Description
1. Product Presentation
The flat mixed BOP TI pouch of 100 x 300 mm from Pergut Medical Sterimed (manufactured by SPS Medical, a member of the Sterimed Group) is the professional solution for packaging and sterilising small medical devices: fine forceps, scissors, burs, motors, small surgical sets and consulting room instruments. Its mixed structure combines the strength of a Premium medical-grade paper with the transparency of a lilac-coloured film, allowing the contents to be checked without opening the pouch. With triple chemical process indicator and heat-sealable closure, it is the usual choice in sterilisation departments, operating theatres, dental clinics and practices that require traceability and regulatory compliance in every cycle. It is supplied in a box of 1,200 units.
2. Technical Explanation
It is a preformed sterile barrier system (SBS) of mixed type with peelable opening, in flat format (without gusset). The porous side is made of Premium medical-grade paper of 60 g/m² (ISO 536), with controlled porosity that allows the penetration and evacuation of the sterilising agent during the cycle, and high resistance to perforation, tearing and rupture. The transparent side is a multilayer polyester-polypropylene (PET/PP) film of 53 g/m², coloured lilac, providing mechanical strength and full visibility of the contents during handling and storage.
The paper incorporates 3 Class 1 chemical process indicators (ISO 11140-1) printed under the seals, one for each compatible method: steam (colour change from pink to brown), ethylene oxide (from blue to green) and formaldehyde (from pink to green). On completion of the cycle, the corresponding indicator changes colour, confirming the package’s exposure to the sterilising agent; these indicators verify the process but do not replace biological indicator, which remains mandatory according to protocol. Sealing is carried out with a validated heat sealer in accordance with EN ISO 11607-2 and opening occurs aseptically thanks to its controlled peelability (fibre tear of less than 10 mm). It is a single-use pouch: do not re-sterilise. Latex-free and TSE/BSE-free.
3. Technical Specifications
Technical data for the BOP TI flat pouch range from Pergut Medical / SPS Medical (material technical data sheet and manufacturer’s catalogue).
| Property | Standard / Method | Value |
|---|---|---|
| Reference Cahnos (SKU) | — | S2T100X300 |
| Dimensions (width x length) | — | 100 x 300 mm |
| Packaging type | — | Preformed flat mixed pouch with peelable opening |
| Premium medical-grade paper | ISO 536 · EN 868-3 | 60 g/m² |
| Polyester – polypropylene film (PET/PP) | ISO 536 | 53 g/m² (12/40 µm), transparent, lilac colour |
| Seal width | EN 868-5 annex C | > 9 mm |
| Peelability (fibre tear) | EN 868-5 annex C | < 10 mm |
| Sealing temperature | EN 868-5 · 4.7.a | 170 – 185 °C |
| Sealing pressure | EN 868-5 · 4.7.a | 90 – 120 N/mm² |
| Tolerance in measurements (width / length) | EN 868-5 annex E | ± 3 mm / ± 10 mm |
| Maximum recommended fill | SPS Medical instructions | 2/3 of the capacity |
| Presentation | — | Box of 1,200 units |
| Expiry | — | 5 years from date of manufacture |
| Recommended storage | — | 10 – 30 °C · 30 – 60 % relative humidity |
| Conformity | — | EN 868-5 · ISO 11607-1 and 2 |
| Regulatory classification | — | Medical device Class I (accessory) · CE marking · MDR 2017/745 |
Chemical process indicators (printed on the paper under the seals):
| Indicator (Class 1, ISO 11140-1) | Process | Colour change |
|---|---|---|
| Steam process indicator | Steam / autoclave | Pink → Brown |
| Ethylene oxide process indicator | Ethylene oxide (EO) | Blue → Green |
| Formaldehyde process indicator | Formaldehyde | Pink → Green |
Size range of the BOP TI flat pouch range (S2T references):
| Reference S2T | Format (width x length) |
|---|---|
| S2T050X250 | 50 x 250 mm |
| S2T100X200 | 100 x 200 mm |
| S2T100X300 (this product) | 100 x 300 mm |
| S2T100X420 | 100 x 420 mm |
4. Equivalences and Cross-References
Code reference: the SKU Cahnos S2T100X300 is consistent with the S2T coding system of the manufacturer’s BOP TI flat pouch range (SPS Medical · Sterimed Group), used consistently across the entire family of flat pouch sizes. No official factory code from the manufacturer has been found in public sources for this exact size: the manufacturer’s code is not verifiable in public sources and must be confirmed with Pergut Medical / SPS Medical before use in tenders or technical specifications.
Within the same manufacturer and the same BOP TI family: the flat pouches S2T050X250, S2T100X200 and S2T100X420 share composition, triple indicator and sealing, but each size is a different article: composition equivalence does not imply dimensional interchangeability. BOP TI gusseted pouches (S2F series, e.g. S2F100X300) and flat rolls (G2T series) or gusseted rolls (G2F series) are comparable at category level, not interchangeable 1:1 if the same pouch geometry is required.
Other brands: flat mixed pouches with triple indicator (steam, ethylene oxide and formaldehyde) from other manufacturers or distribution brands, such as Gallesteril, Laboquimia or DH Material Médico, are comparable only at category level, not interchangeable 1:1: the paper weight and type, film composition, peelability, seal width, dimensional tolerances and conformity standards must be validated one by one before any substitution. BOP DI self-adhesive pouches (closure without heat sealer and double indicator) constitute a different family within the same brand and are not equivalent. There is no equivalence table validated by the manufacturer between the BOP TI range and other brands; consult the technical department before use in tenders.
5. Selection tips and variants of interest
- Filling clearance: the device must not occupy more than 2/3 of the capacity to allow circulation of the sterilising agent and avoid stress on the seals. The 100 x 300 mm format is ideal for short, compact instruments; for longer devices consider the 100 x 420 mm.
- Correct sealing: use a validated heat sealer (EN ISO 11607-2) at 170 – 185 °C and 90 – 120 N/mm²; place the pouches plastic-to-plastic and paper-to-paper in the steriliser.
- Cycle verification: confirm the colour change of the 3 indicators after sterilisation before deeming the material suitable, and record batch, date and contents on the labelling. Biological control remains mandatory according to protocol.
- Storage: store in a cool, dry place (10 – 30 °C, 30 – 60 % relative humidity), in its original packaging and protected from direct light. Do not re-sterilise.
- BOP TI range variants: the same family is available in other flat bag sizes (S2T series), in gusseted bags (S2F series) for bulky instruments and in flat or gusseted rolls (G2T / G2F series) for custom sealing, adapting the packaging to each type of instrument and workload.
- Compatibility: this packaging is not suitable for gamma radiation or hydrogen peroxide plasma; for these processes, consult the specific alternatives in the Cahnos catalogue.