Description
1. Introduction
The safety intravenous catheter with wings and injection port, 24G x 19 mm is the peripheral vascular access solution designed to cannulate small or fragile veins with maximum protection for the professional. It combines holding wings, a Y injection port and a passive safety mechanism that covers the needle tip when withdrawn, making it a common choice in paediatrics, neonatology, geriatrics, emergency departments, oncology and hospitalisation services, for injections as well as infusions and transfusions. Supplied in a box of 50 sterile units, it is an essential ally in occupational risk prevention programmes against sharps (Directive 2010/32/EU).
2. Technical explanation of the device
This peripheral catheter incorporates a safety system with passive and irreversible activation: upon needle withdrawal after venepuncture, the protector advances automatically and covers the tip, without changing the puncture technique or requiring additional manoeuvres from the professional, drastically reducing the risk of accidental needlestick and blood exposure. The transparent structure of the device allows easy confirmation of blood flashback during cannulation, ensuring precise insertion.
The cannula is made of thermosensitive and radiopaque polyurethane (PUR), a flexible and biocompatible material that softens at body temperature, adapting to the venous bed, minimising kinking and irritation of the venous wall. The flexible wings provide a firm grip during puncture and facilitate fixation of the device to the patient’s limb. The Y injection port with universal Luer connector allows administration of medication boluses or flushing of the system without disconnecting the main line, reducing manipulation and risk of contamination. All components are latex-free, DEHP and PVC.
3. Technical Specifications
| Brand / Line | Nipro, SafeTouch family with wings and injection port |
| Reference (SKU Cahnos) | NIC-SF-2425PWP-ET |
| Gauge | 24G (yellow colour code) |
| Size (diameter x length) | 0.7 x 19 mm (3/4″) |
| Reference flow rate | ≈ 17–22 ml/min (SafeTouch 24G series; verify IFU) |
| Cannula material | Thermosensitive, biocompatible and radiopaque polyurethane (PUR) |
| Needle material | Stainless steel with triple bevel |
| Safety system | Passive needle protection (anti-needlestick) |
| Design | Flexible wings + Y injection port with universal Luer connector |
| Connection | Universal Luer |
| Sterility | Sterile, single-use |
| Composition | Latex-free, DEHP and PVC |
| Packaging / Presentation | Individual blister / Box of 50 units |
| Indications | Intravenous injection, infusion and transfusion; small-calibre veins |
| Regulatory classification | CE class IIa medical device; compliant with ISO 10555-5, ISO 7864 and ISO 23908 |
The same SafeTouch line with wings and injection port is available in gauges 16G, 18G, 20G and 22G with different lengths, allowing the entire range of flow rates to be covered from a single supplier.
4. Equivalences and Cross-References
There is no equivalence table validated by the manufacturer that cross-references this reference with specific codes from other brands as direct 1:1 equivalents. However, after verifying the official catalogues of the manufacturers and the regulatory databases, safety catheters of gauge 24G have indeed been identified with which it shares clinical function and indication:
| Manufacturer / Brand | Family / Model | Code / SKU | Equivalence level | Technical differences / Remarks |
|---|---|---|---|---|
| Nipro (same family) | SafeTouch with wings (without port) | NIC-SF-2425PW-ET | Comparable by category (not interchangeable 1:1) | Same gauge 24G and length 19 mm, same material and safety system; without Y injection port. Not interchangeable if the protocol requires a port. |
| Nipro (same family) | SafeTouch with wings and injection port | NIC-SF-2419-PWP-ET | Comparable by category (not interchangeable 1:1) | Same gauge 24G, design with wings and Y port; usable length of 25 mm (longer). Reference verified in family data sheet. |
| Terumo | Surshield™ Versatus (with needle guard) | SR+DS2419PD | Comparable by category (not interchangeable 1:1) | 24G x 0.74 x 19 mm, yellow, box of 50 units. Same gauge and length and needle protection system; differs in design (without wings) and needle guard mechanism. Reference verified in official data sheet. |
| B. Braun | Introcan Safety®-W (PUR, with wings) | 4253523-01 | Comparable by category (not interchangeable 1:1) | 24G (0.7 x 19 mm), PUR, with wings and passive safety with clip; differs in hub design and absence of injection port. Verify manufacturer’s technical data sheet before substituting. |
At category level, the product is also comparable with peripheral safety catheters from brands such as BD (Insyte Autoguard and Nexiva lines) and Vygon (Leader-Cath). Before substituting in a tender or internal protocol, the professional must verify in each brand’s technical data sheet: equivalent gauge, cannula length, material (polyurethane versus other polymers), type of safety mechanism (active/passive), presence of wings and injection port, and packaging per box. There is no equivalence table validated by the manufacturer that cross-references these references with other brands; consult the technical department before use in a tender.
5. Selection tips and variants of interest
The 24G gauge is the smallest in the conventional range of peripheral catheters and is indicated for small, fragile or difficult-to-access veins: paediatric patients, neonates, geriatric patients and low-flow intravenous therapies. If you require a higher infusion flow rate, consider the larger gauges of the same SafeTouch family with wings and injection port: 22G x 25 mm (ref. NIC-SF-2225PWP-ET), 22G x 32 mm (ref. NIC-SF-2232PWP-ET), 20G, 18G or 16G for high-flow transfusions. If your protocol requires an injection port, always check that the code includes the PWP suffix (Y port), as the line offers variants without port (P/PW suffix) that are not interchangeable 1:1. Verify the integrity of the sterile packaging before each use, use the device only once and dispose of it in an approved sharps container.