Description
1. Commercial Introduction
The SEDA Suture PREMISUT in its ½ circle configuration with HR40 taper-point needle is the classic reference in the Cahnos range of non-absorbable silk sutures for soft tissue approximation, ligatures, ophthalmology and skin surgery. It combines the tradition and handling of natural silk with a cylindrical-body, 40 mm taper-point needle, offering atraumatic passage through tissue and excellent knot security. Supplied in a box of 12 individual sterile sachets, with batch traceability and stable factory reference 542HR4020.
2. Technical Explanation
The braided silk PREMISUT is made from natural protein fibres (fibroin from Bombyx mori) processed to remove natural waxes and gums, allowing a compact and uniform braid. The wax and silicone coating reduces capillarity (wicking effect), improves passage through tissues and provides excellent knot security (recommended technique 2 = 1 = 1 = 1). Being non-absorbable, the thread maintains high tensile strength for more than three months, followed by gradual encapsulation by connective tissue, ensuring prolonged wound support during the critical healing phase. The 40 mm HR40 needle, with ½ circle curvature, cylindrical body and grooved taper point, is especially suitable for soft, well-vascularised tissues (fasciae, peritoneum, muscle, mucosa and conjunctiva), as the point separates fibres without cutting them and minimises tissue trauma. USP 2/0 gauge (metric EP 3) with thread diameter between 0.30 and 0.339 mm, black colour for high visibility in the surgical field. Sterilised by ethylene oxide (EO); single-use product, not re-sterilisable.
3. Technical specifications and tables
3.1. General characteristics
| Parameter | Specification |
|---|---|
| Brand / Quality | PREMISUT (standard line) |
| Factory reference | 542HR4020 |
| Reference Cahnos (SKU) | AM542HR4020 |
| Thread material | 100 % natural multifilament silk, coated with wax and silicone |
| Thread structure | Braided with grooved surface, coated |
| Absorption | Non-absorbable; gradual encapsulation, wound support for more than 3 months |
| Gauge / Size | USP 2/0 (metric EP 3); conforming to USP Pharmacopoeia and European Pharmacopoeia |
| Approximate thread diameter | 0.30 – 0.339 mm |
| Thread length | 75 cm (free end, without loop) |
| Thread colour | Black identifier (high visibility in surgical field) |
| Needle shape | ½ circle (1/2 circular), cylindrical body |
| Tip type | Conical / cylindrical (Taperpoint), grooved |
| Needle size | HR 40 (40 mm) |
| Configuration | 1 needle per unit, 1 thread of 75 cm |
| Recommended knotting technique | 2 = 1 = 1 = 1 |
| Sterilisation | Ethylene oxide (EO) – sterile; single use, not re-sterilisable |
| Indications | Approximation of soft tissues, ligatures, ophthalmology and skin surgery |
| Commercial presentation | Box of 12 sterile individual sachets; batch traceability |
3.2. Gauge equivalence according to USP/EP Pharmacopoeia (non-absorbable silk suture)
| USP size | Metric measurement (EP) | Approximate thread diameter |
|---|---|---|
| 2/0 | 3 | 0.30 – 0.339 mm |
| 0 | 3.5 | 0.35 – 0.399 mm |
| 1 | 4 | 0.40 – 0.499 mm |
| 2 | 5 | 0.45 – 0.499 mm |
The thread diameter complies with the official USP Pharmacopoeia ranges for non-absorbable sutures, without exceeding its limits.
3.3. Family PREMISUT HR40 (½ circle needle of 40 mm, taperpoint conical point, black thread)
| Needle profile | USP size | Thread | Reference |
|---|---|---|---|
| HR 40 (standard) | 2/0 | 1 × 75 cm (black) | 542HR4020 / AM542HR4020 (this product) |
| HR 40 Prime | 2/0 | 1 × 75 cm (black, Prime range) | P542HR4020 / AMP542HR4020 |
| HR 40 (standard) | 0 | 1 × 75 cm (black) | 542HR4010 / AM542HR4010 |
| HR 40 (standard) | 1 | 1 × 75 cm (black) | 542HR4001 / AM542HR4001 |
| HR 40 (standard) | 2 | 1 × 75 cm (black) | 542HR4002 / AM542HR4002 |
| HR 40 Prime | 2/0 | 2 × 75 cm (black, double thread) | P542HR4020.2 / AMP542HR4020.2 |
The same HR40 taper-point needle is available in other gauges (USP 0, 1 and 2), in Prime quality, in double-thread format and in shorter needle profiles (HR20, HR30, HR37) depending on the surgical field access.
4. Equivalences and Cross-References
Reference 542HR4020 corresponds to the official nomenclature of the PREMISUT range (HR40 profile: ½ circle needle with cylindrical body and 40 mm taper-point tip, USP 2/0 gauge, 75 cm black grooved thread) and Cahnos markets it without altering characters: no digit may be added, removed or reordered when ordering or tendering. Non-absorbable black braided silk is one of the most standardised natural suture families on the market (commercial types Silkam®, MERSILK® / PERMA-HAND™). The following comparisons are functional, at category level, and must be confirmed against the technical data sheet or instructions for use (IFU) of each manufacturer before use in tendering or direct substitution.
| Manufacturer / Brand | Family / Model | Reference Code / SKU | Equivalence Level | Technical Differences / Remarks |
|---|---|---|---|---|
| B. Braun / Aesculap | Silkam® Black – natural multifilament braided and impregnated silk, wax-coated | C0770477 – SILKAM BLACK 2/0 (3) 75 CM HR40 | Comparable by category (not interchangeable 1:1) | Code confirmed in the official B. Braun catalogue and in GUDID. Material (coated braided silk), USP 2/0 gauge (metric EP 3), black colour, ½ circle curvature, round body, taper point, HR40 needle size of 40 mm and thread length of 75 cm all match. The exact coating (wax/paraffin versus wax and silicone) and the sterilisation method differ. Confirm IFU and CE/MDR classification before substituting. |
| Ethicon (Johnson & Johnson MedTech) | PERMA-HAND™ / MERSILK® – non-absorbable black braided silk, coated | W584G – MERSILK / PERMA-HAND 2/0 (3) 75 cm LH 40 mm | Comparable by category (not interchangeable 1:1) | Code verified in GUDID and in authorised distribution. Material (braided silk), USP 2/0 gauge, black colour, thread length (75 cm), ½ circle curvature, LH needle with 40 mm taper point and presentation in a box of 12 units all match. The exact coating (wax without silicone declared in the public data sheet) and the internal needle designation (LH) differ. Confirm IFU and CE/MDR classification before substituting. |
Other homologous families of non-absorbable braided silk —such as MERSILK® / PERMA-HAND™ (Ethicon / Johnson & Johnson MedTech), SILKAM® (B. Braun / Aesculap), Sofsilk™ (Medtronic / Covidien), DemeSILK (DemeTECH) or black silk from Clarben / Laboratorio Aragó— appear in public sources with needles, gauges and lengths different from this configuration; they must be verified property by property (material, gauge, curvature, point family, needle size, thread length, colour, coating and CE/MDR regulatory classification) in their official technical data sheets and in GUDID/EUDAMED before any substitution. There is no manufacturer-validated equivalence table; consult the technical department before use in tendering. This needle geometry (HR40, USP 2/0, 75 cm black thread) is only fully safely interchangeable within the PREMISUT system itself. The brands cited belong to their respective manufacturers and are mentioned solely as an equivalence reference.
5. Selection tips and variants of interest
Tissue to suture: the HR taper-point needle (cylindrical body) is indicated for lax, well-irrigated soft tissues, vascular planes, ligatures and precision approximation in ophthalmology and skin surgery; for skin, mucous membranes and firm tissues, consider the reverse-cutting profiles (HS/DS series) from the same PREMISUT silk range.
Gauge: USP 2/0 (metric EP 3) is very versatile, ideal for skin, subcutaneous tissue and soft tissues under moderate tension. For finer planes, consider USP 3/0 from the same family; for higher loads, USP 0 (Ref. AM542HR4010), USP 1 (Ref. AM542HR4001) or USP 2 (Ref. AM542HR4002) are available on the same HR40 needle.
Needle length: the 40 mm HR40 profile covers medium-depth access. For smaller fields, the HR20 (Ref. AM542HR2020), HR30 (Ref. AM542HR3020) and HR37 families are available in the same USP 2/0 gauge.
Prime quality: if the procedure requires maximum sharpness consistency from batch to batch, the same geometry and gauge are available in the Prime range (Ref. AMP542HR4020), a direct equivalent within the same manufacturer.
Double presentation: to reduce the number of sachets opened per procedure, the same HR40 Prime USP 2/0 configuration is available in double thread 2×75 cm (Ref. AMP542HR4020.2).
Sterility verification: product sterilised by ethylene oxide and for single-use; check the integrity of the blister and the expiry date before each use, and retain the reference 542HR4020 and the batch number for traceability in the operating theatre. Store below 30 °C, away from moisture and direct sunlight.
For special volume-based B2B conditions, hospital supply agreements or validation of equivalents from other commercial houses, contact the technical-commercial team at Cahnos.





Reviews