Description
1. Introduction
The BOP TI flat mixed pouch, 250 x 360 mm, from Pergut Medical Sterimed (manufactured by SPS Medical, a member of the Sterimed Group) is the professional solution for packaging and sterilising medium-sized medical devices: forceps, scissors, small surgical sets, motors and consultation instruments. Its mixed format, with one side of Premium medical-grade paper and the other of transparent film coloured lilac, combines mechanical strength, microbial barrier and full visibility of the contents during the process and subsequent storage.
A reliable sterile barrier packaging, with triple chemical process indicator, designed for sterilisation departments, operating theatres, clinics and any environment requiring traceability and regulatory compliance in instrument reprocessing.
2. Technical Explanation
The BOP TI range (easy opening with triple indicator) is a preformed sterile barrier system (SBS) of mixed type with peelable opening. The Premium medical-grade paper side, 60 g/m², with controlled porosity, allows penetration and evacuation of the sterilising agent during the cycle, while the transparent lilac-coloured multilayer polyester-polypropylene (PET/PP) film side provides visibility of the contents and mechanical strength during handling and storage.
The packaging incorporates 3 Class 1 chemical process indicators (ISO 11140-1) printed on the paper under the seals, one for each compatible method: steam (colour change from pink to brown), ethylene oxide (from blue to green) and formaldehyde (from pink to green). On completion of the cycle, the corresponding indicator changes colour, confirming the package’s exposure to the sterilising agent; these indicators verify the process but do not replace biological control of the cycle. Sealing is carried out with a validated heat sealer and opening is aseptic thanks to its controlled peelability, without tearing the paper. It is a single-use product: do not re-sterilise. Not suitable for gamma radiation or hydrogen peroxide plasma.
3. Technical Specifications
General data
| Property | Specification |
|---|---|
| Reference Cahnos (SKU) | S2T250X360 |
| Manufacturer | Pergut Medical, S.L.U. (SPS Medical · Grupo Sterimed) |
| Packaging type | Flat BOP TI mixed pouch, peelable, for sterilisation |
| Measurements | 250 x 360 mm (tolerance ±3 mm width and ±10 mm length, EN 868-5 annex E) |
| Presentation | Peelable mixed pouch, box of 1000 units |
| Use | Single-use · do not re-sterilise |
Composition and structure
| Property | Specification |
|---|---|
| Material (porous side) | «Premium» medical-grade paper, 60 g/m², high resistance to puncture, tearing and rupture (ISO 536) |
| Material (film side) | Multilayer PET/PP film, 53 g/m² (12/40 µm), transparent, coloured lilac (ISO 536) |
| Chemical indicators | 3 process indicators printed on the paper under the seals |
| Sterilisation methods | Steam, ethylene oxide (EO) and formaldehyde |
| Not compatible | Gamma radiation and hydrogen peroxide plasma |
Process indicators (Class 1, ISO 11140-1)
| Sterilising agent | Standard | Colour change |
|---|---|---|
| Steam | ISO 11140-1 | Pink → Brown |
| Ethylene oxide | ISO 11140-1 | Blue → Green |
| Formaldehyde | ISO 11140-1 | Pink → Green |
Certified properties (BOP TI family)
| Property | Standard | Value |
|---|---|---|
| Seal width | EN 868-5 annex C | > 9 mm |
| Dry seal strength | ISO 1924-2 | > 1.5 N/15 mm |
| Wet seal strength | ISO 1924-2 | > 1.2 N/15 mm |
| Peelability (fibre tear) | EN 868-5 annex C | < 10 mm |
| Sealing temperature | EN 868-5 · 4.7.a | 170 – 185 °C |
| Sealing pressure | EN 868-5 · 4.7.a | 90 – 120 N/mm² |
| Expiry | — | 5 years from date of manufacture |
| Recommended storage | — | 10 – 30 °C · 30 – 60 % relative humidity |
| Conformity | — | EN 868-5 · ISO 11607-1 y 2 · ISO 11140-1 |
| Regulatory classification | — | Class I medical device · CE marked · MDR 2017/745 |
4. Equivalences and Cross-References
The reference S2T250X360 corresponds to the code used consistently in the Cahnos catalogue for the BOP TI flat pouch series from SPS Medical / Pergut Medical. The BOP TI family shares, in all its sizes, the same composition (Premium paper 60 g/m² + PET/PP film 53 g/m²), the triple chemical indicator (steam, ethylene oxide and formaldehyde) and the same sealing conditions. The specific factory code for this size is not verified in public sources from the manufacturer; confirm the exact reference with the technical department before use in tendering.
Variants from the same manufacturer and BOP TI family (flat pouches)
| SKU Cahnos | Dimensions (mm) | Difference compared to S2T250X360 |
|---|---|---|
| S2T210X360 | 210 x 360 | Same length of 360 mm, 40 mm narrower |
| S2T250X360 (this product) | 250 x 360 | Analysed format |
| S2T250X380 | 250 x 380 | Same opening of 250 mm, 20 mm longer |
| S2T250X420 | 250 x 420 | Same opening of 250 mm, 60 mm longer |
| S2T250X500 | 250 x 500 | Same opening of 250 mm, 140 mm longer |
These variants are comparable by category and replaceable with clearance adjustment, but not interchangeable 1:1 if the exact length is critical. The gusseted version (S2F series) has the same possible opening and length but a different geometry, intended for bulky instruments; rolls from the same range (G2T series) allow custom-made pouches to be produced; and the self-adhesive BOP DI pouches (closure without heat sealer and double indicator) constitute a different, non-equivalent family.
At category level, flat mixed pouches with triple indicator from other manufacturers in the hospital sterilisation sector are comparable but not interchangeable 1:1: the paper weight and type, film composition, peelability, seal width and conformity standards must each be validated before any substitution. There is no equivalence table validated by the manufacturer between the BOP TI range and other brands; consult the technical department before use in tendering.
5. Selection tips and variants of interest
Filling rule: the device must not occupy more than 2/3 of the packaging capacity, leaving clearance for circulation of the sterilising agent, the sealing area and labelling.
Sealing: use a validated heat sealer in accordance with EN ISO 11607-2 at 170 – 185 °C and 90 – 120 N/mm²; place the packages face to face in the steriliser (film to film and paper to paper).
Verification: confirm the colour change of the 3 indicators after the cycle before deeming the material suitable; biological control remains mandatory according to protocol.
Storage: keep in a cool, dry place (10 – 30 °C, 30 – 60 % relative humidity) in its original packaging, protected from direct sunlight, and respect the 5-year expiry date. Do not re-sterilise.
Variants of the same family: the BOP TI range is available in other flat sizes (e.g. S2T210X360, S2T270X450, S2T320X450, S2T420X600), in a gusseted version (S2F series) for bulky instruments and in rolls for custom sealing, adapting the packaging to each type of instrument.
Compatibility: this packaging is not suitable for gamma radiation or hydrogen peroxide plasma; for these processes, please refer to the specific alternatives in the Cahnos catalogue.





Reviews