box |
| SCX220X450 | 220 x 450 | 482S220450 | 500 |
| SCX285X350 | 285 x 350 | 482S285350 | 800 |
| SCX285X450 | 285 x 450 | 482S285450 | 500 |
| SCX285X750 | 285 x 750 | 482S285750 | 400 |
| SCX320X550 | 320 x 550 | 482S320550 | 300 |
| SCX420X600 | 420 x 600 | 482S420600 | 300 |
| SCX4201180 | 420 x 1180 | 482S042118 | 200* |
| SCX480X640 | 480 x 640 | 482S480640 | 250 |
| SCX480X750 | 480 x 750 | 482S480750 | 250 |
| SCX660X900 | 660 x 900 | 482S660900 | 200 |
| SCX7501200 | 750 x 1200 | 482S750120 | 200 |
*Cahnos/Pergut Portugal presentation; SPS Medical France catalogue indicates 300 units for this reference.
4. Equivalences and Cross-References
Exact reference equivalence: the manufacturer’s official code (SPS Medical, Cleantex® range) for this format is SCX4201180, verified in the SPS Medical online catalogue and in the manufacturer’s official technical data sheet (document FT-SCX_01/EN). The same product is marketed under the group’s brands depending on the market: SPS Medical (France), Pergut Medical (Spain/Portugal) and Bromeda Medical (Germany). The Cahnos own SKU is 482S042118, which matches the programme nomenclature used by authorised distributors.
Not verified: a Dutscher distributor reference for this specific 420 x 1180 mm format could not be confirmed in current public sources. The historical references of the range verified in other data sheets (150535, 150537, 150538, 150541) correspond to other Cleantex sizes. A Dutscher reference for this size must not be cited without explicit confirmation from the distributor.
Functional category comparison (not 1:1 interchangeable): due to its non-woven/film construction with steam indicator, this bag is comparable to the peelable paper-film pouches of the Pergut range itself (e.g. BOP DI or BOP TI) and to mixed non-woven bags from other manufacturers in the sector (e.g. non-woven/film ranges from Bolsaplast or Igaltex). However, it is not 1:1 interchangeable: the barrier material (60 g/m² non-woven versus medical paper), the non-woven grammage, the film composition and the dimensions differ, so these parameters must be confirmed before replacing one reference with another in a packaging protocol.
There is no cross-reference equivalence table validated by the manufacturer for cross-brand comparison; consult the technical department before using in tenders or replacing references in sterilisation department protocols.
Important: Cleantex® is indicated for steam, ethylene oxide and formaldehyde. For low-temperature processes (plasma / hydrogen peroxide), Tyvek®-type material (e.g. the Esterivec range) must be used, not this bag.
5. Selection tips and variants of interest
- Filling rule: the device must occupy at most up to 2/3 of the package capacity, leaving sufficient margin for the perimeter seal, the labelling area and proper penetration of the sterilising agent.
- Heavy and long loads: this 420 x 1180 mm format is ideal for elongated instruments over one metre (channels, endoscopes, tubes, long surgical guides) and for heavy devices that are difficult to condition with standard paper. Its superior mechanical strength reduces the risk of breakage during handling and transport.
- Sealing: use a sealing machine validated in accordance with EN ISO 11607-2, at 170 – 185 °C, and place the packages in the autoclave face to face (film to film or porous to porous). Verify the indicator colour change (pink → brown) after each cycle, according to ISO 11140-1.
- Size selection: for very bulky and heavy devices, consider the larger formats in the range: 480 x 750, 660 x 900 and 750 x 1200 mm. For more contained packages, the 420 x 600, 320 x 550 or 285 x 450 mm formats are more compact alternatives. All share the same material and construction.
- Compatibility: this bag is not suitable for sterilisation by gamma radiation or hydrogen peroxide plasma; for these processes, the Cahnos catalogue offers specific alternatives such as the Esterivec range.
- Storage: store in a cool, dry place (10 – 30 °C; 30 – 60 % RH), in its original packaging, protected from direct sunlight and sterilising agents.




Reviews