Description
1. Product Presentation
The AEROS 4800 from Dixion is the reference ICU ventilator for intensive care units, emergency departments and critical areas that need to cover with a single device from paediatric to adult patients, including backup ventilation in the event of apnoea. It is the superior evolution of the AEROS family and is designed to offer complete invasive and non-invasive mechanical ventilation, with optimised patient-ventilator synchronisation that improves comfort and clinical safety at every stage of treatment. Distributed by Cahnos, it is the ideal solution for equipment supply projects, ventilator fleet renewal and expansion of critical care beds.
2. Technical explanation of the equipment
The AEROS 4800 is an electronic intensive care ventilator that combines volume-controlled and pressure-controlled ventilation with advanced lung protection strategies (PRVC and BIVENT/APRV). Its triggering system combines flow trigger and pressure trigger to synchronise with the patient’s inspiratory effort, avoiding false triggers and reducing the work of breathing. It incorporates automatic tube compensation (ATC), intelligent suction that maintains ventilation without interruption during aspiration, manual breath, inspiratory and expiratory hold, waveform freeze and nebulisation. Continuous monitoring of volumes, pressures, FiO2 and expired CO2 (etCO2), together with measurement of lung mechanics, P/V manoeuvres and the spontaneous breathing trial (SBT), allow the clinician to assess the patient’s response and adjust treatment in a personalised way during the initiation, stabilisation, weaning and rehabilitation phases.
3. Technical Specifications
The parameter configuration of the AEROS 4800 ventilator allows independent ranges for adult and paediatric patients, optimising respiratory treatment with precise control of each variable.
Parameter configuration — Ranges by patient type
| Parameter | Adult | Paediatric |
|---|---|---|
| Ventilation frequency (f) — VCV | 1–120/min | 2–120/min |
| Ventilation frequency (f) — NIV | 4–20/min | 4–40/min |
| Inspiratory time (Ti) | 0.1–12 s | 0.1–5 s |
| Tidal volume (VT) | 50–2500 ml | 10–300 ml |
| Pause (Tpause) | 0–4 s | 0–2.5 s |
| SIMV ventilation frequency | 1–60/min | 1–50/min |
| Expiratory sensitivity (% of maximum flow) | 5%–80% | 5%–80% |
Pressure and oxygen parameters
| Parameter | Range / specification |
|---|---|
| Inspiratory pressure (Pinsp) | 0–100 cmH2O |
| Supplement | 0–90 cmH2O |
| Pressure support in NIV (Psupp NIV) | 0 to (32-CPAP) cmH2O |
| PEEP | 0–50 cmH2O |
| PEEP in NIV | 2–20 cmH2O |
| CPAP in NIV | 2–20 cmH2O |
| O2 concentration (FiO2) | 21%–100% |
Trigger sensitivity
| Trigger type | Range / specification |
|---|---|
| Flow trigger (adult) | 0.3–20 L/min |
| Flow trigger (paediatric) | 0.5–20 L/min |
| Pressure trigger | −20 to 0 cmH2O |
Ventilation modes
| Type | Modes |
|---|---|
| Controlled | VCV (A/C), PCV (A/C), PRVC |
| Synchronised | SIMV (VCV)+PSV, SIMV (PCV)+PSV, SIMV (PRVC)+PSV |
| Spontaneous | SPONT/CPAP+PSV |
| Advanced | BIVENT/APRV+PSV |
| Non-invasive (NIV) | NIV/CPAP+PSV, NIV/PCV |
Advanced functions
| Function | Description |
|---|---|
| Apnoea ventilation | Automatic backup ventilation for respiratory pauses |
| Lung mechanics | Monitoring of compliance, resistances, intrinsic PEEP and time constant |
| Automatic tube compensation (ATC) | Compensates for the endotracheal tube resistance in each cycle |
| Intelligent suction | Maintains ventilation and pressure during suction of secretions |
| Manual breathing | Manual triggering of cycles at the clinician’s request |
| Inspiration/expiration hold | Pauses for measurement and recruitment manoeuvres |
| Waveform freeze | Detailed analysis of pressure-time and flow-time curves |
| Nebulisation | Administration of nebulised drugs without interrupting ventilation |
| Non-invasive ventilation (NIV) | NIV/CPAP and NIV/PCV support with leak compensation |
| EtCO2 measurement | CO2 concentration at the end of expiration |
| P/V manoeuvre | Pressure-volume curve to assess respiratory mechanics |
| Spontaneous breathing trial (SBT) | Assessment of ventilator weaning |
| Patient type | Adult and paediatric |
4. Equivalences and Cross-References
The AEROS 4800 belongs to the Dixion ventilation range (Vertrieb medizinischer Geräte GmbH, Düsseldorf, Germany), distributed by Cahnos. It shares the configuration platform and family with the AEROS 4600 (SKU AEROS4600), a dual invasive and non-invasive ventilator for intensive care and emergency care, with which it shares much of the same parameter ranges and ventilation modes. As lower references in the same family, the AEROS 4300 (SKU AEROS4300) and the integrated station AEROS 4300S (SKU AEROS4300S) stand out, aimed at emergency care, resuscitation and intra-hospital transport. On the market, functionally comparable equipment from other adult-paediatric ICU ventilation manufacturers with invasive and non-invasive modes, PRVC and BIVENT/APRV is used, with which it shares the philosophy of lung protection and advanced monitoring. Our technical-commercial team can confirm equivalences, accessory compatibilities and conditions for equipment supply and ventilator fleet renewal projects.
5. Selection tips and variants of interest
To choose the most appropriate configuration, consider the usage scenario: if you need advanced invasive and non-invasive ventilation with lung protection (PRVC, BIVENT/APRV) for adult and paediatric patients in the ICU and critical areas, the AEROS 4800 is the reference option in the family. If your need focuses on a dual device for intensive care and emergency care with a more compact profile, the AEROS 4600 may be considered. For emergency care, resuscitation or intra-hospital transport, consult the AEROS 4300 or the AEROS 4300S station with trolley, humidifier, circuit and O2 cylinder support. Always verify the availability of medical gas (air and oxygen outlets), the support system (column, trolley or bed mount) and the electromagnetic compatibility regulations applicable in your facility. Our technical-commercial department will advise you on the selection, commissioning, training and maintenance of the equipment.