Description
1. Introduction
ELINE flat sterilisation rolls are the bespoke packaging solution for sterilisation departments, operating theatres, clinics and laboratories that need to make instrument packs of different sizes. With 200 m length per roll and widths from 50 to 400 mm, they allow the packaging to be cut exactly to the size of the device, optimising consumption and reducing waste. They are manufactured in Italy from medical-grade materials and are intended for steam (autoclave) and ethylene oxide sterilisation.
2. Technical Explanation
The ELINE flat roll consists of a transparent polyester-polypropylene film bonded to medical-grade paper. This combination allows steam and gas to penetrate during the sterilisation process, ensuring that the sterilising agent reaches the instruments and that they subsequently dry. The paper side provides a suitable surface for writing and traceability labelling, while the transparent film side allows the contents to be visually checked without opening the package.
The range incorporates steam and gas indicators that show a clear colour change when the package has been exposed to the sterilisation process, allowing rapid visual verification of the cycle. This is a single-use package: it must not be re-sterilised.
3. Technical Specifications
The ELINE flat roll range offers a length of 200 m per roll and is available in the following widths, with their corresponding outer and inner channel dimensions:
| Reference | Size | Sales unit | Details |
|---|---|---|---|
| GM35803 | 50 mm AUTOCLAVE | Box of 8 units | Flat rolls · Outer: 200 m x 50 mm · Inner: 200 m x 36 mm |
| GM35804 | 75 mm AUTOCLAVE | Box of 8 units | Flat rolls · Outer: 200 m x 75 mm · Inner: 200 m x 53 mm |
| GM35805 | 100 mm AUTOCLAVE | Box of 4 units | Flat rolls · Outer: 200 m x 100 mm · Inner: 200 m x 78 mm |
| GM35806 | 150 mm AUTOCLAVE | Box of 4 units | Flat rolls · Outer: 200 m x 150 mm · Inner: 200 m x 128 mm |
| GM35807 | 200 mm AUTOCLAVE | Box of 2 units | Flat rolls · Outer: 200 m x 200 mm · Inner: 200 m x 178 mm |
| GM35808 | 250 mm AUTOCLAVE | Box of 2 units | Flat rolls · Outer: 200 m x 250 mm · Inner: 200 m x 228 mm |
| GM35809 | 300 mm AUTOCLAVE | Box of 2 units | Flat rolls · Outer: 200 m x 300 mm · Inner: 200 m x 278 mm |
| GM35816 | 400 mm AUTOCLAVE | 1 unit | Flat rolls · Outer: 200 m x 400 mm · Inner: 200 m x 378 mm |
Product data recorded in the manufacturer’s catalogue:
| Manufacturer | GIMA |
|---|---|
| Device type | Medical device |
| Device class | I |
| CND | S01020101 |
| GMDN | 13735 |
4. Equivalences and Cross-References
References 35803 to 35816 correspond to the manufacturer’s official catalogue. As of today, there is no equivalence table validated by the manufacturer identifying another manufacturer’s code as a direct substitute for these rolls. At category level, they can be compared with other paper-film mixed sterilisation packaging available on the market (for example, the self-adhesive mixed pouches of the BOP DI series from SPS Medical / Pergut Medical or the BOP TI flat pouches), but these are products with a different closure system or a different format (sealed pouch versus roll), so they are not interchangeable 1:1. Before proposing a substitution in a tender, check against the corresponding manufacturer’s technical data sheet and IFU the exact dimensions, materials, available indicators and applicable regulations.
5. Selection tips and variants of interest
- Width selection: choose the roll width according to the device to be packaged: 50–75 mm for small instruments, 100–150 mm for medium instruments and 200–400 mm for trays and large packs.
- Filling rule: do not fill the package to more than two thirds of its internal capacity and leave sufficient margin for sealing, allowing proper circulation of the sterilising agent and drying.
- Sealing: flat rolls are sealed with a heat sealer suitable for paper-film packaging; cut the required length and apply the seal at the ends following the machine instructions.
- Cycle verification: always check the colour change of the steam/gas indicators after sterilisation before considering the process valid, and record batch, date and contents on the labelling.
- Storage: store the rolls in a cool, dry place protected from direct light; observe the single-use principle (do not re-sterilise).