Description
1. Introduction
The 320 x 500 mm flat BOP TI mixed pouch from Pergut Medical Sterimed (manufactured by SPS Medical, a member of the Sterimed Group) is the professional packaging solution for medium and large medical devices: instrument trays, forceps, scissors, motors, surgical kits and material that must maintain its sterility up to the point of use. Its mixed format, with one side of premium medical-grade paper and the other of transparent lilac-coloured film, combines mechanical strength, microbial barrier and full visibility of the contents throughout the sterilisation and storage process. A reliable sterile barrier pouch, with triple chemical process indicator, designed for sterilisation departments, operating theatres, clinics and practices that require traceability and regulatory compliance in instrument reprocessing.
2. Technical Explanation
It is a preformed mixed-type sterile barrier system (SBS). The premium medical-grade paper side, 60 g/m², with controlled porosity and reinforcing surface treatment, allows penetration and evacuation of the sterilising agent; the multilayer polyester-polypropylene (PET/PP) film side, transparent and lilac-coloured, provides visibility of the contents and mechanical strength during handling and storage.
The pouch incorporates 3 Class 1 chemical process indicators (ISO 11140-1) printed on the paper under the seals, one for each compatible method: steam (change from pink to brown), ethylene oxide (from blue to green) and formaldehyde (from pink to green). On completion of the cycle, the corresponding indicator changes colour, confirming that the package has been exposed to the agent; they are process indicators and do not replace biological, physical or external chemical control of the cycle. Sealing is carried out with a validated heat sealer in accordance with EN ISO 11607-2 at 170–185 °C and 90–120 N/mm², and opening is aseptic thanks to its controlled peelability (fibre tear of less than 10 mm). It is a single-use product: do not re-sterilise. Not suitable for gamma radiation or hydrogen peroxide plasma.
3. Technical Specifications
General data:
| Reference (SKU) | S2T320X500 |
| Brand / manufacturer | Pergut Medical, S.L.U. (SPS Medical · Grupo Sterimed) |
| Packaging type | Flat BOP TI mixed pouch, peelable, for sterilisation |
| Measurements | 320 x 500 mm |
| Composition | Premium medical-grade paper 60 g/m² + lilac PET/PP film |
| Valid processes | Steam, ethylene oxide (EO) and formaldehyde |
| Presentation | Box of 500 units |
| Use | Single-use · do not re-sterilise |
| Expiry | 5 years from date of manufacture |
Certified technical properties:
| Property | Standard | Value |
| Premium medical-grade paper | ISO 536 | 60 g/m² |
| Polyester – polypropylene film (PET/PP) | ISO 536 | 53 g/m² (12/40 µm) |
| Seal width | EN 868-5 annex C | > 9 mm |
| Dry seal strength | ISO 1924-2 | > 1.5 N/15 mm |
| Wet seal strength | ISO 1924-2 | > 1.2 N/15 mm |
| Peelability (fibre tear) | EN 868-5 annex C | < 10 mm |
| Tolerance in measurements (width / length) | EN 868-5 annex E | ± 3 mm / ± 10 mm |
| Sealing temperature | EN 868-5 · 4.7.a | 170 – 185 °C |
| Sealing pressure | EN 868-5 · 4.7.a | 90 – 120 N/mm² |
| Maximum recommended fill | Manufacturer’s instructions | 2/3 of the capacity |
| Recommended storage | — | 10 – 30 °C · 30 – 60 % RH |
| Conformity | — | EN 868-5 · ISO 11607-1 y 2 · ISO 11140-1 |
| Regulatory classification | — | Class I medical device · CE marked · MDR 2017/745 |
Chemical process indicators (Class 1, ISO 11140-1):
| Process | Indicator colour change |
| Steam | Pink → Brown |
| Ethylene oxide | Blue → Green |
| Formaldehyde | Pink → Green |
4. Equivalences and Cross-References
Code verification status: the SKU S2T320X500 is consistently confirmed in the channel: it is used by Cahnos for this 320 x 500 mm BOP TI pouch and it also appears with sector distributors (e.g. Mevesur markets it as a 320 x 500 mm mixed sterilisation pouch in a pack of 500 units). It also follows the official coding system of the manufacturer SPS Medical for flat pouches with triple indicator (codes S2T***X***, verified in the technical data sheet TDS FT-S2T/G2T-01). However, the internal factory code of Pergut Medical/Amcor for this size could not be verified in public sources: confirm with the manufacturer before use in tenders.
Distribution-level equivalence (not verified in this review): the same 320 x 500 mm format could be distributed in the channel under reference 1004S2T320X500 (pattern used by DH Medical for other sizes in the same BOP TI range, e.g. 1004S2T300X500 or 1004S2T320X550). This code could not be directly cross-checked for the 320 x 500 format in public sources: pending confirmation with distributor/manufacturer.
Category comparables (not 1:1 interchangeable) within the same BOP TI family of Pergut Medical Sterimed: S2T300X500 (same 500 mm length, 300 mm opening), S2T320X450 (same 320 mm opening, shorter length), S2T320X550 and S2T380X500 (same length, greater width). They share composition, triple indicator and sealing; they can only be substituted with an adjustment in slack and never if the exact size is critical. The gusseted version S2F320X500 (320 x 60 x 500 mm) provides volume for bulky instruments: comparable by category, different geometry. Flat rolls (e.g. G2T320X200) allow the pouch to be made to measure: same family, different format. BOP DI self-adhesive pouches (closure without heat sealer and double indicator) belong to another family within the same brand: not equivalent.
Flat mixed pouches with triple indicator from other manufacturers are comparable at category level, but the paper weight and type, film composition, peelability, seal width and conformity standards must be validated one by one before any substitution. There is no equivalence table validated by the manufacturer between the BOP TI range and other brands; consult the technical department before use in tenders.
5. Selection tips and variants of interest
Filling rule: the device must not occupy more than 2/3 of the pouch capacity, leaving slack for the sterilising agent to circulate, the sealing zone and labelling. Sealing: use a heat sealer validated to EN ISO 11607-2 at 170 – 185 °C and 90 – 120 N/mm², and place the pouches in the steriliser with film against film and paper against paper. Verification: confirm the colour change of the 3 indicators after each cycle before considering the material suitable; biological control remains mandatory according to protocol. Storage: keep in a cool, dry place (10 – 30 °C, 30 – 60 % relative humidity), in its original packaging and protected from direct sunlight; observe the 5-year expiry. Do not re-sterilise.
The BOP TI range from Pergut Medical Sterimed is available on Cahnos in other sizes of flat heat-sealable pouch (e.g. S2T100X200, S2T150X400, S2T210X360, S2T270X450, S2T300X500, S2T320X450, S2T320X550, S2T380X500), in a gusseted version (S2F series) for bulky instruments and in flat or gusseted rolls (G2T / G2F series) for making the pouch to measure with a heat sealer, optimising material consumption in high-volume sterilisation departments.