Description
1. Commercial Introduction
The NYLON Suture PREMISUT Prime 1/2 Triangular HS36 is the reference of the Prime range of non-absorbable sutures from Cahnos for procedures requiring a high-performance permanent thread mounted on a medium-bodied reverse cutting needle. It combines a nylon monofilament (polyamide 6 or 6.6) in identifying black colour, USP 1 gauge and 75 cm of uncoated grooved thread, with a ½ circle needle with reverse cutting triangular point of 36 mm (HS 36 profile).
It is the usual choice in ophthalmology, neurosurgery and general surgery departments, as well as in purchasing centres and tenders that value in the Prime range superior sharpening precision, batch-to-batch consistency and full traceability. Sterile product, ready to use, presented in a box of 12 units in individual sterile envelopes.
2. Technical Explanation
Nylon (polyamide 6 or 6.6) is a non-absorbable synthetic polymer that remains in the tissue indefinitely, gradually encapsulating and maintaining high tensile strength for more than three months, comfortably covering the critical healing phase. Being an uncoated monofilament, the thread slides easily through tissues, reduces the wicking effect and bacterial drag compared with multifilaments, and minimises the inflammatory reaction. The grooved finish of the strand improves knot retention and handling in the field.
The HS 36 needle of 36 mm has a ½ circle curvature with a reverse cutting triangular point: the cutting edge cuts in the opposite direction to the wound, which preserves the skin edge and prevents the cut from widening in the puncture bed, facilitating passage through firm and non-distensible tissues. The USP 1 gauge (metric EP 4) provides high resistance for tissues with medium tension, and the 75 cm of thread offer sufficient length for continuous sutures or multiple stitches without re-threading. Sterilisation by ethylene oxide (EO); single-use product, not re-sterilisable.
3. Technical Specifications
3.1. General characteristics
| Parameter | Specification |
| Brand / Quality | PREMISUT Prime (premium range) |
| Thread material | Nylon monofilament (polyamide 6 or 6.6), uncoated |
| Thread structure | Monofilament with grooved surface |
| Coating | None |
| Absorption | Non-absorbable; gradual encapsulation, wound support for more than 3 months |
| Gauge / Size | USP 1 (EP metric 4), conforming to USP and European Pharmacopoeia gauges |
| Approximate thread diameter | 0.40 – 0.499 mm |
| Thread length | 75 cm (free end, without loop) |
| Thread colour | Black identifier (high visibility in surgical field) |
| Needle shape | 1/2 circle (½ circular), grooved |
| Tip type | Reverse cutting triangular |
| Needle size | HS 36 (36 mm) |
| Configuration | 1 needle per unit, grooved thread |
| Sterilisation | Ethylene oxide (EO) – sterile |
| Use | Single-use, sterile. Do not re-sterilise |
| Indications | Approximation of soft tissues, ligatures, ophthalmology and neurosurgery |
| Commercial presentation | Box of 12 units (sterile individual sachets); batch traceability |
| Factory reference | P642HS3601 |
| Reference Cahnos (SKU) | AMP642HS3601 |
| Regulatory standard | Complies with USP/EP Pharmacopoeia; CE marking (MDR) |
3.2. Gauge equivalence according to USP/EP Pharmacopoeia (synthetic non-absorbable suture)
| USP size | Metric measurement (EP) | Approximate thread diameter |
| 2/0 | 3.0 | 0.30 – 0.339 mm |
| 0 | 3.5 | 0.35 – 0.399 mm |
| 1 (this product) | 4.0 | 0.40 – 0.499 mm |
| 2 | 5.0 | 0.50 – 0.599 mm |
3.3. PREMISUT HS 36 family (½ circle needle, 36 mm, reverse cutting triangular point, black 75 cm thread)
| Needle profile | USP size | Thread | Factory reference | SKU Cahnos |
| HS 36 (standard) | 0 | 1 × 75 cm (black) | 642HS3610 | AM642HS3610 |
| HS 36 (standard) | 1 | 1 × 75 cm (black) | 642HS3601 | AM642HS3601 |
| HS 36 (standard) | 2 | 1 × 75 cm (black) | 642HS3602 | AM642HS3602 |
| HS 36 Prime | 0 | 1 × 75 cm (black) | P642HS3610 | AMP642HS3610 |
| HS 36 Prime | 1 | 1 × 75 cm (black) | P642HS3601 | AMP642HS3601 (this product) |
| HS 36 Prime | 2 | 1 × 75 cm (black) | P642HS3602 | AMP642HS3602 |
4. Equivalences and Cross-References
The reference P642HS3601 corresponds to the official nomenclature of the Prime range PREMISUT (HS 36 profile: ½ circle needle with reverse cutting triangular point of 36 mm, USP 1 gauge, black grooved 75 cm thread) and Cahnos markets it without altering characters: no digit may be added, removed or reordered when ordering or tendering.
Black non-absorbable nylon monofilament is one of the most standardised permanent suture families on the market. There is no manufacturer-validated equivalence table cross-referencing this reference with other brands for the exact combination of USP 1 + ½ circle reverse cutting needle of 36 mm + black 75 cm strand; the following comparison is functional, at category level, and must be confirmed against the technical data sheet or instructions for use (IFU) of each manufacturer before use in tendering or direct substitution.
| Manufacturer / Brand | Family / Model | Reference code / SKU | Equivalence level | Technical differences / Remarks |
| B. Braun (Aesculap) | DAFILON® blue – uncoated polyamide monofilament, USP 1 (EP 4), 90 cm, HS37 needle of 37 mm (½ circle, cutting triangular body) | C0934046 (verified in official catalogue B. Braun) | Comparable by category (not interchangeable 1:1) | They match in material (polyamide 6/6.6), USP 1 gauge (EP 4), ½ circle curvature, cutting triangular point family (HS) and 1-needle configuration; they differ in strand colour (blue vs black), thread length (90 cm vs 75 cm) and exact needle size (37 mm vs 36 mm, within ±1 mm). Confirm IFU and CE/MDR classification before substituting. |
Other homologous families on the market that must be reviewed property by property (material, gauge, curvature, point family, needle and strand length, colour and CE/MDR regulatory classification) before any substitution: ETHILON® (Ethicon / J&J MedTech, black nylon monofilament), MONOSOF™ / DERMALON™ (Medtronic / Covidien, nylon monofilament), DemeLON™ (DemeTECH), LUXAMID® (LuxSutures) and SUPRAMID® (SMI / B. Braun, polyamide). No codes from these companies have been verified in public sources for the exact HS 36 configuration of 36 mm, USP 1 and black 75 cm strand; they must be confirmed with the manufacturer before use in substitution or tendering.
There is no manufacturer-validated equivalence table; consult the technical department before use in tendering. This needle geometry (HS 36, USP 1, 75 cm) is only fully safely interchangeable within the PREMISUT system itself: the direct equivalent within the same manufacturer is the standard range version with identical configuration (Ref. 642HS3601, SKU AM642HS3601).
5. Selection tips and variants of interest
- Tissue to suture: USP 1 gauge (EP 4) offers high resistance for subcutaneous tissue, skin with moderate tension, fasciae and deep planes. The reverse cutting needle passes through firm tissues without tearing the puncture bed. For finer planes consider USP 0 (AMP642HS3610) or USP 2/0; for a larger gauge, USP 2 (AMP642HS3602) from the same family.
- Prime quality: the Prime range provides maximum sharpening precision and batch-to-batch consistency; for standard use, the same geometry and gauge can be considered in the conventional line (Ref. 642HS3601, SKU AM642HS3601), a direct equivalent within the same manufacturer.
- Needle length: the HS 36 profile of 36 mm covers medium-depth accesses. For smaller fields, the HS 26 family is available; for deeper accesses, the HS 40 profile in USP 1 (AM642HS4001 / AMP642HS4001) or HS 48 with 100 cm thread (AM647HS4801 / AMP647HS4801).
- Thread length: the 75 cm covers continuous sutures or several stitches without re-threading; if you need a longer run, the same HS 40 Prime configuration is available in 100 cm (P647HS4001 / AMP647HS4001) and there are 45 and 100 cm series within the nylon family.
- Sterility verification: sterile product by ethylene oxide and single-use; check the integrity of the blister and the expiry date before each use and keep the reference P642HS3601 together with the batch number for traceability in the operating theatre. Store below 30 °C, away from moisture and direct sunlight.





Reviews