Description
1. Product Presentation
The flat mixed pouch BOP TI of 210 x 360 mm from Pergut Medical Sterimed (SPS Medical · Grupo Sterimed) is the professional solution for packaging and sterilisation of medium-sized medical devices: surgical instruments, forceps, scissors, motors and small surgical kits. Its mixed structure combines the strength of a Premium medical-grade paper with the transparency of a film coloured lilac, allowing the contents to be checked without opening the package. With triple chemical process indicator and heat-sealable closure, it is the usual choice in sterilisation centres, operating theatres, clinics and hospitals that demand traceability and regulatory compliance in every cycle.
2. Technical Explanation
It is a preformed sterile barrier system (SBS) of mixed type with peelable opening. The Premium medical-grade paper side of 60 g/m², with controlled porosity and reinforcing treatment, allows penetration and evacuation of the sterilising agent during the cycle; the transparent lilac-coloured multilayer polyester-polypropylene (PET/PP) film side provides total visibility of the contents and mechanical resistance during handling and storage.
The paper incorporates 3 chemical process indicators (Class 1, ISO 11140-1) printed under the seals, one for each compatible method: steam (change from pink to brown), ethylene oxide (from blue to green) and formaldehyde (from pink to green). On completion of the cycle, the corresponding indicator changes colour, confirming the package’s exposure to the sterilising agent; these indicators verify the process but do not replace the biological control of the cycle, which is mandatory according to protocol. Sealing is carried out with a validated heat sealer in accordance with EN ISO 11607-2 at 170 – 185 °C and 90 – 120 N/mm², and opening occurs aseptically thanks to its controlled peelability (fibre tear of less than 10 mm). It is a single-use package: do not re-sterilise. It is not suitable for gamma radiation or hydrogen peroxide plasma.
3. Technical Specifications
General data
| Characteristic | Detail |
|---|---|
| Reference Cahnos (SKU) | S2T210X360 |
| Manufacturer | Pergut Medical, S.L.U. (SPS Medical · Grupo Sterimed) |
| Packaging type | Flat BOP TI mixed pouch, peelable, for sterilisation |
| Measurements | 210 x 360 mm (tolerance ±3 mm width / ±10 mm length, EN 868-5 annex E) |
| Material (porous side) | Premium medical-grade paper of 60 g/m² (ISO 536), high resistance to puncture, tear and rupture |
| Material (film side) | Multilayer polyester–polypropylene (PET/PP) film, 53 g/m² (12/40 µm), transparent, lilac-coloured |
| Chemical indicators | 3 Class 1 process indicators (ISO 11140-1) printed under the seals |
| Sterilisation methods | Steam, ethylene oxide (EO) and formaldehyde |
| Not compatible | Gamma radiation and hydrogen peroxide plasma |
| Presentation | Peelable mixed pouch, box of 1000 units |
| Use | Single-use · do not re-sterilise |
| Expiry | 5 years from date of manufacture |
| Recommended storage | 10 – 30 °C · 30 – 60 % relative humidity |
| Conformity | EN 868-5 · ISO 11607-1 y 2 · ISO 11140-1 |
| Regulatory classification | Class I medical device · CE marked · MDR 2017/745 |
Process indicators (Class 1, ISO 11140-1)
| Sterilising agent | Colour change |
|---|---|
| Steam | Pink → Brown |
| Ethylene oxide | Blue → Green |
| Formaldehyde | Pink → Green |
Certified properties of the BOP TI family
| Property | Standard | Value |
|---|---|---|
| Seal width | EN 868-5 annex C | > 9 mm |
| Dry seal strength | ISO 1924-2 | > 1.5 N/15 mm |
| Wet seal strength | ISO 1924-2 | > 1.2 N/15 mm |
| Peelability (fibre tear) | EN 868-5 annex C | < 10 mm |
| Sealing temperature | EN 868-5 · 4.7.a | 170 – 185 °C |
| Sealing pressure | EN 868-5 · 4.7.a | 90 – 120 N/mm² |
4. Equivalences and Cross-References
The reference S2T210X360 is the code consistently used in the Cahnos catalogue for the BOP TI flat pouch series from SPS Medical / Pergut Medical; the BOP TI family shares across all its sizes the same composition (Premium paper 60 g/m² + PET/PP film of 53 g/m²), the triple chemical indicator (steam, ethylene oxide and formaldehyde) and the same sealing conditions. The official factory code from Pergut Medical / Sterimed for the exact 210 x 360 mm format could not be verified in public sources; it remains pending confirmation with the manufacturer’s technical department before use in tenders.
Same BOP TI family (comparable by category, not interchangeable 1:1 if the exact size is critical): S2T210X320 (210×320), S2T210X400 (210×400), S2T210X420 (210×420), S2T210X450 (210×450) and S2T210X600 (210×600) for sharing the 210 mm opening; S2T250X360 (250×360) shares the 360 mm length with a larger opening. For other geometries, the family is completed with S2T150X400, S2T270X450, S2T320X450 and S2T420X600. In roll form for custom sealing, the G2T series of the same width exists (G2T210X200).
Flat mixed pouches of 60 g/m² medical-grade paper with triple indicator (steam, ethylene oxide and formaldehyde) sold under distribution brands such as Gallesteril, DH Material Médico or Laboquimia are comparable at category level, but not interchangeable 1:1: the measurements, paper weight and type, film composition, peelability, seal width and certifications (EN 868-5 / ISO 11607) must be validated one by one. Self-adhesive BOP DI pouches (closure without heat sealer and double indicator) constitute a different family within the same brand: they are not equivalent. There is no equivalence table validated by the manufacturer between the BOP TI range and other brands; consult the technical department before use in tenders.
5. Selection tips and variants of interest
- Filling rule: the device must not occupy more than 2/3 of the package capacity, leaving slack for the circulation of the sterilising agent, the sealing area and labelling; the item must not touch the seals.
- Correct sealing: use a validated heat sealer in accordance with EN ISO 11607-2 at 170 – 185 °C and 90 – 120 N/mm²; place the packages face to face in the steriliser (film to film and paper to paper).
- Cycle verification: confirm the colour change of the 3 indicators after sterilisation before deeming the material suitable; biological control remains mandatory according to protocol.
- Storage: keep the box in a cool, dry place (10 – 30 °C, 30 – 60 % relative humidity), in its original packaging and protected from direct sunlight; respect the 5-year expiry date. Do not re-sterilise.
- Range variants: the same series is available in other flat heat-sealable sizes (e.g. S2T210X400, S2T210X450, S2T270X450, S2T320X450, S2T420X600), in a gusseted version (S2F series) for bulky instruments and in rolls for custom sealing (G2T series).
- Compatibility: this package is not suitable for gamma radiation or hydrogen peroxide plasma; for these processes consult the specific alternatives in the Cahnos catalogue.